pORTAL® Steerable Hydrophilic Guidewire PORT-14-200-1Recalls & reported problems

Phenox Limited · The pORTAL Guidewire is a disposable medical device designed for single use only. It is designed to selectively introduce and position catheters and other interventional devices within target vessels. In order to achieve this purpose, the guidewire must be capable of being steered through blood vessels. This necessitates pushing and torqueing capability within the product. The design of the distal section of the guidewire provides steerability, while maintaining the flexibility necessary to negotiate the tortuous path of the blood vessel network. Coatings are placed on the device to improve the lubricity and ease in its advancement through the guide catheter and the blood vessels. Two accessories are provided within the package. These are a Torquer device and an introducer Needle. Both can be used to aid the successful use of the device. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
1
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05391530910040), received between Jan 13, 2021 and Feb 26, 2026 — classified as 1 death, 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Break2
  • Fracture1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body Embolism1
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 26, 2026Fracture

    BASED ON THE LIMITED EVENT INFORMATION PROVIDED: THE GUIDEWIRE BROKE OR SNAPPED WHILE THE PHYSICIAN WAS PERFORMING A MECHANICAL THROMBECTOMY OF A BLOOD CLOT. THE BREAKAGE LED TO AN OCCLUSION IN A SMALLER BLOOD VESSEL FURTHER DOWNSTREAM DISTAL IN THE BRAIN, LIKELY BECAUSE A FRAGMENT OF THE BROKEN WIRE OR A DISPLACED CLOT BLOCKED BLOOD FLOW. ON THE (B)(6) 2025, PHENOX LTD RECEIVED MORE INFORMATION REGARDING THE EVENT INVOLVING PORTAL GUIDEWIRE: EVENT DESCRIPTION AS DESCRIBED BY THE COMPLAINANT: A PATIENT WITH AN ACUTE ISCHEMIC STROKE UNDERWENT MECHANICAL THROMBECTOMY VIA RIGHT FEMORAL ACCESS. INITIAL CATHETER NAVIGATION WAS DIFFICULT DUE TO A SEVERE ANATOMICAL LOOP IN THE LEFT INTERNAL CAROTID ARTERY, PREVENTING FULL ADVANCEMENT OF THE ASPIRATION CATHETER. PROCEDURE: A LONG 7F 70 CM FLEXOR SHEATH CATHETER WAS INTRODUCED FOR SELECTIVE CATHETERIZATION OF THE LEFT INTERNAL CAROTID ARTERY, BUT ITS LENGTH WAS INSUFFICIENT TO REACH THE PROXIMAL SEGMENT OF THE LEFT ICA. AFTER DIAGNOSTIC IMAGING, THE DIAGNOSTIC BERENSTEIN 5F CATHETER WAS EXCHANGED FOR A LONG 7F FLEXOR SHEATH POSITIONED AT THE ORIGIN OF THE LEFT INTERNAL CAROTID ARTERY, USING A 0.35 HYDROPHILIC GUIDEWIRE (STIFF 260 CM). CONTI

    Patient problems: Foreign Body Embolism

    MAUDE report 24455336 ↗
  • InjuryJul 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    FOLLOWING THE COMPLICATION DURING THE ENDOVASCULAR PROCEDURE ,WHERE A PORTION OF THE PORTAL GUIDEWIRE (DISTAL SEGMENT) DETACHED AND REMAINED WITHIN THE INTERNAL CAROTID ARTERY-THE PHYSICIAN TOOK IMMEDIATE REMEDIAL STEPS TO ENSURE PATIENT SAFETY AND COMPLETE THE TREATMENT WITHOUT ADVERSE OUTCOMES. THE DETACHED SEGMENT OF THE MICROGUIDE WAS PROMPTLY IDENTIFIED AS RESIDING IN THE CEREBRAL VASCULATURE (MIDDLE CEREBRAL ARTERY), . THE PHYSICIAN ATTEMPTED 3 INTERVENTIONAL TECHNIQUES TO RETRIEVE THE RETAINED FRAGMENT: AN INFLATED BALLOON WAS USED IN AN ATTEMPT TO DISLODGE AND WITHDRAW THE GUIDE SEGMENT. SNARE CATHETER RETRIEVAL FOR MECHANICAL CAPTURE. A STENT RETRIEVER WAS DEPLOYED AS A FINAL MANEUVER, WHICH SUCCESSFULLY RETRIEVED THE RETAINED FRAGMENT. SOFIA DISTAL ACCESS CATHETER AND NEURON MAX SHEATH SUPPORTED PROCEDURAL SUCCESS. BASED ON THE INITIAL INFORMATION PROVIDED - THE PATIENT EXPERIENCED NO NEUROLOGICAL DEFICITS, INDICATING THAT THE REMEDIAL ACTIONS WERE EFFECTIVE AND DID NOT COMPROMISE PATIENT SAFETY.

    Patient problems: Foreign Body In Patient

    MAUDE report 22479890 ↗
  • InjuryJul 2, 2025Break; Break

    INITIAL EVENT AS REPORTED: IT WAS REPORTED THAT A MICROGUIDE 0.014 PORTAL, WAS BROKEN INSIDE A VASCO 10 MICROCATHETER AND BECAME AN INTRAVASCULAR FOREIGN BODY IN THE CEREBRAL ARTERIES. THE REASON FOR TREATMENT WAS CEREBRAL EMBOLIZATION. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JUN-2025 AS REQUESTED BY THE MANUFACTURER: PRIOR TO THE PROCEDURE, THE ENTIRE GUIDEWIRE WAS SUFFICIENTLY HYDRATED IN PHYSIOLOGICAL SOLUTION IN ACCORDANCE WITH THE DEVICE'S INSTRUCTIONS FOR USE (IFU). THE TIP WAS SHAPED USING THE PROVIDED INSERTION TOOL BEFORE USE. NO DAMAGE, SUCH AS KINKS, WAS NOTED ON THE GUIDEWIRE BEFORE THE PROCEDURE. THE GUIDEWIRE WAS NOT REMOVED OR REINSERTED DURING THE PROCEDURE, AND THEREFORE WAS NOT RE-EXAMINED OR WIPED WITH GAUZE AT ANY POINT. THE USER DID NOT EXPERIENCE RESISTANCE DURING EITHER ADVANCEMENT OR RETRACTION OF THE GUIDEWIRE. THERE WAS NO ADVANCEMENT OR WITHDRAWAL AGAINST RESISTANCE WITHIN THE PATIENT'S ANATOMY, AND THE GUIDEWIRE WAS NOT EXCESSIVELY ROTATED DURING NAVIGATION. THE PROCEDURE WAS CONDUCTED UNDER CONTINUOUS FLUOROSCOPIC MONITORING. AN ANCILLARY DEVICE, A VASCO 10 MICROCATHETER (BALT), WAS USED IN CONJUNCTION WITH THE GUIDEWIRE. THIS MICROCATHETER WAS HYDRAT

    Patient problems: Insufficient Information

    MAUDE report 22383093 ↗
  • InjuryDec 12, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A PATIENT WITH AN ACOM ANEURYSM AND TORTUOUS VESSELS WAS TREATED WITH PORTAL STEERABLE HYDROPHILIC GUIDEWIRE. A 6F GUIDING CATHETER AND A COMPETITOR MICROCATHETER WERE ALSO USED DURING THE PROCEDURE. IT WAS REPORTED THAT, DURING THE PROCEDURE, THE DEVICE WAS INSPECTED PRIOR TO USE AND NO ISSUES WERE FOUND. BOTH GUIDEWIRE AND MICROCATHETER WERE HYDRATED BEFORE USE. THE DEVICE DISTAL END WAS WIPED USING CLOTH/GAUZE BEFORE USE. NO ABNORMAL RESISTANCE WAS EXPERIENCED WHEN ADVANCING OR RETRACTING THE DEVICE. THE PHYSICIAN DID THE PUNCTURE, PLACED THE INTRODUCER, RAISED THE CATHETER AND PASSED THE PORTAL GUIDEWIRE, WHICH NAVIGATED VERY EASILY. BUT HALFWAY THROUGH THE PROCEDURE, PHYSICIAN NOTICED THAT THE CATHETER HAD MOVED AWAY FROM THE LOCATION. AN ATTEMPT WAS MADE TO CROSS FROM LEFT A1 TO RIGHT A2 WITH THE MICROCATHETER AND THE PORTAL, BUT BECAUSE THE GUIDING CATHETER WAS UNSTABLE DUE TO THE CERVICAL TORTUOSITY, THE GUIDEWIRE WAS DISLOCATED FROM POSITION. THEN IT WAS DECIDED TO START THE PROCEDURE AGAIN. THE CATHETER WAS NAVIGATING WHEN THE GUIDEWIRE BROKE. THE POSITION OF THE DISTAL TIP WAS AT A1 AND THE BROKEN PROXIMAL PART WAS AT THE PETROUS SEGMENT OF THE ICA.THE PHYSICIAN MANAGED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18311434 ↗
  • InjurySep 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE FOLLOWING DESCRIPTION WAS PROVIDE FROM THE HEALTHCARE PROFESSIONAL: "THEY STARTED TO USE THE PORTAL, AND BEFORE REACHED THE LESION, REALIZED THAT THE GUIDE WIRE PORTAL HAD BROKEN IN HALF. THEY DIDN'T PUSH OR DID ANY PRESSION AND FEEL ANY KINKING BECAUSE THE ANATOMY WASN'T COMPLEX. THEY TRIED TO PULL IT OUT WITH A SOLITAIRE, A SNARE AND IT DIDN'T WORK. THEY ENDED UP USING A DERIVO´S FLOW DIVERTER. THE PATIENT WAS HOSPITALISED FOR 1 DAY, AS USUAL." DURING THE PROCEDURE, ONE PORTAL GUIDEWIRE WAS USED AND, BEFORE THE DEVICE REACHED THE LESION, IT WAS NOTED THAT THE PORTAL WIRE HAD BROKEN IN HALF. A MICROCATHETER WAS ALSO USED DURING THE PROCEDURE. IT WAS REPORTED THAT THE DEVICE WAS NOT PUSHED AND NO PRESSURE WAS ADDED. KINKING WAS NOT FELT AS PATIENT'S ANATOMY WAS NOT COMPLEX. AN ATTEMPT WAS MADE TO REMOVE THE DETACHED PORTION OF THE DEVICE USING A THROMBECTOMY RETRIEVER AND SNARE BUT THE ATTEMPT WAS NOT SUCCESSFUL. A FLOW DIVERTER DEVICE WAS THEN USED AND THE DETACHED PART OF THE PORTAL DEVICE WAS SUCCESSFULLY REMOVED FROM THE PATIENT. THERE WAS NO PROLONGATION OF PATIENT HOSPITALIZATION AND NO FURTHER INJURY WAS REPORTED. THE ADDITIONAL INFORMATION WILL BE OBTAINED, AND INVESTIG

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17841578 ↗
  • MalfunctionJan 13, 2021Break; Material Integrity Problem; Break; Material Integrity Problem

    THIS IS A FOLLOW UP REPORT - SEE INTINAL REPORT FOR THIS INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11164795 ↗

Other Guide, Wire, Catheter, Neurovasculature devices

All Guide, Wire, Catheter, Neurovasculature devices →