Skip to content

Sublime™ 014 Rx PTA Balloon Dilatation Catheter SRA-DC14-35150Recalls & reported problems

Creagh Medical Limited · 014 Radial Catheter 3.5mm x 150mm x 250cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05391525783888), received between Mar 18, 2024 and Mar 18, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Failure to Deflate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deflate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2024Failure to Deflate; Failure to Deflate

    THE BALLOON WAS USED DURING AN ATHERECTOMY PROCEDURE THROUGH THE LEFT RADIAL ARTERY. THE PHYSICIAN PERFORMED AND ANGIOGRAM WITH DISTAL SUPERFICIAL FEMORAL ARTERY (SFA) ATHERECTOMY USING A COMPETITOR'S LASER CATHETER TECHNOLOGY. UPON CONCLUSION OF THAT THERAPY, THE BALLOON WAS USED. THE BALLOON WAS DELIVERED TO THE TARGET LESION AND INFLATED WITHOUT ISSUE. DURING REMOVAL, THE BALLOON COULD NOT BE DEFLATED. SEVERAL TECHNIQUES WERE ATTEMPTED WITHOUT RESOLUTION. THE PHYSICIAN DECIDED TO USE A 21G NEEDLE TO POP THE BALLOON AND REMOVE. THE ISSUE OCCURRED AFTER THE INITIAL INFLATION THAT INTENDED TO BE INFLATED FOR APPROXIMATELY 30 SECONDS. THE BALLOON WAS INFLATED TO 6ATM. THE TORTUOSITY WAS REPORTED AS NORMAL. TROUBLESHOOTING INCLUDE MULTIPLE DEFLATION ATTEMPTS WITH THE INFLATOR, MANUAL PULL BACK WITH A SYRINGE ON SIDE, AND MANUAL PULL BACK DIRECT FROM THE BALLOON. THE PHYSICIAN ALSO INSERTED A GUIDEWIRE THROUGH THE BALLOON AND APPLYING MANUAL PRESSURE TO THE OUTSIDE OF THE PATIENT'S LEG TO POP. THE REPORTED ISSUE RESULTED IN A PROCEDURE DELAY OF APPROXIMATELY 25 MINUTES. THERE WAS NO FURTHER PATIENT HARM. THERE WAS NO FURTHER PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18927266 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page