Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05391525783666), received between May 9, 2024 and May 9, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 9, 2024Fracture; Fracture

    THE BALLOON CATHETER WAS USED IN A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE WITH LEFT COMMON FEMORAL ARTERY (CFA) ACCESS. THE TARGET VESSEL WAS THE POSTERIOR TIBIAL ARTERY. THE BALLOON CATHETER WAS USED TO FINISH THE PROCEDURE. IT WAS NOTED THE TARGET VESSEL WAS HIGHLY CALCIFIED. THERE WAS NO REPORT OF RESISTANCE BEING ENCOUNTERED DURING ADVANCEMENT/REMOVAL. THE PHYSICIAN WAS A HIGH VOLUME PROVIDER AND CONFIRMED DEFLATION PRIOR TO REMOVAL. DURING REMOVAL, THE CATHETER SHAFT SEPARATED IN THE MID TO DISTAL PORTION OF THE SHAFT. THE DISTAL SEGMENT COULD NOT BE RETRIEVED FOLLOWING SEVERAL HOURS OF EFFORT. DUE TO THE LENGTH OF THE CATHETER OUTSIDE OF THE PATIENT, THE REMAINING CATHETER WAS KEPT STRAIGHT OR IN LARGE LOOPS. THE PHYSICIAN STATED THE MIDSHAFT JOINT WAS RETAINED IN THE PATIENT POPLITEAL ARTERY DURING THE PROCEDURE. THE PATIENT WAS TRANSFERRED TO A HOSPITAL AND PLACED ON HEPARIN OVERNIGHT. THE NEXT MORNING, AN ADDITIONAL OPEN AND CUT-DOWN PROCEDURE WAS PERFORMED TO SURGICALLY REMOVE THE DISTAL SEGMENT. THE BALLOON PORTION OF THE CATHETER WAS STILL CONNECTED TO THE SHAFT. NO FURTHER CATHETER SEGMENTS REMAINED. THERE WAS NO FURTHER PATIENT HARM.

    Patient problems: Foreign Body In Patient

    MAUDE report 19280215 ↗