Sustain™ PM1134Recalls & reported problems
St. Jude Medical, Inc. · Pulse GeneratorSSI · Class 3 · Implant
All models of Sustain (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 05414734504744), received between Mar 11, 2019 and Jan 15, 2026 — classified as 9 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Misconnection” (2 reports).
9 other recalls cover devices cleared under the same FDA submission (P880086); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Misconnection2
- Loose or Intermittent Connection1
- No Pacing1
- Pacemaker Found in Back-Up Mode1
- Therapy Delivered to Incorrect Body Area1
- Unexpected Mode Switch1
Patient problems
- Unspecified Infection7
- Pocket Erosion5
- Discomfort3
- No Clinical Signs, Symptoms or Conditions3
- Erosion2
- No Known Impact Or Consequence To Patient2
- Wound Dehiscence1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryJan 15, 2026Therapy Delivered to Incorrect Body Area; Unexpected Mode Switch
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL AFTER EXPERIENCING INAPPROPRIATE STIMULATION IN THE POCKET AREA. UPON INTERROGATION, THE PACEMAKER WAS FOUND IN BACKUP MODE. THE PACEMAKER WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE.
Patient problems: Discomfort
MAUDE report 24082656 ↗InjuryJun 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S PACEMAKER WAS INFECTED. THE PACEMAKER WAS EXPLANTED. NO ADDITIONAL INFORMATION WAS REPORTED.
Patient problems: Unspecified Infection
MAUDE report 17131547 ↗InjuryMar 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE CLINIC DUE TO PACEMAKER POCKET INFECTION AND EROSION. IT WAS NOTED THAT THE PACEMAKER WAS ERODED AND VISIBLE THROUGH THE PACEMAKER POCKET. THE PACEMAKER WAS EXPLANTED DUE TO PACEMAKER POCKET INFECTION AND EROSION. THE PATIENT'S CONDITION WAS STABLE.
Patient problems: Pocket Erosion; Unspecified Infection
MAUDE report 16535111 ↗InjuryOct 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE HOSPITAL DUE TO PACEMAKER POCKET INFECTION. IT WAS NOTED THAT THE PACEMAKER WAS ERODED THROUGH THE PACEMAKER POCKET AND POCKET INFECTION WAS DISCOVERED. THE PACEMAKER WAS EXPLANTED DUE TO PACEMAKER POCKET INFECTION AND EROSION. THE PATIENT'S CONDITION WAS STABLE.
Patient problems: Pocket Erosion; Unspecified Infection
MAUDE report 15623601 ↗MalfunctionSep 28, 2022Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR AN IMPLANT PROCEDURE. DURING THE PROCEDURE, IT WAS NOTED THAT THE PACEMAKER SETSCREW WAS FAILING TO BE TIGHTENED. THE PACEMAKER WAS NOT USED. THE PHYSICIAN CONTINUED TO USE THE SECOND PACEMAKER AND COMPLETED THE PROCEDURE. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15500501 ↗InjurySep 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2022-38457. IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL FOR OUTPATIENT REVIEW. UPON EXAMINATION, THE PACEMAKER POUCH WAS FOUND TO BE RUPTURED AND INFECTED. THE ENTIRE PACEMAKER SYSTEM WAS EXPLANTED DUE TO INFECTION. THE PATIENT'S CONDITION WAS STABLE.
Patient problems: Unspecified Infection; Wound Dehiscence
MAUDE report 15443173 ↗InjuryAug 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2022-21284. IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A FOLLOW-UP IN CLINIC WITH THE SKIN OF THE DEVICE POCKET APPEARING SWOLLEN. THE PACEMAKER AND RIGHT VENTRICULAR LEAD WERE EXPLANTED AND REPLACED DUE TO INFECTION. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: Unspecified Infection
MAUDE report 15280808 ↗MalfunctionJan 10, 2022Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode
IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR AN INITIAL IMPLANT. DURING THE PROCEDURE, THE PHYSICIAN NOTED THE PACEMAKER WAS IN BACKUP MODE. THE PACEMAKER WAS REPLACED. THE PATIENT WAS STABLE THROUGHOUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13213856 ↗InjurySep 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NEW INFORMATION RECEIVED NOTES THAT THE PATIENT RECEIVED ANTIBIOTIC TREATMENT IN (B)(6) 2021.
Patient problems: Discomfort; Erosion; Pocket Erosion
MAUDE report 12534930 ↗InjurySep 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2021-30262. IT WAS REPORTED THAT THE PATIENT PRESENTED EXPERIENCING PAIN DUE TO A SWOLLEN WOUND. UPON INSPECTION, IT WAS NOTED THAT THE PATIENT HAD AN INFECTION AND EROSION. THE PACEMAKER WAS EXPLANTED AND A PORTION OF THE RIGHT VENTRICULAR LEAD WAS REMOVED, WHILE THE REST REMAINED IMPLANTED. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: Discomfort; Pocket Erosion; Unspecified Infection
MAUDE report 12460132 ↗MalfunctionJun 15, 2021No Pacing; No Pacing
IT WAS REPORTED THAT DURING INITIAL IMPLANT, A PATIENT'S PACEMAKER EXHIBITED NO LOW VOLTAGE OUTPUTS EVEN AFTER MULTIPLE ATTEMPTS TO PACE. THE PACEMAKER WAS NOT USED AND ANOTHER PACEMAKER WAS USED TO COMPLETED THE PROCEDURE. THE PATIENT WAS STABLE THROUGHOUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12005406 ↗InjuryJan 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2021-00144. IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO POCKET RUPTURE AND PACEMAKER EXPOSURE. AFTER EXAMINATION, IT WAS CONFIRMED THAT THE PACEMAKER POCKET WAS INFECTED AND THE POCKET WAS ULCERATED AT THE SAME TIME. THE PHYSICIAN CONSIDERED THAT THE POSSIBLE CAUSE OF INFECTION IS THAT THE PATIENT IS TOO THIN. THE SYSTEM WAS EXPLANTED ON (B)(6) 2020 AND REPLACED ON (B)(6) 2020. THE PATIENT WAS STABLE.
Patient problems: Erosion; Pocket Erosion; Unspecified Infection
MAUDE report 11113922 ↗MalfunctionApr 9, 2020Misconnection; Misconnection
IT WAS REPORTED THAT DURING PROCEDURE, THE NEW PACEMAKER WAS ATTEMPTED BE IMPLANTED HOWEVER, THE SET SCREW WAS UNABLE TO BE FIXED AND EXHIBITED A CONNECTION PROBLEM. THE DEVICE WAS REMOVED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9946471 ↗MalfunctionMar 11, 2019Misconnection; Misconnection
IT WAS REPORTED THE PATIENT PRESENTED THE DAY FOLLOWING INITIAL IMPLANT OF A RIGHT VENTRICULAR LEAD. THE LEAD WAS OBSERVED TO BE DISLODGED. DURING AN OPERATION TO REVISE THE LEAD, THE SETSCREW OF THE PACEMAKER COULD NOT BE LOOSENED. THE PACEMAKER AND LEAD WERE EXPLANTED. THE PATIENT WAS IN STABLE CONDITION FOLLOWING THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8410345 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P880086. One submission often covers many models, so check the product and lot details in each record.
- Z-1497-2022 · Jun 17, 2022 · Open, ClassifiedMerlin.net model MN5000 Software
- Z-1496-2022 · Jun 17, 2022 · Open, ClassifiedMerlin 2 PCS MER37000 programmer Model MER3400 software
- Z-1495-2022 · Jun 17, 2022 · Open, ClassifiedMerlin PCS 3650 programmer Model 3330 software
- Z-0429-2012 · Sep 23, 2011 · TerminatedImplantable Pacemaker Accent DR models PM2110 and PM2210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
- Z-0430-2012 · Sep 23, 2011 · TerminatedAnthem CRT-P models PM3110 and PM3210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
- Z-0298-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)
- Z-0296-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical APS III Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3500/3510
- Z-0297-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650
- Z-0052-04 · Jul 31, 2003 · TerminatedPulse Generator
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page