Spectrum Medical Limited · Quantum Ventilation Module - DISS (Hook Mount) · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Quantum QVM2 ventilation module is a hook-mounted gas control unit used during cardiopulmonary bypass to help manage the patient's breathing. The reports mostly describe gas output and suction problems, alarms that would not clear, and pressure and carbon dioxide readings drifting far from the values measured by other equipment, sometimes forcing staff to swap in another device mid-procedure. There are 37 reports, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 05060434422071), received between Jul 5, 2024 and Jun 8, 2026 — classified as 36 malfunctions.

The most frequently reported specific device problem is “Gas Output Problem” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 5 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 6 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas Output Problem12
  • Suction Problem4
  • Incorrect Measurement3
  • Decrease in Pressure2
  • Device Sensing Problem2
  • Imprecise or Erratic Results2
  • Pressure Problem2
  • Sensing Intermittently2

Patient problems

  • No Clinical Signs, Symptoms or Conditions35
  • Low Oxygen Saturation1

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2026Gas Output Problem

    NOMOLINE ICON FLASHING RED AFTER CPB INITIATION WHEN THEY TRIED SYNCING THE FIRST BLOOD GAS AT 10:37. PIPELINE PROTECTOR CHANGED BUT CONTINUED ALARMING. NOMOLINE REPLACED WITH NEW DISPOSABLE BUT ALARMING CONTINUED. USER ATTEMPTED TO USE OUTSIDE SUCTION ON THE NOMOLINE PORT ON THE QVM2, TO SEE IF ANY CONDENSATION COULD BE REMOVED FROM THE PORT ITSELF. NO RESOLUTION. UNPLUGGED/REPLUGGED NOMOLINE IN MULTIPLE TIMES AND EVEN REPLACED WITH A SECONDARY NOMOLINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428992 ↗
  • MalfunctionJun 8, 2026Gas Output Problem

    DURING CASE, PCO2 @37 VALUE TRENDING VERY DANGEROUSLY LOW, DESPITE REPEATED CAPTURE/SYNC. EVEN IMMEDIATELY AFTER PERFORMING A SYNC, PCO2 VALUE WOULD DROP BACK IMMEDIATELY TO HIGH-20S VALUE. USER CHECKED WAG LINE CONNECTIONS, SET SWEEP TO 10 LPM TO BLOW OFF ANY CONDENSATION, AND ALSO EMPTIED THE SPUTUM TRAPS. PIPELINE PROTECTORS WERE REPLACED. FACTORY RESET OF PCO2 WAS NOT PERFORMED, NOR WAS THE NOMOLINE REPLACED AT THE TIME. PCO2 CONTINUED ALARMING BELOW THEIR LOWER LIMIT THRESHOLD THROUGHOUT THE CASE AND REQUIRED SERIAL ABGS TO BE PERFORMED FOR PATIENT SAFETY. ALL OTHER PARAMETERS TRENDED AS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428946 ↗
  • MalfunctionMar 20, 2026Incorrect Measurement

    USER REPORTED AN ERROR WITH THE PRESSUREMENT MEASUREMENT ON THE VENTILATION MODULE. THE VALUE FROM THE PRESSURE SENSOR DID NOT MATCH AN EXTERNAL PRESSURE SENSOR. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650378 ↗
  • MalfunctionMar 20, 2026Incorrect Measurement

    USER REPORTED AN ERROR WITH THE PRESSURE MEASUREMENT ON THE VENTILATION MODULE. THE VALUE MEASURED BY THE SENSOR PRESSURE DID NOT MATCH THE EXTERNAL PRESSURE SENSOR. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650276 ↗
  • MalfunctionMar 19, 2026Sensing Intermittently

    USER REPORTED THAT THE SAFE LEVEL WAS INTERMITTENTLY INOPERABLE THROUGHOUT THE PROCEDURE. THIS COULD RESULT IN POTENTIAL PATIENT HARM IF SAFE FLOW WAS INACTIVE WHEN EXPECTED TO ENGAGE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641554 ↗
  • MalfunctionMar 18, 2026Gas Output Problem

    USER REPORTED THAT THE VENTILATION MODULE ISA FAILED, RESULTING IN HARDWARE ERRORS DURING THE PROCEDURE. THERE WAS NO PATIENT HARM; HOWEVER, THE DEVICE MALFUNCTIONED AND SPECTRUM MEDICAL CONSIDERS THIS EVENT TO BE REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632325 ↗
  • MalfunctionMar 18, 2026Gas Output Problem

    USER REPORTED THAT THE VENTILATION MODULE MALFUNCTIONED, CAUSING AN INTERRUPTION IN SWEEP. THERE WAS NO PATIENT HARM; HOWEVER, THIS TYPE OF EVENT IS CONSIDERED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626507 ↗
  • MalfunctionMar 12, 2026Incorrect Measurement

    USER REPORTED INACCURATE PRESSURE MEASUREMENT FROM PRESSURE SENSOR. THERE WAS NO PATIENT HARM; HOWEVER, THIS TYPE OF EVENT IS CONSIDERED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24580312 ↗
  • MalfunctionJan 19, 2026Unexpected Shutdown

    USER REPORTED THAT THE VENTILATION MODULE TURNED OFF TWICE DURING USE, WHICH IS CONSIDERED REPORTABLE BY SPECTRUM MEDICAL. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109422 ↗
  • MalfunctionDec 19, 2025Pressure Problem; Pressure Problem

    USER REPORTED THE INABILITY TO REGULATE THE VACUUM PRESSURE. THERE WAS NO PATIENT HARM; HOWEVER, SPECTRUM MEDICAL CONSIDERS THIS TYPE OF EVENT TO BE REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850405 ↗
  • MalfunctionDec 18, 2025Decrease in Pressure; Decrease in Pressure

    USER REPORTED THAT VACUUM REGULATION WAS LOST DURING THE PROCEDURE. THERE WAS NO PATIENT HARM; HOWEVER, SPECTRUM MEDICAL CONSIDERS THE INABILITY TO REGULATE VACUUM TO BE REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838939 ↗
  • MalfunctionDec 18, 2025Decrease in Pressure; Decrease in Pressure

    USER REPORTED THAT VACUUM REGULATION WAS LOST DURING THE PROCEDURE. THERE WAS NO PATIENT HARM; HOWEVER, SPECTRUM MEDICAL CONSIDERS THE INABILITY TO REGULATE VACUUM TO BE REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838791 ↗
  • MalfunctionDec 12, 2025Device Sensing Problem; Device Sensing Problem

    USER REPORTED THAT THE LEVEL SENSOR DID NOT TRIGGER AN ALERT TO INDICATE A LACK OF FLUID. THIS TYPE OF EVENT IS CONSIDERED REPORTABLE AS THERE IS THE POTENTIAL FOR SERIOUS PATIENT HARM TO OCCUR IF THE RESERVOIR LEVEL WERE TO BE EMPTIED WITHOUT THE USER REALIZING. THERE WAS NO PATIENT HARM IN THIS INSTANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789006 ↗
  • MalfunctionDec 11, 2025Gas Output Problem; Gas Output Problem

    USER REPORTED THAT ALTHOUGH THE PO2 MEASUREMENT WAS IN THE 300S ON THE DEVICE, ANOTHER THIRD-PARTY DEVICE SHOWED THAT THE PO2 HAD DROPPED SIGNIFICANTLY.

    Patient problems: Low Oxygen Saturation

    MAUDE report 23778353 ↗
  • MalfunctionOct 22, 2025Pressure Problem; Pressure Problem

    USER REPORTED AN ISSUE WITH THE VENTILATION MODULE VENTING TO ATMOSPHERE TO PREVENT PRESSURE BUILD-UP. THERE WAS NO PATIENT HARM; HOWEVER, THIS ISSUE COULD CREATE A RISK TO THE PATIENT / USER IF THIS WERE TO PERSIST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356189 ↗
  • MalfunctionOct 17, 2025Gas Output Problem; Gas Output Problem

    USER REPORTED SWEEP WOULD NOT CHANGE FROM ZERO. THIS IS CONSIDERED REPORTABLE PER SPECTRUM MEDICAL'S CRITERIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23326315 ↗
  • MalfunctionAug 8, 2025Suction Problem; Suction Problem

    USER REPORTED ISSUES DURING THE CASE REGULATING THE VACUUM. THERE WAS NO PATIENT HARM; HOWEVER, THIS TYPE OF EVENT IS CONSIDERED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22750088 ↗
  • MalfunctionAug 7, 2025Suction Problem; Suction Problem

    THE PERFUSIONIST REPORTED THE INABILITY TO ADJUST THE VACUUM DURING THE PROCEDURE. THE ISSUE WAS ABLE TO BE RESOLVED AND THERE WAS NO PATIENT HARM; HOWEVER, SPECTRUM MEDICAL CONSIDERS THIS TYPE OF EVENT TO BE REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22737865 ↗
  • MalfunctionAug 6, 2025Device Sensing Problem; Device Sensing Problem

    THE LEVEL SENSOR WAS TRIGGERING INAPPROPRIATELY WHEN THERE WAS SUFFICIENT VOLUME IN THE RESERVOIR. SPECTRUM MEDICAL CONSIDERS THIS TO BE A REPORTABLE EVENT DESPITE THERE BEING NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22726692 ↗
  • MalfunctionMay 29, 2025Gas Output Problem; Gas Output Problem

    USER REPORTED THEY WERE UNABLE TO CHANGE SWEEP SET POINT AND RECEIVED A HARDWARE ERROR. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22111630 ↗
  • MalfunctionMay 20, 2025Inaccurate Delivery; Inaccurate Delivery

    USER REPORTED THAT SET-POINTS FOR SWEEP WERE NOT BEING MET. USER WAS ABLE TO RESOLVE ISSUE AND USE THE DEVICE SUCCESSFULLY SINCE THIS EVENT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22055347 ↗
  • MalfunctionMay 19, 2025Suction Problem; Suction Problem

    USER REPORTED INABILITY TO REGULATE VACUUM. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22045854 ↗
  • MalfunctionMay 1, 2025Gas Output Problem; Gas Output Problem

    INITIALLY THE USER REPORTED "ISSUES" WITH PCO2 VALUES. SERVICE TEAM CHECKED THE VENTILATION MODULE, CALIBRATED AND DOWNLOADED THE LOGS, AND FOUND NO FAULTS. USER THEN REPORTED THAT THEY WERE NOT GETTING AIRFLOW AT THE OUTLET. SERVICE MADE A VISIT TO THE HOSPITAL AND CONFIRMED LACK OF FLOW. THE SERVICE TEAM WAS REQUESTED TO REPLACE THE QVM AND CONDUCT TESTS. USER REPORTED THAT AFTER QVM WAS REPLACED THERE WERE NO FURTHER ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941108 ↗
  • MalfunctionMar 7, 2025Sensing Intermittently; Sensing Intermittently

    INTERMITTENT COMMUNICATION WITH FLOW SENSORS CAUSED UNINTENDED PUMP SLOWING OR STOPPING. THERE WAS NO PATIENT HARM AND THE FAULT WAS DURING VALIDATION; HOWEVER, SPECTRUM MEDICAL CONSIDERS A PUMP STOP TO BE A REPORTABLE EVENT DUE TO THE POTENTIAL EFFECT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21546389 ↗
  • MalfunctionJan 31, 2025Improper Flow or Infusion; Improper Flow or Infusion

    THIS VMSR REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT MODULE QVM2 (PART NUMBER: 51-000082-00) EXPERIENCED AN ISSUE WHEN REGULATING GAS SWEEP. THERE WAS NO PATIENT HARM ASSOCIATED WITH THESE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21282762 ↗

FDA clearance and classification

Cleared through Special 510(k) K202733 on Oct 16, 2020 (decision: Substantially Equivalent). Submitted by Spectrum Medical , Ltd. as “Quantum Ventilation Module”. The FDA review took 28 days. FDA record ↗

The FDA regulates product code DTX (gas control unit, cardiopulmonary bypass) under 21 CFR 870.4300 as a Class II device. 21 CFR 870.4300 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Quantum QVM2-D-HM

Has the Quantum QVM2-D-HM been recalled?

No FDA recall record in our data names this exact device identifier (DI 05060434422071), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Quantum QVM2-D-HM?

The FDA's MAUDE database lists 36 adverse event reports for this device, received between Jul 5, 2024 and Jun 8, 2026. In those reports, the most common specific device problems are “Gas Output Problem” (12), “Suction Problem” (4) and “Incorrect Measurement” (3); the most common patient problem is “Low Oxygen Saturation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Quantum QVM2-D-HM?

None of the reports are classified as deaths, 0 as injuries and 36 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Quantum QVM2-D-HM FDA approved?

It was cleared through Special 510(k) K202733 on Oct 16, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Quantum QVM2-D-HM, and is it still sold?

The FDA's device database lists Spectrum Medical Limited as the company responsible for it (brand Quantum, model QVM2-D-HM). Its record shows it as in commercial distribution.