Quantum QLS1-200Recalls & reported problems
Spectrum Medical Limited · Quantum Reservoir Level Sensor - Single Head 2.0m · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05060434421890), received between Oct 17, 2024 and Apr 28, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem4
- Failure to Sense2
- Electrical /Electronic Property Problem1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2026Failure to Sense
THE USER REPORTED A LEVEL SENSOR INCORRECT MEASUREMENT DURING THE CASE. NO ADVERSE CONSEQUENCES OCCURRED FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25015711 ↗MalfunctionJul 25, 2025Failure to Sense; Failure to Sense
USER REPORTED THAT DURING A CASE, THE PUMP WAS STOPPED AS THE SAFE FLOW BUBBLE DETECTOR HAD TRIGGERED. THE USER NOTICED THAT THE LEVEL SENSOR HAD NOT RECOGNISED THAT THE RESERVOIR LEVEL HAD DROPPED BELOW THE PROTECTED LEVEL WHICH SHOULD HAVE STOPPED THE PUMP PRIOR TO THE SAFE FLOW BUBBLE DETECTOR. ONCE THE PUMP STOPPED, THE RESERVOIR VOLUME RETURNED, THE SAFE FLOW BUBBLE ALARM WAS CLEARED, THE PUMPED RESTARTED AND THE CASE CONTINUED WITHOUT ISSUE. THE PUMP WAS STOPPED FOR LESS THAN 20 SECONDS. TESTING THE LEVEL SENSOR AFTERWARDS, THE USER PHYSICALLY MANIPULATED THE SENSOR AND SAW THE LEDS INTERMITTENTLY FLASH ON AND OFF. THE USER OPTED TO CHANGE OUT THE LEVEL SENSOR. THERE WAS NO PATIENT HARM AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22607334 ↗MalfunctionJul 2, 2025Device Sensing Problem; Device Sensing Problem
THE USER REPORTED THAT THE SAFE FLOW LEVEL WAS TRIGGERED WHEN ADDING CRYSTALLOID INTO THE RESERVOIR. THE EVENT SELF-RESOLVED QUICKLY AND THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22382395 ↗MalfunctionJun 11, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
USER REPORTED THAT AFTER 25 SECONDS OF THE COMMENCEMENT OF BYPASS, WITH FLUID LEVEL ABOVE SAFE FLOW LEVEL AND DURING BLOOD MIXING, THE LEVEL SENSOR MALFUNCTION (10810) ALARMED WITH A PUMP STOP FOR 4.5 SEC. SENSOR ERROR SELF CORRECTED AND FLOWS RETURNED TO PRE-ALARM SETTING. THERE WAS NO PATIENT HARM AS A RESULT OF THIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22182781 ↗MalfunctionJun 11, 2025Device Sensing Problem; Device Sensing Problem
USER REPORTED THAT PROTECTED LEVEL TRIGGERED INCORRECTLY ON COMMENCEMENT OF BYPASS. NO ERROR DISPLAYED JUST TRIGGERING PROTECTED LEVEL ALARM. USER DISABLED LEVEL AND SYNCED WITH BLOOD. REST OF CASE CONTINUED WITHOUT ISSUE. NO PATIENT HARM AS A RESULT OF THIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22182780 ↗MalfunctionMay 14, 2025Output Problem; Output Problem
USER REPORTED THAT DURING BYPASS, ALL LOOKED GOOD, BUT THEN THE PUMP STOPPED DUE TO LEVEL ALARM. ACCORDING TO PERFUSIONIST IT WAS NOT POSSIBLE TO DISABLE THE LEVEL AT ALL BECAUSE IT WAS "GREYED OUT". AS BLOOD FLOW WAS ZERO AND IT WAS NOT POSSIBLE TO RESTART THE PUMP, THEY CLOSED VENOUS AND STARTED HAND CRANKING GIVING VOLUME BACK TO THE PATIENT. AFTER THAT THEY REPLACED THE CRADLE OF THE LEVEL SENSOR. THEN SPECTRUM CLINICAL SPECIALIST ENTERED ROOM. WHILE RESERVOIR FULL YELLOW LED WAS OFF AND RED WAS ON. THE USER DISCONNECTED THE LEVEL SENSOR AND RECONNECTED. THE USER HAD TO PRESS THE SENSOR BETTER INTO CRADLE AND THEN BOTH LEDS WERE ON. THE USER CONTINUED BYPASS WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT HARM AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22019260 ↗MalfunctionOct 17, 2024Device Sensing Problem; Device Sensing Problem
SAFE FLOW ALARM STARTED 'FLICKERING' ON AND OFF, AFTER CIRCULATING FLUID FOR AROUND 1.5HR. THE PERFUSIONIST OPTED TO REPLACE THE LEVEL SENSOR BEFORE GOING ON BYPASS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20470898 ↗MalfunctionOct 17, 2024Device Sensing Problem; Device Sensing Problem
PERFUSIONIST COMPLAINED OF MULTIPLE 'SAFE FLOW' LEVEL ALARMS WHEN BLOOD VOLUME IN RESERVOIR WAS SUFFICIENT (HIGHER THAN LEVEL OF THE SENSOR CRADLE). THESE WERE COINCIDING WITH THE ADDITION OF CRYSTALLOID VOLUME TO THE RESERVOIR THROUGHOUT THE CASE. ADDITIONALLY, THE PERFUSIONIST COMPLAINED OF A PROTECTED FLOW LEVEL ALARM APPROXIMATELY 20 SECONDS AFTER THE ADDITION OF PACKED RED CELLS TO THE RESERVOIR. THE PUMP STOPPED MOMENTARILY BUT RESTARTED ALMOST IMMEDIATELY. WHEN WEANING FROM BYPASS (MANUALLY WITH A TUBING CLAMP), FLOW HAD BEEN REDUCED TO AROUND 1LPM; SURGEON REQUESTED TO COME BACK UP TO 3LPM. PERFUSIONIST RELEASED CLAMP AND INCREASED FLOW TO 3LPM. REPORTED 'SAFE FLOW' AGAIN WITH ADEQUATE VOLUME IN RESERVOIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20470891 ↗
FDA clearance and classification
Cleared through Special 510(k) K202733 on Oct 16, 2020 (decision: Substantially Equivalent). Submitted by Spectrum Medical , Ltd. as “Quantum Ventilation Module”. The FDA review took 28 days. FDA record ↗
The FDA regulates product code DTX (gas control unit, cardiopulmonary bypass) under 21 CFR 870.4300 as a Class II device. 21 CFR 870.4300 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Quantum QLS1-200
Has the Quantum QLS1-200 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05060434421890), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Quantum QLS1-200?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Oct 17, 2024 and Apr 28, 2026. In those reports, the most common specific device problems are “Device Sensing Problem” (4), “Failure to Sense” (2) and “Electrical /Electronic Property Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Quantum QLS1-200?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Quantum QLS1-200 FDA approved?
It was cleared through Special 510(k) K202733 on Oct 16, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Quantum QLS1-200, and is it still sold?
The FDA's device database lists Spectrum Medical Limited as the company responsible for it (brand Quantum, model QLS1-200). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gas Control Unit, Cardiopulmonary Bypass devices
- Electronic Gas Blender 10 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass95 reports
- Quantum QVM2-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass36 reports
- Quantum QVM4-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass18 reports
- Quantum QVM2-N-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass12 reports
- Quantum QVMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass10 reports
- Electronic Gas Blender 5 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass5 reports
- Electronic Gas Blender 2 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass3 reports
- Quantum QLS1-120Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page