CardioMEMS™ CM1000Recalls & reported problems
Cardiomems, Inc. · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734509169), received between Jun 20, 2018 and Mar 8, 2023 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Feb 7, 2023Class IIOpen, Classified
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
Z-1218-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Displays Incorrect Message4
- Material Frayed2
- Smoking1
- Thermal Decomposition of Device1
Patient problems
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 8, 2023Material Frayed; Material Frayed
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING EXPOSED WIRES OF THE POWER SUPPLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16504232 ↗MalfunctionMar 8, 2021Thermal Decomposition of Device; Thermal Decomposition of Device
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING THERMAL DAMAGE TO THE PRINTED CIRCUIT BOARD OF THE CARDIOMEMS PATIENT ELECTRONIC UNIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11441003 ↗MalfunctionApr 17, 2019Material Frayed; Material Frayed
IT WAS REPORTED THAT THE PATIENT'S UNIT HAD SMOKE LIKE SMELL. THE PATIENT WAS NOT HARMED AND ISSUE WAS RESOLVED BY REPLACING THE UNIT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8523217 ↗MalfunctionDec 10, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT ELECTRONIC UNIT RECEIVED AN ERROR 5 MESSAGE, WHICH INDICATES AN ISSUE RELATED TO TEMPERATURE AND BAROMETRIC PRESSURE. A REPLACEMENT UNIT MAY RESOLVE THE ISSUE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8147623 ↗MalfunctionJul 12, 2018Smoking; Smoking
IT WAS REPORTED THAT THE PATIENT'S ELECTRONIC UNIT WOULD NOT TURN ON AFTER THE PATIENT HEARD A POPPING SOUND AND SMELLED SMOKE COMING FROM IT. THE PATIENT WAS NOT INJURED WHEN THE POPPING OCCURRED. A REPLACEMENT UNIT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7681934 ↗MalfunctionJun 21, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT ELECTRONIC UNIT RECEIVED AN ERROR 5 MESSAGE, WHICH INDICATES AN ISSUE RELATED TO TEMPERATURE AND BAROMETRIC PRESSURE. A REPLACEMENT UNIT MAY RESOLVE THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7627765 ↗MalfunctionJun 20, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT ELECTRONIC UNIT RECEIVED AN ERROR 5 MESSAGE, WHICH INDICATES AN ISSUE RELATED TO TEMPERATURE AND BAROMETRIC PRESSURE. A REPLACEMENT UNIT MAY RESOLVE THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7623912 ↗MalfunctionJun 20, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT ELECTRONIC UNIT RECEIVED AN ERROR 5 MESSAGE, WHICH INDICATES AN ISSUE RELATED TO TEMPERATURE AND BAROMETRIC PRESSURE. A REPLACEMENT UNIT MAY RESOLVE THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7620350 ↗
Other System, hemodynamic, implantable devices
- CardioMEMS™ CM1100Cardiomems, Inc. · System, hemodynamic, implantable3 recalls4,257 reports
- CardioMEMS™ CM2000Cardiomems, Inc. · System, hemodynamic, implantable1 recall170 deaths4,294 reports
- CardioMEMS™ CM3100Cardiomems, Inc. · System, hemodynamic, implantable1 recall419 reports
- CardioMEMS™ CM3000Cardiomems, Inc. · System, hemodynamic, implantable3 recalls65 reports
- CardioMEMS™ CM1010Cardiomems, Inc. · System, hemodynamic, implantable1 recall1 report
- Cordella Sensor Delivery Kit 100303-05Endotronix, Inc. · System, Hemodynamic, Implantable1 death47 reports
- CardioMEMS™ CM3000Cardiomems, Inc. · System, hemodynamic, implantable5 reports
- CardioMEMS™ CM1110St. Jude Medical, Inc. · System, hemodynamic, implantable1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page