EnSite™ ENSITE-SW-3.1Recalls & reported problems
St. Jude Medical, Inc. · X EP System Software Upgrade · Class 2
All models of EnSite (37)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05415067049704), received between Nov 26, 2024 and Sep 9, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Application Program Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Problem4
- Display or Visual Feedback Problem2
- Electromagnetic Interference1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2025Display or Visual Feedback Problem; Application Program Problem; Display or Visual Feedback Problem; Application Program Problem
DURING A VENTRICULAR TACHYCARDIA PROCEDURE, DISPLAY ISSUES WITH THE ENSITE X SOFTWARE RESULTED IN CANCELLATION OF THE PROCEDURE. WHEN USING THE ENSITE X SYSTEM IN VOXEL MODE AND MAPPING WAS DONE USING THE ADVISOR HD GRID CATHETER, EVERYTHING WAS NORMAL. HOWEVER, WHEN ARRIVING AT THE ANTERIOR APICAL PART OF THE LEFT VENTRICLE, THE CATHETER HAD DISAPPEARED. EVERYTHING ELSE REMAINED THE SAME AND THE GREEN LIGHT WAS ON BUT, ONLY IN THAT PARTICULAR SPOT DID THE CATHETER DISAPPEAR. SWITCHING TO NAVX MODE CAUSED THE CATHETER TO REAPPEAR BUT, AFTER CARDIOVERSION, WHEN THE MODE WAS SWITCHED BACK TO VOXEL, THE DISTORTION METER WAS FROZEN IN RED. THE MODE WAS SWITCHED OFF VOXEL AND THE CATHETER WAS REINSERTED INTO THE PATIENT WHICH RESOLVED THE DISTORTION METER ISSUE, BUT THE CATHETER DISAPPEARED. CHANGING THE CATHETER DID NOT RESOLVE THE ISSUE. AFTER TRYING TO MAP FOR A WHILE LONGER, THE DECISION TO END THE PROCEDURE WAS MADE DUE TO THE ISSUE PERSISTING. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23012347 ↗MalfunctionJun 16, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
DURING A SUPRAVENTRICULAR TACHYCARDIA VOXEL PROCEDURE, DISPLAY ISSUES RESULTED IN A PROCEDURAL DELAY. THE MAP SHIFTED NOTICEABLY WITHOUT PATIENT MOVEMENT. THE CASE WAS RESTARTED TWICE AND REMAPPED, AND THE METAL BASELINE WAS RESET. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22211849 ↗MalfunctionMay 5, 2025Application Program Problem; Application Program Problem
DURING AN ATRIAL TACHYCARDIA PROCEDURE, ISSUES WITH THE SOFTWARE RESULTED IN A PROCEDURAL DELAY. WHEN USING THE ENSITE X V3.1.1 IN VOXEL MODE, MAPPING WAS PERFORMED WITHOUT ANY ISSUES WHEN USING THE ADVISOR HD GRID CATHETER. ONCE THE ADVISOR VL CATHETER WAS CONNECTED TO CONTINUE MAPPING, THE STUDY CRASHED WITH THE ERROR MESSAGE: "UNEXPECTED ERROR OCCURRED. APPLICATION EXITED". TROUBLESHOOTING INCLUDED REPLACING THE DIRECT CABLE, REPLACING THE CATHETER, USING A DIFFERENT PORT, TRYING "RESUME STUDY", AND RESTARTING THE DWS AND THE AMPLIFIER BUT ISSUE PERSISTED. THE ISSUE WAS RESOLVED AFTER SWITCHING FROM VOXEL MODE TO NAVX MODE AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21960518 ↗MalfunctionJan 17, 2025Electromagnetic Interference; Application Program Problem; Electromagnetic Interference; Application Program Problem
DURING THE PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, ISSUES WITH THE SOFTWARE RESULTED IN A PROCEDURAL DELAY. THE ELECTRODE KIT WAS APPLIED TO THE PATIENT, THE ABLATION CATHETER AND THE MAPPING CATHETER WERE INSERTED INTO THE CARDIAC CAVITY AND CONNECTED TO ENSITE. IT WAS NOTED THAT DISTORTION METER ON THE DWS EXCEEDED THE LIMIT (II), TURNED RED, AND FROZE. THE NAVIGATION DISPLAYS OF THE CATHETERS ALSO FROZE. NO ERROR MESSAGE WAS DISPLAYED. THE SHEATH WAS MOVED, THE FIELD FRAME WAS ADJUSTED, THE LOCATION OF THE PRS-A WAS CHECKED WITH A TOUCH, EACH PRS CABLE WAS RECONNECTED, THE CABLES OF BOTH CATHETERS WERE RECONNECTED, THE ENSITE WAS RESTARTED, AND THE ABLATION CATHETER WAS REPLACED. THE ENSITE WAS RESTARTED AGAIN AND THE DISTORTION METER AND NAVIGATION DISPLAY FREEZE WERE RESOLVED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21179889 ↗MalfunctionNov 26, 2024Application Program Problem; Application Program Problem
DURING A PREMATURE VENTRICULAR CONTRACTION PROCEDURE, THERE WAS A CRASH ON THE DISPLAY WORKSTATION RESULTING IN A 20 MINUTE DELAY. WHEN CONNECTING A NEW TACTIFLEX CATHETER TO THE AMPLIFIER. THE DISPLAY WORKSTATION SHOWED AN UNEXPECTED ERROR, AND THE STUDY CLOSED ITSELF. THE STUDY COULD NOT BE RESTARTED UNTIL THE TACTIFLEX SE WAS DISCONNECTED. THIS WAS THE FIRST TIME USING THE DISPLAY WORKSTATION AFTER PERFORMING THE UPDATE TO V3.1. THE PROCEDURE WAS COMPLETED IN NAVX MODE WITH A NON-SE CATHETER WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20786806 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page