Quantum QVM4-D-HMRecalls & reported problems
Spectrum Medical Limited · Quantum ECLS Ventilation Module - DISS (Hook Mount) · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 05060434422859), received between Sep 11, 2024 and May 1, 2026 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Gas Output Problem” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas Output Problem11
- Device Sensing Problem2
- Calibration Problem1
- Electrical /Electronic Property Problem1
- Failure to Calibrate1
- Failure to Sense1
- Incorrect Interpretation of Signal1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- Low Oxygen Saturation1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2026Gas Output Problem
USER REPORTED THAT THE SWEEP WENT TO ZERO WITHOUT ANY INDICATION TO THE CLINICIAN. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25056112 ↗MalfunctionApr 16, 2026Calibration Problem
THE USER REPORTED INACCURATE MEASUREMENTS DISPLAYED ON THE SYSTEM AND INSTABILITY OF THE SYSTEM TO REACH THE SETPOINT. THIS BEHAVIOR IS CONSIDERED REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24896428 ↗MalfunctionMar 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
USER REPORTED INCORRECT MEASUREMENTS FROM THE SAMPLE LINE ON THE QUANTUM VENTILATION MODULE. THERE WAS NO PATIENT HARM; HOWEVER, THIS TYPE OF INACCURATE MEASUREMENT IS CONSIDERED REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24727614 ↗MalfunctionMar 23, 2026Device Sensing Problem
THE USER REPORTED GAS FLOW ISSUES THAT CAUSE THE PATIENT'S OXYGEN SATURATION TO DROP, REQUIRING CLINICAL INTERVENTION TO RE-STABILIZE PATIENT LEVELS.
Patient problems: Low Oxygen Saturation
MAUDE report 24660975 ↗MalfunctionFeb 12, 2026Gas Output Problem
USER REPORTED THAT SWEEP DROPPED TO ZERO UNEXPECTEDLY. THERE WAS NO PATIENT HARM; HOWEVER, THIS TYPE OF ISSUE IS CONSIDERED REPORTABLE BY SPECTRUM MEDICAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24322745 ↗MalfunctionDec 22, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT THE DEVICE WOULD NOT ALLOW THE USER TO BLEND IN SWEEP GAS. THERE WAS NO PATIENT HARM; HOWEVER, THIS TYPE OF EVENT IS CONSIDERED REPORTABLE BY SPECTRUM MEDICAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23864457 ↗MalfunctionSep 5, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT THE SWEEP REVERTED TO ZERO AFTER CHANGING THE SETPOINT. THERE WAS NO PATIENT HARM; HOWEVER, INTERRUPTION TO SWEEP IS CONSIDERED REPORTABLE BY SPECTRUM MEDICAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22976311 ↗MalfunctionJun 12, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT THE DEVICE WAS NOT DELIVERING OXYGEN, WHICH WAS IDENTIFIED DURING SET-UP. WHILE THERE WAS NO PATIENT HARM, IF THE ERROR WERE TO RECUR DURING A CASE, IT COULD RESULT IN PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22194897 ↗MalfunctionJan 30, 2025Failure to Sense; Failure to Sense
USER REPORTED THE QUANTUM VENTILATION MODULE READING A NEGATIVE PRESSURE INCORRECTLY. INCORRECT MEASUREMENT FROM A LEVEL SENSOR IS CONSIDERED A REPORTABLE MALFUNCTION. THERE WAS NO PATIENT HARM DUE TO THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21272134 ↗MalfunctionDec 9, 2024Gas Output Problem; Gas Output Problem
CUSTOMER IDENTIFIED A SETPOINT ERROR WHEN EXCEEDING A SWEEP OF 1 LPM DURING A CASE AND CLINICIAN REBOOTED THE MODULE TO ADDRESS THE ERROR. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878088 ↗MalfunctionDec 5, 2024Gas Output Problem; Gas Output Problem
VENT MODULE IDENTIFIED A HARDWARE ERROR AND USER WAS UNABLE TO SET SWEEP GAS FLOW RATE. USER PERFORMED HARD REBOOT OF DEVICE, WHICH RESOLVED THE ERROR. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20853737 ↗MalfunctionNov 28, 2024Gas Output Problem; Gas Output Problem
PERFUSIONIST REPORTED THAT THE LEVELS OF SWEEP AND FIO2 WERE RAPIDLY FLUCTUATING DURING A CASE. A BACKUP WAS USED TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER INABILITY TO MANAGE GAS FLOW TO BE A REPORTABLE EVENT, DESPITE NO HARM TO THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20805694 ↗MalfunctionNov 28, 2024Gas Output Problem; Gas Output Problem
PERFUSIONIST REPORTED THAT THAT THEY WERE UNABLE TO CHANGE THE VALUE FOR SWEEP. WE CONSIDER THE LOSS OF ABILITY TO CONTROL GAS MANAGEMENT TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20804291 ↗MalfunctionNov 27, 2024Gas Output Problem; Gas Output Problem
PERFUSIONIST REPORTED THAT THEY WERE UNABLE TO CONTROL THE SWEEP DURING A CASE. WE CONSIDER INABILITY TO MANAGE GAS FLOW TO BE A REPORTABLE EVENT, DESPITE NO HARM TO THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20796542 ↗MalfunctionNov 11, 2024Failure to Calibrate; Failure to Calibrate
PERFUSIONIST REPORTED THAT THE GAS CALIBRATION FAILED. FOLLOWING THIS, ERROR MESSAGES APPEARED REPEATEDLY, INCLUDING WHEN THE SPECIALIST ATTEMPTED TO INCREASE THE FLOW. THERE WAS NO HARM TO THE PATIENT INVOLVED, DESPITE SWITCHING TO A DIFFERENT GAS MANAGEMENT SYSTEM. WE CONSIDER THE LOSS OF ABILITY TO CONTROL GAS MANAGEMENT TO BE REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20650135 ↗MalfunctionOct 29, 2024Gas Output Problem; Gas Output Problem
ERROR CODES 2883 AND 2884 DISPLAYED. THE PERFUSIONIST WAS UNABLE TO ADJUST SWEEP AND FIO2. AS GAS FLOW WAS INTERRUPTED, THIS IS CONSIDERED A REPORTABLE EVENT. THERE WAS NO EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20555881 ↗MalfunctionOct 25, 2024Electrical /Electronic Property Problem; Incorrect Interpretation of Signal; Electrical /Electronic Property Problem; Incorrect Interpretation of Signal
USER REPORTED THAT THEY WERE EXPERIENCING ISSUES WITH NOMOLINE READING EITHER SHOWING "0" OR ">180". THIS IS BELIEVED TO BE AN INCORRECT MEASUREMENT FROM THE SENSOR. ALSO OF NOTE THE VENTILATION MODULE SHOWED "HARDWARE ERROR (2881)" FOR BOTH PCO2 AND PO2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20532437 ↗MalfunctionSep 11, 2024Device Sensing Problem; Device Sensing Problem
USER REPORTED THAT DURING A CASE A PACO2 READING WAS FLUCTUATING BETWEEN READING CORRECTLY AND THEN DROPPING TO 0 FOR AN EXTENDED PERIOD. PAO2 ALSO BEGAN READING "!!!". THE USER TRIED REPLACING THE NOMOLINE AND DRYING OUT THE PORT ON THE QVM4 WHERE THE NOMOLINE IS ATTACHED USING A FAN FOR OVER 8 HOURS BUT THE READING NEVER RECOVERED, AND THE PATIENT WAS TAKEN OFF THE PUMP THE FOLLOWING DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20189619 ↗
Other Gas Control Unit, Cardiopulmonary Bypass devices
- Electronic Gas Blender 10 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass98 reports
- Quantum QVM2-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass36 reports
- Quantum QVM2-N-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass11 reports
- Quantum QVMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass10 reports
- Quantum QLS1-200Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass8 reports
- Electronic Gas Blender 5 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass5 reports
- Electronic Gas Blender 2 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass3 reports
- Quantum QLS1-120Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page