SustainRecalls & reported problems across all models
St. Jude Medical, Inc. · 4 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 189
- Deaths reported
- 1
- Injuries reported
- 88
- Recalls
- 0
We found 4 models of Sustain by St. Jude Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 189 adverse event reports in the FDA's MAUDE database, received between Jun 11, 2018 and Sep 3, 2026 — classified as 1 death, 88 injuries, 100 malfunctions.
The most frequently reported specific device problem is “Pacemaker Found in Back-Up Mode” (50 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Sustain problems
Device problems
- Pacemaker Found in Back-Up Mode50
- Premature Discharge of Battery30
- Failure to Interrogate22
- Adverse Event Without Identified Device or Use Problem21
- No Pacing17
- Unexpected Mode Switch14
- Data Problem11
- Incorrect Measurement11
Patient problems
- No Clinical Signs, Symptoms or Conditions91
- No Known Impact Or Consequence To Patient47
- Unspecified Infection19
- Arrhythmia11
- No Consequences Or Impact To Patient10
- Pocket Erosion10
- Discomfort7
- Dizziness6
Sustain models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Sustain™ PM2136Pulse generator, permanent, implantable107 reports
- Sustain™ PM2134Pulse generator, permanent, implantable1 death43 reports
- Sustain™ PM1136Pulse generator, permanent, implantable25 reports
- Sustain™ PM1134Pulse generator, permanent, implantable14 reports
Recent Sustain adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 3, 2026Difficult to Interrogate; Unexpected Mode Switch
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50595914 ↗MalfunctionAug 18, 2026Incorrect Measurement
Following a magnetic resonance imaging (MRI), a false alert for the elective replacement indicator (ERI) was observed on the device. The device was reprogrammed to resolve event. The patient was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50308280 ↗MalfunctionMay 7, 2026Computer Software Problem; Unexpected Mode Switch
DURING AN IN-CLINIC FOLLOW-UP, THE PATIENT WAS FEELING DISCOMFORT. AFTER INTERROGATION, THE DEVICE WAS FOUND TO BE IN BACK-UP MODE (BVVI) DUE TO BIT FLIP. ADDITIONALLY, DURING INTERROGATION, WHILE IN BVVI, THE MODEL NUMBER WAS SHOWING INCORRECT. THE DEVICE HAD A FIRMWARE UPDATE WHICH RESOLVED THE BVVI AND CORRECTED THE MODEL NUMBER IN THE PROGRAMMER RECORDS EVENT. THE PATIENT WAS STABLE WITH NO CONSEQUENCES.
Patient problems: Discomfort
MAUDE report 25111336 ↗MalfunctionFeb 17, 2026Pacing Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED AT CLINIC FOR A PACEMAKER REPLACEMENT PROCEDURE DUE TO PACING INHIBITION AND LOSS OF OUTPUT. THE PATIENT EXPERIENCED EPISODES OF ASYSTOLE. THE PACEMAKER WAS EXPLANTED AND SUCCESSFULLY REPLACED. THE PATIENT REMAINED STABLE THROUGHOUT THE PROCEDURE.
Patient problems: Arrhythmia
MAUDE report 24373759 ↗MalfunctionFeb 13, 2026Failure to Interrogate
IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. DEVICE EXPLANT WAS ANTICIPATED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED. THERE WERE NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24335302 ↗InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
NEW INFORMATION RECEIVED INDICATED THAT WHEN THE PATIENT PRESENTED TO THE CLINIC WITH AN UNSPECIFIED INFECTION, THE PHYSICIAN NOTED THAT SOME FUNCTIONS OF THE PACEMAKER WERE DISABLED AND THAT THE DEVICE WAS NOT IN A FULLY OPERATIONAL STATE.
Patient problems: Hematoma; Pocket Erosion; Unspecified Infection
MAUDE report 24320116 ↗InjuryJan 15, 2026Therapy Delivered to Incorrect Body Area; Unexpected Mode Switch
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL AFTER EXPERIENCING INAPPROPRIATE STIMULATION IN THE POCKET AREA. UPON INTERROGATION, THE PACEMAKER WAS FOUND IN BACKUP MODE. THE PACEMAKER WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE.
Patient problems: Discomfort
MAUDE report 24082656 ↗InjuryDec 22, 2025Failure to Interrogate; Failure to Interrogate
DURING AN IN CLINIC FOLLOW UP, THE DEVICE WAS UNABLE TO BE INTERROGATED. IT WAS NOTED, THE PATIENT HAD RECEIVED CANCER RADIATION THERAPY, IT IS UNKNOWN IF THIS WAS THE CAUSE OF IN THE INABILITY TO INTERROGATE THE DEVICE. IT WAS ALSO NOTED; THE PATIENT HAD NOT HAD A DEVICE CHECK IN A LONG TIME. THE DEVICE WAS EXPLANTED AND REPLACED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23865242 ↗InjuryDec 18, 2025Premature Discharge of Battery; Premature Discharge of Battery
DURING IN-CLINIC FOLLOW-UP, AN ELECTIVE REPLACEMENT INDICATOR (ERI) ALERT WAS RECEIVED. PREMATURE BATTERY DEPLETION IS SUSPECTED, ALTHOUGH NOT CONFIRMED. THE DEVICE WAS EXPLANTED AND REPLACED TO RESOLVE THE EVENT. THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23844016 ↗InjuryNov 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN ULCER ON THE RIGHT CHEST AND THE PACEMAKER WAS EXPOSED. THE PHYSICIAN SUSPECTED THAT THE INFECTION WAS RELATED TO THE PATIENT'S REPEATED PACEMAKER REPLACEMENTS. THE PACEMAKER WAS EXPLANTED DUE TO THE INFECTION. THE NEW PACEMAKER WAS IMPLANTED ON (B)(6) 2025. THE PATIENT WAS STABLE THROUGHOUT.
Patient problems: Pocket Erosion; Unspecified Infection
MAUDE report 23529064 ↗MalfunctionAug 22, 2025Unexpected Mode Switch; Unexpected Mode Switch
DURING AN IN-CLINIC FOLLOW-UP WHILE ATTEMPTING TO PERFORM A SOFTWARE DOWNLOAD, THE DEVICE WENT INTO BACK-UP VVI (BVVI) MODE. TECHNICAL SUPPORT WAS CONTACTED AND THE DEVICE WAS RESTORED BY A FIRMWARE DOWNLOAD TO RESOLVE THE EVENT. THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22874239 ↗MalfunctionJul 23, 2025Unexpected Mode Switch; Unexpected Mode Switch
DURING AN IN-CLINIC FOLLOW-UP, THE DEVICE WAS FOUND IN BACKUP MODE (BVVI). THE SUSPECTED CAUSE OF THE EVENT WAS DUE TO LOW BATTERY VOLTAGE. THE DEVICE WAS EXPLANTED AND REPLACED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22575539 ↗MalfunctionJun 22, 2025Unexpected Mode Switch; Unexpected Mode Switch
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL FOR A SCHEDULED DEVICE EXCHANGE PROCEDURE. DURING THE PROCEDURE, IT WAS NOTED THAT THE PULSE GENERATOR HAD ENTERED BACKUP VVI MODE SHORTLY BEFORE BEING REMOVED FROM THE POCKET. THE PULSE GENERATOR WAS EXPLANTED AND REPLACED. THE PATIENT HAD NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22304783 ↗MalfunctionMay 28, 2025Loose or Intermittent Connection; Failure to Disconnect; Loose or Intermittent Connection; Failure to Disconnect
DURING A ROUTINE DEVICE EXCHANGE PROCEDURE, THE ATRIAL SET SCREW COULD NOT BE LOOSENED AND THE SET SCREW APPEARED TO BE STRIPPED. ADDITIONALLY, THE HEX WRENCH WOULD NOT FIT INTO THE SET SCREW. THE ATRIAL LEAD WAS CUT AND THE DEVICE WAS EXPLANTED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22102028 ↗MalfunctionMay 14, 2025Unexpected Mode Switch; Unexpected Mode Switch
IT WAS REPORTED THAT THE DEVICE WAS FOUND IN BACKUP MODE. ABBOTT TECHNICAL SUPPORT WAS CONTACTED AND RECOMMENDED REPROGRAMMING, WHICH WAS PERFORMED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE AND THERE WERE NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22023388 ↗
Other St. Jude Medical, Inc. products
- Tendril15 models · 49,498 reports
- Octrode10 models · 36,740 reports
- Proclaim12 models · 35,038 reports · 17 recalls
- Durata28 models · 19,805 reports
- Assurity MRI2 models · 17,838 reports · 1 recall
- Gallant8 models · 12,885 reports · 5 recalls
- Quartet10 models · 11,001 reports
- MitraClip19 models · 9,056 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page