CardioMEMS™ CM1110Recalls & reported problems

St. Jude Medical, Inc. · Power Supply · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05414734509916), received between Apr 26, 2019 and Apr 26, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Frayed” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Frayed1
  • Sparking1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 26, 2019Material Frayed; Sparking; Material Frayed; Sparking

    IT WAS REPORTED ALLEGED SPARKS CAME FROM THE POWER CORD OF THE PATIENT'S EXTERNAL DEVICE UNIT WHILE DISCONNECTED FROM THE PATIENT'S DEVICE. FURTHER EXAMINATION OF THE POWER CORD REVEALED THE CAUSE WAS BELIEVED TO BE CHEWING FOR A PET. A REPLACEMENT DEVICE WAS SENT TO THE PATIENT AND ADDRESSED THE ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8555878 ↗

Other System, hemodynamic, implantable devices

All System, hemodynamic, implantable devices →