EnSite™ NavX™ EN0010-100003331Recalls & reported problems
St. Jude Medical, Inc. · Surface Electrode Kit · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414734210003), received between Aug 26, 2019 and Feb 22, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Communication or Transmission Problem1
- Display or Visual Feedback Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 22, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2021-00032. DURING THE PROCEDURE, AFTER THE PATIENT WAS PREPPED, WHEN THE PATCH WAS VALIDATED, THE SYSTEM SHOWED A LOCATION SHIFT ERROR. THE SYSTEM WAS POWER CYCLED SEVERAL TIMES AND THE AMPLIFIER WAS DISCONNECTED AND RECONNECTED, BUT THE ISSUE PERSISTED. THE PROCEDURE WAS CANCELLED DUE TO THESE ISSUES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11357631 ↗InjuryAug 26, 2019Communication or Transmission Problem; Communication or Transmission Problem
RELATED MANUFACTURING REF: 2184149-2019-00153. DURING A PULMONARY VEIN ISOLATION PROCEDURE, THE TACTISYS WOULD NOT CONNECT AND THE PROCEDURE WAS SIGNIFICANTLY DELAYED. AFTER VALIDATION, THE LIGHT ON THE TACTISYS WAS AMBER COLORED AND THE SYSTEM STATED, "SEARCHING FOR TACTISYS." THE LIGHT WOULD NOT TURN GREEN SO THE TACTISYS WAS RESTARTED AND THE NETWORK CABLES WERE VERIFIED WITH NO RESOLUTION. THE ETHERNET CABLE WAS REPLACED AND ENSITE WAS RESTARTED WITH NO RESOLUTION. THE CASE WAS CLOSED OUT AND RESUME STUDY AND REVALIDATION OF THE PATCHES WAS NOT POSSIBLE. THE ERROR MESSAGE "NAVX VALIDATION FAILED. THE DATA MODULE IS INTENDED FOR SINGLE USE ONLY" DISPLAYED. AFTER 30 MINUTES OF TROUBLESHOOTING, THE PROCEDURE WAS CONTINUED WITH A DIFFERENT CATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8932316 ↗
Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page