Skip to content

Biotronik Passeo-35 HP 12/40/75Recalls & reported problems

Creagh Medical Limited · Class 2

All models of Biotronik Passeo-35 HP (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05391525781617), received between Apr 20, 2016 and Jul 10, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Deflation Problem1
  • Difficult to Insert1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2020Deflation Problem; Deflation Problem

    AN 8F 45CM ARROW SHEATH WAS USED TO INTRODUCE THE BALLOON. A COOK ROADRUNNER 035 SYSTEM 180CM WIRE WAS USED TO GUIDE THE BALLOON TO THE LESION. A PASSEO-35 4/60/90 BALLOON WAS USED TO PRE-DILATE THE LESION IN THE SUBCLAVIAN VEIN. AFTER PRE-DILATION, THE PASSEO-35HP 12/40/75 BALLOON WAS INTRODUCED TO FURTHER DILATE THE VESSEL. BALLOON WAS INFLATED THROUGH THE 8F 45CM ARROW SHEATH AND DEFLATED AFTER 2 MINUTES AT 18ATM PRESSURE. WHEN BALLOON WAS ABOUT TO BE REMOVED FROM THE SHEATH, PHYSICIAN EXPERIENCED DIFFICULTY IN PULLING THE BALLOON OUT, DESPITE USING FORCE. AFTER NUMEROUS ATTEMPTS, PHYSICIAN REMOVED THE SHEATH ALONG WITH THE BALLOON AND REVEALED THE BALLOON WAS STUCK AT THE DISTAL OPENING OF THE SHEATH AS BALLOON DID NOT DEFLATE AND REWRAP PROPERLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10258894 ↗
  • MalfunctionApr 20, 2016Difficult to Insert; Difficult to Insert

    THE PASSEO-35 HP 12*40*75 CATHETER CANNOT RE-ENTER INTO THE 8FR SHEATH AFTER 1ST INFLATION. PHYSICIAN HAD NO CHOICE BUT TO CHANGE TO 9FR SHEATH FOR THE SECOND TIME ENTRY BY USING THE SAME BALLOON.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5588710 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates