Biotronik Passeo-35 HP 10/40/75Recalls & reported problems
Creagh Medical Limited · Class 2
All models of Biotronik Passeo-35 HP (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05391525781600), received between Apr 20, 2016 and Mar 18, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Detachment of Device or Device Component1
- Difficult to Remove1
- Failure to Deflate1
- Fracture1
- Inflation Problem1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient2
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2024Fracture; Fracture
THE BALLOON CATHETER WAS USED FOR A TREATMENT OF A LESION WITH 70% STENOSIS IN THE MILD LEFT BRACHIOCEPHALIC FISTULA (BCF). THE BALLOON WAS PREPARED AS PER IFU. THE BALLOON WAS INFLATED AT 18 ATM AND OPERATED AS EXPECTED DURING THE FIRST INFLATION. A SECOND INFLATION AT 18 ATM WAS PERFORMED AND THE BALLOON BURST BEFORE THE RATED BURST PRESSURE WAS REACHED. THERE WAS NO CALCIFICATION AND NOT TORTUOUS ANATOMY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18926895 ↗InjuryJan 10, 2023Material Fragmentation; Material Rupture; Material Fragmentation; Material Rupture
THE 10MM HIGH PRESSURE BALLOON BURST WHILE INFLATING AT THE RATED BURST PRESSURE, 20ATM. THE BALLOON FRAGMENT WAS STUCK IN THE VEIN FIBROTIC TISSUE AND WAS UNABLE TO BE REMOVED. IT WAS ALSO NOTED THAT A PART OF THE BALLOON WAS ALSO STUCK AT THE TIP OF THE GUIDE SHEATH. SOME FORCE WAS APPLIED TO REMOVE THE BALLOON CATHETER AND GUIDE SHEATH FROM THE VEIN. PARTS OF THE BALLOON WERE ALSO BROKEN AND STUCK IN THE PATIENT ANATOMY NEAR THE PUNCTURE SITE. A VENOUS CUT DOWN WAS PREFORMED TO REMOVE THE REMAINING BALLOON. NO BALLOON FRAGMENTS REMAINED IN THE PATIENT ANATOMY. THE REPORTED ISSUE RESULTED IN PROLONGED PATIENT HOSPITALIZATION. A PERMANENT CATHETER WAS ALSO INSERTED INTO THE PATIENT FOR DIALYSIS DUE TO THE TRAUMA SUSTAINED TO THE FISTULA BRACHIAL CEPHALIC VEIN. IT WAS NOTED THE LESION LOCATION WAS LOCATED DISTALLY. THE APPROXIMATE VESSEL DIAMETER WAS REPORTED TO BE 10MM WITH A STENOSIS DEGREE OF 70% AND LESION LENGTH OF 60MM. THE VESSEL HAD MILD TORTUOSITY AND SEVERE CALCIFICATION.
Patient problems: Foreign Body In Patient
MAUDE report 16133573 ↗MalfunctionSep 30, 2022Failure to Deflate; Failure to Deflate
ON 08SEP2022, MANUFACTURER BECAME AWARE OF THE EVENT, DESCRIBED AS "BALLOON DID NOT DEFLATE PROPERLY". PROCEDURE WAS PERFORMED ON (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15510236 ↗MalfunctionMay 25, 2021Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
DURING INFLATION THE BALLOON BURST OCCURED LONGITUDINALLY AT 16ATM. THE BALLOON WAS SUCCESSFULLY RETRIEVED WITHOUT COMPLICATION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11879574 ↗InjuryAug 2, 2017Material Rupture; Material Rupture
DURING THE TREATMENT OF A TIGHT LESION OF THE CENTRAL VEIN, THE BALLOON RUPTURED RADIALLY INTO TWO PARTS. WHILE WITHDRAWING THE TIP DISLODGED AND STUCK AT THE NECK OF THE STENOSIS AND COULD NOT BE REMOVED. AFTER 10 MINUTES THROMBUS DEVELOPED AT THE LESION. VASCULAR SURGEON ASSESSED THAT MIGRATION OF THE DEVICE WAS NOT POSSIBLE. PATIENT BEING MONITORED.
Patient problems: Foreign Body In Patient
MAUDE report 6760939 ↗MalfunctionJun 22, 2017Difficult to Remove; Difficult to Remove
PROCEDURE WAS SMOOTH UNTIL DEFLATION AND WITHDRAWAL FROM SHEATH. BALLOON WINGS DID NOT REWRAP WELL AND GOT STUCK AT THE SHEATH. DOCTOR REQUESTED PRODUCT INVESTIGATION AND REPORT. PRODUCT WILL BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6661125 ↗MalfunctionApr 20, 2016Inflation Problem; Inflation Problem
THE PASSEO-35 HP 10/40/75 PRODUCT LEAKS DURING INFLATION TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5589985 ↗
Other Catheter, Angioplasty, Peripheral, Transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page