Skip to content

Sublime™ Radial Access 018 RX PTA Dilatation Catheter SRA-DC18-50120Recalls & reported problems

Creagh Medical Limited · 018 Radial RX PTA, 5Fr, 5.0 mm x 120 mm, 220 cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05391525786360), received between Sep 18, 2024 and Sep 18, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 18, 2024Fracture; Material Separation; Fracture; Material Separation

    THE BALLOON CATHETER WAS USED FOR A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE. RADIAL ANGIOPLASTY WAS PERFORMED IN ORDER TO AVOID AN OPEN PROCEDURE. THE SITE REPORTED THE BALLOON BURST AFTER INFLATION TO APPROXIMATELY 10ATMS. THE PHYSICIAN WAS USING THE "KISSING BALLOON TECHNIQUE" FOR BIFURCATIONS. DURING REMOVAL, THE BALLOON CATHETER SHAFT SEPARATED FROM THE MAIN BODY NEAR THE RAPID EXCHANGE PORT DURING REMOVAL FROM A 6FR SHEATH. THE THOUGHT WAS THAT THE SITE DID NOT ALLOW FULL DEFLATION OF THE BALLOONS PRIOR TO PULLING BACK INTO THE SHEATH. THE PHYSICIAN HAD TO PERFORM A VESSEL CUTDOWN TO RETRIEVE THE SEPARATED SEGMENTS AND FINISH WITH ENDARTERECTOMY. THE SITE INDICATED THE BALLOONS SEPARATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) AND PROXIMAL POPLITEAL ARTERY. MULTIPLE ATTEMPTS AT DEFLATION WERE MADE. THE WIRE WAS NOT REMOVED. THE BALLOON SEPARATED APPROXIMATELY 10-20 CM FROM THE BALLOON MARKER BAND. THE FURTHER INTERVENTION OF ENDARTERECTOMY WAS REQUIRED FOR THE CALCIFIED ILIAC ARTERY AND BALLOON SEGMENT REMOVAL. BOTH BALLOON SEGMENTS WERE INTACT UPON REMOVAL. THE ENTIRE PROCEDURE LASTED APPROXIMATELY SEVEN HOURS WITH A FOUR HOUR DELAY DUE TO THE BALLOON SEPA

    Patient problems: Foreign Body In Patient

    MAUDE report 20251878 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page