Dynamis AV D1604081475Recalls & reported problems
Creagh Medical Limited · Dynamis PTA UHP 16mm x 40mm x 75cm · Class 2
All models of Dynamis AV (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05391525783369), received between Jan 16, 2019 and May 26, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Remove2
- Accessory Incompatible1
- Difficult to Open or Close1
- Failure to Deflate1
- Failure to Fold1
- Mechanical Jam1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Thrombosis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2021Accessory Incompatible; Failure to Fold; Difficult to Open or Close; Mechanical Jam; Failure to Deflate; Accessory Incompatible; Failure to Fold; Difficult to Open or Close; Mechanical Jam; Failure to Deflate
DURING PROCEDURE THE BALLOON DID NOT DEFLATE/ REFOLD CORRECTLY. THERE WAS DIFFICULTY IN RETRIEVAL AND GUIDE CATHETER TIP WAS DAMAGED. NO COMPLICATIONS OCCURED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11890670 ↗MalfunctionApr 11, 2019Difficult to Remove
CUSTOMER REPORTS THAT AFTER THE PROPER BALLOON INFLATION AND DEFLATION TECHNIQUE WAS PERFORMED, WHEN TRYING TO REMOVE THE BALLOON FROM THE SHEATH, IT COULD NOT RETURN THROUGH THE 8FR SHEATH. IT INSERTED INTO THE SHEATH WITH NO DIFFICULTY, BUT IT WOULD NOT PULL BACK THROUGH THE 8FR SHEATH AFTER THE BALLOON HAD BEEN INFLATED. THE SHEATH HAD TO BE REMOVED FROM THE PATIENT AND A NEW 8FR SHEATH INSERTED BECAUSE THE BALLOON WOULD NOT EXIT THROUGH THE SHEATH. THE PATIENT HAD NO COMPLICATIONS, BUT THIS DID CAUSE UNNEEDED PROCEDURE COMPLICATIONS, INCREASED PROCEDURE TIME AND THE USE OF EXTRA SUPPLIES. THE PATIENT DID HAVE A DVT POST PROCEDURE AND IS BEING TREATED FOR A DVT. IT IS ALSO REPORTED THAT NO PRODUCT WILL BE RETURNING.
Patient problems: Insufficient Information; Thrombosis
MAUDE report 8503381 ↗MalfunctionJan 16, 2019Difficult to Remove
THE CUSTOMER REPORTED THAT THEY WERE NOT ABLE TO WITHDRAW THE BALLOONS THEY SHEATH. THEY REPORTED HAVING TO REMOVE THE SHEATH EACH TIME TO REMOVE THE BALLOON. THE CUSTOMER REPORTED THIS HAPPENING ON SEVERAL OCCASIONS.
Patient problems: Insufficient Information; No Known Impact Or Consequence To Patient
MAUDE report 8252095 ↗
Other Catheter, Angioplasty, Peripheral, Transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page