Skip to content

EnSite™ ENSITE-R-DWS-1.1Recalls & reported problems

St. Jude Medical, Inc. · X EP System Display Workstation · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05415067037756), received between Apr 8, 2026 and Apr 8, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Freezes or Fails to Launch1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2026Application Program Freezes or Fails to Launch

    DURING AN ATRIAL FIBRILLATION PROCEDURE, THERE WAS A PROCEDURAL DELAY. IN VOXEL MODE, ON ENSITE X 3.1.1 SOFTWARE, IT WAS NOTED THAT THE SYSTEM SUDDENLY BECAME CHOPPY AND ABNORMAL WHEN THE STUDY WAS ENTERED AND ANY OPERATION WAS ATTEMPTED TO BE PERFORMED. THE SURFACE AND INTRACARDIAC ELECTROGRAMS DISPLAYED AS A MAGENTA-COLORED BASELINE, AND THE SYSTEM FROZE. THE CONNECTIONS WERE CHECKED AND CONFIRMED TO BE CORRECT. A FORCED LOGOUT OF THE SYSTEM WAS PERFORMED USING CTRL + ALT + DELETE AND THE SYSTEM WAS REBOOTED, BUT THE ISSUE PERSISTED EVEN AFTER FIVE TO TEN REBOOTS. WHEN SHUTTING DOWN THE SYSTEM, AN ERROR STATING "HARDWARE ERROR" DISPLAYED JUST BEFORE THE SYSTEM POWERED OFF. THE ISSUE INTERFERED WITH LOADING PRESETS, ADDING CATHETERS, AND PERFORMING SET METAL AND INITIATING MAPPING. WHEN THE ADVISOR HD GRID X AND TACTIFLEX CATHETERS WERE AUTOMATICALLY RECOGNIZED, THE SYSTEM WAS ABLE TO MAINTAIN STABLE PERFORMANCE, AND MINIMAL NAVIGATION WAS POSSIBLE IN NAVX MODE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY PERFORMING ONLY PVI IN NAVX MODE UNDER FLUOROSCOPIC AND ICE GUIDANCE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817227 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →

Sources and updates