Synchro2 S2SPS14300Recalls & reported problems

Stryker Corporation · SUPPORT Guidewire with Hydrophilic Coating · Class 2

All models of Synchro2 (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
1
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327459647), received between Sep 4, 2020 and Jan 3, 2025 — classified as 1 death, 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 18, 2024Open, Classified

    Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

    Z-1941-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture5
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • Hemorrhage/Blood Loss/Bleeding1
  • Hemorrhagic Stroke1
  • Perforation of Vessels1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2025Fracture; Fracture

    THE SUBJECT DEVICE WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE GUIDEWIRE WAS BROKEN/FRACTURED DURING USE. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074243 ↗
  • MalfunctionSep 26, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING THE CHRONIC INTRACRANIAL OCCLUSION PROCEDURE, DURING MANIPULATION WITHIN THE MICROCATHETER, THE GUIDEWIRE FRACTURED. THE PHYSICIAN WAS ABLE TO REMOVE THE ENTIRE FRACTURED GUIDEWIRE AS IT STILL REMAINED IN THE MICROCATHETER. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17818697 ↗
  • MalfunctionJun 1, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ACUTE THROMBUS THROMBECTOMY PROCEDURE, WHEN THE SUBJECT GUIDEWIRE WAS INSERTED INTO THE PATIENTS ANATOMY, IT COULD NOT BE ROTATED. THE OPERATOR WITHDREW THE SUBJECT GUIDEWIRE AND FOUND 5CM FROM THE DISTAL TIP WAS STRETCHED AND FRACTURED. THE OPERATOR REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17045421 ↗
  • DeathDec 16, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE WHILE ADVANCING THE SUBJECT GUIDEWIRE THROUGH THE BALLOON CATHETER IN THE CAROTID SIPHON, PHYSICIAN NOTICED A KINK AND ON REMOVAL THE TIP OF THE SUBJECT GUIDEWIRE WAS FRACTURED. HE ATTEMPTED TO REMOVE THE FRAGMENT USING MULTIPLE STENT RETRIEVERS AND A SNARE DEVICES BUT WAS UNSUCCESSFUL. THE BROKEN FRAGMENT WAS REPORTED STENTED AGAINST THE VESSEL WALL SO IT WAS UNABLE TO RETRIEVED. THE PATIENT SUFFERED A SUBARACHNOID HEMORRHAGE. IT WAS NOT CLEAR THAT THE EVENT WAS A DIRECT RESULT OF THE BROKEN SUBJECT GUIDEWIRE. THE PATIENT'S ANATOMY WAS VERY TORTUOUS AND THERE WAS A SURGICAL DELAY OF 2 HOURS. THE PROCEDURE WAS NOT COMPLETED AND THE PATIENT PASSED AWAY.

    Patient problems: Foreign Body In Patient; Hemorrhagic Stroke

    MAUDE report 13019907 ↗
  • MalfunctionOct 14, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE GUIDEWIRE (SUBJECT DEVICE ) BROKE OFF FROM DISTAL PORTION INSIDE THE PATIENT. NO FRAGMENTS LEFT INSIDE THE PATIENT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT ANY CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12636821 ↗
  • InjurySep 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE PROCEDURE, THERE WAS PERFORATION OF A SMALL VESSEL DISTAL TO THE LESION WHICH LED TO BLEEDING. PERFORATION WAS TREATED USING CODMAN GLUE AND BLEEDING CONTAINED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Perforation of Vessels

    MAUDE report 10494992 ↗

Other GUIDE, WIRE, CATHETER, NEUROVASCULATURE devices