Skip to content

CardioMEMS™ CM1010Recalls & reported problems

Cardiomems, Inc. · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05414734509176), received between Oct 24, 2018 and Oct 24, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Sparking” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 7, 2023Class IIOpen, Classified

    Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).

    Z-1219-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Sparking1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2018Sparking; Sparking

    IT WAS REPORTED THE PATIENT OBSERVED SPARKS COMING OUT AND A BURNING SMELL WHILE ATTEMPTING TO POWER ON THE UNIT. THE UNIT WAS REPLACED WHICH RESOLVED THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8001928 ↗

Other System, hemodynamic, implantable devices

All System, hemodynamic, implantable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page