Synchro2 S2SPP14300Recalls & reported problems

Stryker Corporation · SUPPORT PRE-SHAPED Guidewire with Hydrophilic Coating · Class 2

All models of Synchro2 (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327459661), received between Dec 22, 2025 and Apr 6, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 18, 2024Open, Classified

    Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

    Z-1941-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture2
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 6, 2026Fracture; Peeled/Delaminated

    THE GUIDEWIRE (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS AND OBSERVED THAT THE DISTAL TIP OF THE SUBJECT DEVICE WAS DAMAGED, STRETCHED AND BROKEN/FRACTURED DURING USE. ALSO, THE PLOY TETRA FLUORO ETHYLENE (PTFE) COATING OF THE SUBJECT DEVICE WAS PEELING. THE SUBJECT DEVICE WAS REPORTED TO BE REPLACED, AND THE PROCEDURE WAS REPORTED TO COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792780 ↗
  • InjuryDec 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SEVERE STENOSIS CASE OF MIDDLE CEREBRAL ARTERIES (MCA), THE PHYSICIAN USED THE SUBJECT GUIDEWIRE AND EXCHANGED IT TO PASS THROUGH THE LESION SITE AND REACH THE DISTAL END OF THE M2 SEGMENT. DURING THE PROCESS OF WITHDRAWAL, APPROXIMATELY 5 CM OF THE SUBJECT GUIDEWIRE'S TIP FRACTURED AT THE STENOSIS SITE. AFTERWARDS, THE PHYSICIAN USED A RETRIEVER STENT TO REMOVE THE SUBJECT GUIDEWIRE FRAGMENT AND CONCLUDED THE PROCEDURE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23873591 ↗

Other GUIDE, WIRE, CATHETER, NEUROVASCULATURE devices

All GUIDE, WIRE, CATHETER, NEUROVASCULATURE devices →