Synchro SELECT SSFT300STRRecalls & reported problems
Stryker Corporation · SOFT STRAIGHT Guidewire with Hydrophilic Coating · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 18, 2024Open, Classified
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
Z-1940-2024 · Root cause: Component design/selection · FDA record ↗
Other Guide, wire, catheter, neurovasculature devices
- Synchro SELECT SSTD215STRStryker Corporation · Guide, wire, catheter, neurovasculature1 recall97 reports
- Synchro SELECT SSTD215PREStryker Corporation · Guide, wire, catheter, neurovasculature1 recall16 reports
- Synchro SELECT SSUP215PREStryker Corporation · Guide, wire, catheter, neurovasculature1 recall11 reports
- Synchro2 S2SPS14300Stryker Corporation · GUIDE, WIRE, CATHETER, NEUROVASCULATURE1 recall1 death6 reports
- Synchro SELECT SSUP300STRStryker Corporation · Guide, wire, catheter, neurovasculature1 recall5 reports
- Synchro2 S2SPP14300Stryker Corporation · GUIDE, WIRE, CATHETER, NEUROVASCULATURE1 recall2 reports
- Mirage 103-0608Micro Therapeutics, Inc. · Guide, wire, catheter, neurovasculature12 reports
- CHIKAI 008 WAIN-CKI-008-200Asahi Intecc CO., Ltd. · GUIDE, WIRE, CATHETER, NEUROVASCULATURE6 reports