Skip to content

Sublime™ 014 Rx PTA Balloon Dilatation Catheter SRA-DC14-20150Recalls & reported problems

Creagh Medical Limited · 014 Radial Catheter 2mm x 150mm x 250cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05391525783673), received between Dec 3, 2024 and Dec 3, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Loss of or Failure to Bond1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 3, 2024Loss of or Failure to Bond; Fracture; Loss of or Failure to Bond; Fracture

    THE BALLOON CATHETER USED DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE OF A CALCIFIED TIBIAL ARTERY. THE BALLOON WAS INFLATED THREE TIMES. ON THE THIRD INFLATION, THE BALLOON RUPTURED. THE BALLOON WAS INFLATED TO 8ATM, 4ATM, AND 6ATM RESPECTIVELY. THE PHYSICIAN ATTEMPTED TO REMOVE THE BALLOON. IN DOING SO, THE DISTAL PORTION OF THE BALLOON SEPARATED AND WAS LEFT IN THE PATIENT'S ARTERY. SEVERAL ATTEMPTS WERE MADE TO RETRIEVE THE BALLOON SEGMENT, ALL OF WHICH WERE UNSUCCESSFUL. THE PHYSICIAN WAS ABLE TO STENT THE ARTERY, ENSURING THE BALLOON SEGMENT WOULD NOT MIGRATE. THERE WAS NO FURTHER PATIENT HARM. THE PROCEDURE WAS DELAYED DUE TO THE UNSUCCESSFUL RETRIEVAL ATTEMPTS.

    Patient problems: Foreign Body In Patient

    MAUDE report 20839680 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page