EnSite™ Velocity NavLinkRecalls & reported problems
St. Jude Medical, Inc. · Module · Class 2
All models of EnSite (37)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05414734209953), received between Jul 18, 2018 and Jul 20, 2020 — classified as 5 injuries.
The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Communication or Transmission Problem2
- Detachment of Device or Device Component1
- Display or Visual Feedback Problem1
- Output Problem1
Patient problems
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 20, 2020Output Problem; Output Problem
DURING THE PROCEDURE, AN ERROR MESSAGE WAS DISPLAYED AND THE PROCEDURE WAS CANCELLED. AN ENSITE PRECISION ERROR MESSAGE APPEARED ON THE SCREEN AFTER VALIDATION: ¿NECK PATCH DISCONNECTED¿. TROUBLESHOOTING WAS PERFORMED WITH NO RESOLUTION. AF ABLATION COULD NOT BE COMPLETED AND ONLY A CTI ABLATION WAS PERFORMED. THE PATIENT WAS STABLE WITH NO PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10296082 ↗InjuryJul 1, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
DURING AN ATRIAL FIBRILLATION AND ATRIOVENTRICULAR NODAL RE-ENTRANT TACHYCARDIA (AVNRT) ABLATION PROCEDURE, THE NAVLINK PIN DETACHED WITHIN THE PATCH CONNECTOR AND MAPPING WAS NO LONGER POSSIBLE. THE AVNRT ABLATION WAS COMPLETED WITHOUT MAPPING, HOWEVER THE ATRIAL FIBRILLATION PORTION OF THE PROCEDURE WAS UNABLE TO BE COMPLETED AND THE PROCEDURE WAS CANCELLED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE CANCELLATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10220707 ↗InjuryNov 5, 2019Display or Visual Feedback Problem; Display or Visual Feedback Problem
DURING AN ABLATION PROCEDURE, THE 3D NAVIGATION DID NOT FUNCTION AS EXPECTED. TROUBLESHOOTING WAS ATTEMPTED BY REPOSITIONING PATCHES, REVALIDATING, AND RESTARTING THE COMPUTER AND AMPLIFIER, WITH NO RESOLUTION. EXCHANING THE NAVLINK RESOLVED THE ISSUE AND THE PROCEDURE WAS CONTINUED WITH NO CONSEQUENCES TO THE PATIENT. THE LENGTH OF PROCEDURE WAS EXTENDED BY AN HOUR AND WAS CLINICALLY SIGNIFICANT DUE TO INCREASED HEPARIN ADMINISTRATION AND DIFFICULTIES IN MANAGING SEDATION FOR THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9279383 ↗InjuryAug 29, 2018Communication or Transmission Problem; Communication or Transmission Problem
DURING DEVICE PREPARATION FOR A SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, AN ERROR MESSAGE APPEARED STATING THE NECK PATCH WAS DISCONNECTED AND THE PROCEDURE WAS CANCELLED. THE CONNECTIONS FOR THE NAVLINK AND PATCHES WERE VERIFIED AND APPEARED CORRECT. THE NECK AND CHEST PATCHES WERE REPLACED WITH NO RESOLUTION. A REPLACEMENT NAVLINK WAS UNAVAILABLE AND THE PROCEDURE WAS RESCHEDULED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE CANCELLATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7828765 ↗InjuryJul 18, 2018Communication or Transmission Problem; Communication or Transmission Problem
DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE A CANCELLATION OCCURRED DUE TO AN ERROR MESSAGE. FOLLOWING ANESTHESIA AND SHEATH INSERTION, THE SYSTEM REFERENCE PATCHES WERE PLACED ON THE PATIENT. THREE ERRORS WERE DISPLAYED UPON PATCH PLACEMENT AND COULD NOT BE RESOLVED EVEN AFTER THE PATCHES WERE EXCHANGED. THE PROCEDURE WAS CANCELLED WITH NO ADVERSE CONSEQUENCES ONTO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7698177 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page