EP-WorkMate™ SystemRecalls & reported problems
St. Jude Medical, Inc. · 120 Channel SCU · Class 2
All models of EP-WorkMate (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05415067003119), received between Aug 8, 2018 and Apr 5, 2021 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Communication or Transmission Problem1
- Failure to Analyze Signal1
- Failure to Power Up1
- Smoking1
- Sparking1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2021Failure to Analyze Signal; Failure to Analyze Signal
DURING AN ATRIOVENTRICULAR NODAL RE-ENTRANT TACHYCARDIA (AVNRT) PROCEDURE, INTRACARDIAC SIGNALS WERE NOT VISIBLE WHEN THE AMPLIFIER WAS CONNECTED AND THE CASE WAS CANCELLED. THE FEMORAL VEIN WAS PUNCTURED AND CATHETERS WERE INSERTED AND NO INTRACARDIAC ELECTROGRAMS WERE NOTED. THE SYSTEM WAS RESTARTED 3 TIMES WHICH DID NOT RESOLVE THE ISSUE AND THE PROCEDURE WAS POSTPONED. ANOTHER SYSTEM WAS USED TO COMPLETE THE CASE THE NEXT DAY WITH NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11614835 ↗InjurySep 9, 2020Failure to Power Up; Failure to Power Up
DURING THE PROCEDURE, THE AMPLIFIER WOULD NOT BOOT UP AND ONE SINGLE CONTINUOUS BEEP WAS HEARD AND THE PROCEDURE WAS CANCELLED. THE AMPLIFIER WAS REBOOTED SEVERAL TIMES, BUT THE ISSUE REMAINED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE CANCELLATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10508714 ↗InjuryAug 30, 2019Communication or Transmission Problem; Communication or Transmission Problem
DURING A SUPRAVENTRICULAR TACHYCARDIA ABLATION PROCEDURE A CANCELLATION OCCURRED. FOLLOWING PATIENT PREP, THE AMPLIFIER WAS POWERED ON AND INTERMITTENT CONNECTION OCCURRED. THE AMPLIFIER AND MEDIA CONVERTER WERE POWER CYCLED WITHOUT RESOLUTION. THE FIBER OPTIC AND ETHERNET CABLES WERE REPLACED AND THE ISSUE STILL DID NOT IMPROVE. THE PROCEDURE WAS CANCELLED WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8950654 ↗MalfunctionNov 2, 2018Smoking; Smoking
WHEN THE AMPLIFIER WAS POWERED ON SMOKE WAS EMITTED. THERE WERE NO PATIENT OR USER ADVERSE CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8033794 ↗MalfunctionAug 8, 2018Sparking; Sparking
PRIOR TO AN ELECTROPHYSIOLOGY PROCEDURE, THE AMPLIFIER DISPLAYED THE MESSAGE ¿NO DISPLAY - PLEASE CHECK CONNECTION¿ AND A SPARK WAS OBSERVED FOLLOWED BY A BURNING SMELL. THE AMPLIFIER WAS POWER CYCLED AND A LOUD CONTINUOUS BEEPING SOUND CAME FROM THE AMPLIFIER. THE PATIENT WAS MOVED TO ANOTHER ROOM AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7761424 ↗
Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page