EnSite™ ENSITE-DWS-2.0.1Recalls & reported problems

St. Jude Medical, Inc. · X EP System Display Workstation · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05415067046710), received between Jan 19, 2024 and Dec 5, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem4
  • Application Program Freezes or Fails to Launch1
  • Application Program Problem1
  • No Apparent Adverse Event1
  • Power Problem1
  • Signal Artifact/Noise1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 5, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    DURING THE RADIOFREQUENCY ATRIAL FIBRILLATION PROCEDURE IN VOXEL MODE, THERE WAS A DELAY DUE TO THE ENSITE DWS LAGGING AND CRASHING. WHILE MAPPING, THE ENSITE X DWS LAGGED AND CRASHED. THE DWS WAS RESTARTED THREE TIMES (EACH RESTART TOOK 15-20 MINUTES), BUT THE LAGGING ISSUE PERSISTED. THE DWS WAS REPLACED, AND THE ISSUE WAS RESOLVED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23735272 ↗
  • MalfunctionNov 23, 2025Communication or Transmission Problem; Communication or Transmission Problem

    DURING AN ATRIAL FIBRILLATION PROCEDURE, THERE WAS A SHEATH FILTER AND COMMUNICATION ISSUES WITH THE ENSITE X SOFTWARE, AMPLIFIER, AND DISPLAY WORKSTATION RESULTING IN A DELAY. THE ENSITE X AMPLIFIER WAS FLASHING RED AND THE CATHETER WAS UNABLE TO BE CONNECTED. IT ALSO KEPT SHOWING THAT THE CATHETER WAS IN THE SHEATH AND NO CATHETER DATA COULD BE DISPLAYED. ADDITIONALLY, DURING THE PROCEDURE THERE WAS A CONTACT FORCE ERROR WITH LOSS OF THE PRESSURE VALUE. DURING THIS THE MAGNETIC POINT WOULD DISCONNECT. AFTER REPLACING THE ENSITE X DISPLAY WORKSTATION THE CONTACT FORCE ISSUE RESOLVED. TROUBLESHOOTING ALSO INCLUDED REPLACING THE CATHETER AND RESTARTING THE ENSITE X DISPLAY WORKSTATION AT LEAST 3 TIMES, BUT THE SHEATH FILTER ISSUE PERSISTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23627067 ↗
  • MalfunctionJun 4, 2025Communication or Transmission Problem; Communication or Transmission Problem

    DURING AN ATRIAL FIBRILLATION PROCEDURE, COMMUNICATION ISSUES RESULTED IN A PROCEDURAL DELAY. DIFFICULTY WAS NOTED GETTING COMMUNICATION BETWEEN THE AMPLIFIER AND THE ENSITE X DWS. WHEN THE ENSITE X DWS WAS REBOOTED, IT TOOK A LONGER TIME THAN NORMAL. ONCE BOOTED, AN ATTEMPT WAS MADE TO LOGIN AND THE ENSITE X DWS GAVE A "LOGIN SUCCESSFUL" MESSAGE, BUT DID NOT PROCEED TO THE HOME SCREEN AND INSTEAD STAYED ON THE LOGIN SCREEN BEFORE THE LOGIN PROMPT CAME BACK. THE SYSTEM WAS REBOOTED A TOTAL OF FIVE OR SIX TIMES BEFORE THE ISSUE WAS RESOLVED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22141543 ↗
  • MalfunctionMay 27, 2025Application Program Problem; Application Program Problem

    DURING AN ATYPICAL ATRIAL FLUTTER PROCEDURE, FUNCTIONAL ISSUES WITH THE DWS RESULTED IN THE CANCELLATION OF THE PROCEDURE. THE ENSITE X SYSTEM SUDDENLY GOT REALLY SLOW, NOT ANSWERING PROPERLY TO THE CATHETER MOVEMENTS, TAKING LONGER TO RESPOND TO COMMANDS WHEN CLICKING ON BUTTONS, OPENING MENUS, ETC. THE SYSTEM WAS REBOOTED (AMPLIFIER AND DWS), CABLES WERE UNPLUGGED AND PLUGGED, THE CASE WAS CREATED FOR THE SAME PATIENT BUT THE ISSUE REMAINED. THE DWS HARDRIVE HAS ONLY 4 CASES IN IT. THE ISSUE WAS NOT RESOLVED AND THE CASE WAS CANCELLED WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. THE PROCEDURE WILL BE RESCHEDULED FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22090029 ↗
  • MalfunctionMay 22, 2025Signal Artifact/Noise; No Apparent Adverse Event; Signal Artifact/Noise; No Apparent Adverse Event

    DUPLICATE OF (B)(4). DURING AN ATRIAL FIBRILLATION PROCEDURE, THERE WAS NOISE DUE TO THE DISPLAY WORKSTATION RESULTING IN CANCELLATION OF THE PROCEDURE. WHEN PREPARING TO START MAPPING WITH THE MAPPING CATHETER, NOISE WAS SEEN IN SOME OF THE EGMS AND THEN ALL THE SIGNALS GOT NOISY WITH INTERFERENCE AND AT TIMES THE EGMS TURNED PURPLE WITH A FLAT LINE. AFTER UNPLUGGING THE MAPPING CATHETER FROM THE AMPLIFIER, THE OTHER CORONARY SINUS SIGNALS RECOVERED. THE MAPPING CATHETER CABLE WAS CHECKED, DISCONNECTED AND CONNECTED AGAIN, AND REPLACED WITH A NEW CABLE BUT THE ISSUE REMAINED. THE AMPLIFIER AND DISPLAY WORKSTATION WERE THEN REBOOTED, AND A NEW CASE WAS STARTED, AND THE MAPPING CATHETER WAS EXCHANGED BUT THE ISSUE REMAINED. ALL THE TROUBLESHOOTING WAS REPEATED WITHOUT SUCCESS. A THIRD CATHETER, THIS TIME AN ADVISOR HD GRID RATHER THAN AN ADVISOR HD GRID X WAS USED. HOWEVER, THE ISSUE STILL DID NOT RESOLVE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22073367 ↗
  • MalfunctionJan 23, 2025Communication or Transmission Problem; Communication or Transmission Problem

    DURING AN ATRIAL TACHYCARDIA PROCEDURE, COMMUNICATION ISSUES RESULTED IN A PROCEDURAL DELAY. WHILE WORKING IN VOXEL MODE, THE ABLATION CATHETER WAS NO LONGER DISPLAYED. ALL THE CABLES WERE CHECK AND RECONNECTED. THE CATHETER WAS REPLACED WITH NO RESOLUTION. THE DWS WAS RESTARTED AND THE STUDY WAS RESUMED WHICH RESOLVED THE ISSUE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21217400 ↗
  • MalfunctionDec 5, 2024Power Problem; Power Problem

    BEFORE AN ATRIAL FIBRILLATION PROCEDURE, WHILE PATIENT WAS PREPPED, IT WAS NOT POSSIBLE TO TURN ON THE ENSITE X DISPLAY WORKSTATION RESULTING IN A DELAY. THE DWS WAS RESTARTED 3 TIMES, BUT IT WOULD NOT POWER UP. A MESSAGE "THE SYSTEM TIME IS INVALID. THIS MAY BE A RESULT OF A LOSS IN BATTERY POWER" APPEARED. THE DWS WAS EXCHANGED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20854505 ↗
  • MalfunctionJan 19, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS CONFIRMED THAT THIS ISSUE WAS RESOLVED BY REPLACING THE PRS SENSOR. THE DWS IS STILL IN USE IN THE FIELD WITHOUT REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18543083 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →