EnSite™ ENSITE-SW-2.0.1Recalls & reported problems

St. Jude Medical, Inc. · X EP System Software Upgrade · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067046703), received between Nov 23, 2025 and Feb 4, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2026Display or Visual Feedback Problem

    DURING A TYPICAL ATRIAL FLUTTER PROCEDURE, THERE WAS A DELAY DUE TO A DISPLAY ISSUE. USING THE ENSITE X V2.0.1 IN VOXEL MODE, THE ADVISOR HD GRID CATHETER WAS INSERTED IN THE RIGHT ATRIUM. THE CATHETER HAD APPEARED INSIDE THE SHEATH IN THE 3D SYSTEM BUT APPEARED OUTSIDE THE SHEATH ON THE DSA IMAGES. TROUBLESHOOTING INCLUDED REINSERTING THE TAIL WIRE, RESTARTING THE ENSITE X DWS, AND REPLACING THE CATHETER, BUT THE ISSUE PERSISTED. THE ISSUE WAS LATER RESOLVED BY SWITCHING THE VOXEL MODE TO NAVX MODE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257454 ↗
  • MalfunctionNov 23, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    DURING AN ATRIAL FIBRILLATION PROCEDURE, THERE WAS A SHEATH FILTER AND COMMUNICATION ISSUES WITH THE ENSITE X SOFTWARE, AMPLIFIER, AND DISPLAY WORKSTATION RESULTING IN A DELAY. THE ENSITE X AMPLIFIER WAS FLASHING RED AND THE CATHETER WAS UNABLE TO BE CONNECTED. IT ALSO KEPT SHOWING THAT THE CATHETER WAS IN THE SHEATH AND NO CATHETER DATA COULD BE DISPLAYED. ADDITIONALLY, DURING THE PROCEDURE THERE WAS A CONTACT FORCE ERROR WITH LOSS OF THE PRESSURE VALUE. DURING THIS THE MAGNETIC POINT WOULD DISCONNECT. AFTER REPLACING THE ENSITE X DISPLAY WORKSTATION THE CONTACT FORCE ISSUE RESOLVED. TROUBLESHOOTING ALSO INCLUDED REPLACING THE CATHETER AND RESTARTING THE ENSITE X DISPLAY WORKSTATION AT LEAST 3 TIMES, BUT THE SHEATH FILTER ISSUE PERSISTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23627065 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →