Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067047854), received between Oct 25, 2023 and Sep 16, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Application Program Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Problem1
- Display or Visual Feedback Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2024Application Program Problem; Application Program Problem
RELATED MANUFACTURING REFERENCE: 3008452825-2024-00522. DURING THE SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, COMMUNICATION ISSUES RESULTED IN A PROCEDURAL DELAY. WHILE USING ENSITE X V3.0.1 IN VOXEL MODE, A TACTICATH SE WAS CONNECTED TO THE ENSITE X AMPLIFIER WITH AN SE CABLE TO SE1 OR SE2 ON THE AMPLIFIER, AND THE SYSTEM IMMEDIATELY DISPLAYED, "UNEXPECTED ERROR" AND THE STUDY WAS AUTOMATICALLY CLOSED. THE AMPLIFIER, DWS, AND TACTISYS WERE ALL RESTARTED BUT THE ISSUE WAS NOT RESOLVED. THE CATHETER WAS REPLACED BUT THE ISSUE PERSISTED. THE PROCEDURE WAS COMPLETED USING THE SECOND CATHETER IN NAVX MODE AND WITHOUT CONNECTING THE SE CABLE OF THE CATHETER TO THE AMPLIFIER, WITHOUT MAGNETIC NAVIGATION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20229986 ↗MalfunctionOct 25, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
DURING AN ATRIAL FIBRILLATION PROCEDURE, A SHIFT OCCURRED RESULTING IN CANCELLATION OF THE PROCEDURE. THERE WAS A SHIFT IN ENSITEX IN VOXEL MODE; ALIGNMENT COULD NOT BE DONE AND COULD NOT COMPENSATE THE SHIFT. PRS-A AND PRS-P SENSORS WERE CORRECTLY PLACED AND RESPIRATION WAS DONE MULTIPLE TIMES WHICH DID NOT RESOLVE THE ISSUE AND THE PROCEDURE WAS CANCELLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18003158 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page