Skip to content

EnSite™ ENSITE-TFM-3.1Recalls & reported problems

St. Jude Medical, Inc. · X EP System TactiFlex Ablation Catheter, Sensor Enabled Software Upgrade · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05415067049711), received between Feb 6, 2025 and Feb 6, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Component Missing” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025Component Missing; Component Missing

    DURING AN ABLATION PROCEDURE, IT WAS FOUND THAT THE TACTIFLEX SOFTWARE HAD NOT BEEN INSTALLED RESULTING IN CANCELLATION OF THE PROCEDURE. AFTER CONNECTING THE TACTIFLEX SE, AN ERROR APPEARED THAT THE CATHETER HAD BEEN REUSED OR USED ON ANOTHER PATIENT. THE CATHETER WAS RECONNECTED, A NEW STUDY WAS STARTED, THE SYSTEM WAS REBOOTED, AND THE CATHETER WAS EXCHANGED WITH NO RESOLVE. CUSTOMER SERVICE WAS CALLED, AND THE SOFTWARE WAS CHECKED. IT WAS DISCOVERED THAT THE TACTIFLEX SOFTWARE WAS MISSING. THE PROCEDURE WAS ABORTED AT THIS TIME BECAUSE THE SOFTWARE WAS NOT AVAILABLE AND THERE WAS DIFFICULT PATIENT ANATOMY. ONCE THE DISK WAS OBTAINED, AND THE SOFTWARE WAS INSTALLED THE ISSUE RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21322846 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page