St. Jude Medical, Inc. · HF Cardiac Resynchronization Therapy Defibrillator VVED DDDRV · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05415067050120), received between Jan 7, 2026 and May 18, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing3
  • Loose or Intermittent Connection2
  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Capture1
  • High impedance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Pocket Erosion1
  • Unspecified Infection1
  • Wound Dehiscence1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Over-Sensing

    IT WAS REPORTED THAT PATIENT PRESENTED FOR IN CLINIC FOLLOW-UP. DEVICE INTERROGATION REVEALED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED INTERMITTENT CROSSTALK RESULTING IN PACING INHIBITION. PROGRAMMING CHANGES WERE PERFORMED TO RESOLVE THE ISSUE. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25204448 ↗
  • MalfunctionApr 14, 2026Over-Sensing

    IT WAS REPORTED POST IMPLANT PROCEDURE, EPISODES OF CROSSTALK OVERSENSING WERE OBSERVED ON THE DEVICE. TECHNICAL SUPPORT WAS CONTACTED AND RECOMMENDED REPROGRAMMING TO RESOLVE THE EVENT. REPROGRAMMING HAS BEEN PERFORMED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877770 ↗
  • MalfunctionApr 8, 2026Over-Sensing

    DURING A REMOTE FOLLOW-UP, EPISODES OF POST-PACED T-WAVE OVERSENSING (PPTWOS) WERE OBSERVED ON THE DEVICE. TECHNICAL SUPPORT WAS CONTACTED AND RECOMMENDED REPROGRAMMING TO RESOLVE THE EVENT. NO INTERVENTION HAS BEEN PERFORMED AT THIS TIME. THE PATIENT WAS STABLE WITH NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816067 ↗
  • MalfunctionMar 2, 2026Loose or Intermittent Connection

    DURING THE INITIAL IMPLANT PROCEDURE, WHILE ATTEMPTING TO TIGHTEN THE SET SCREW TO SECURE THE LEAD, THE SET SCREW WAS STRIPPED AND WAS NO LONGER ABLE TO ROTATE. A NEW DEVICE WAS SELECTED AND SUCCESSFULLY IMPLANTED TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24480220 ↗
  • MalfunctionFeb 24, 2026Loose or Intermittent Connection

    DURING THE INITIAL IMPLANT PROCEDURE, THE SET SCREW WAS UNABLE TO BE TIGHTENED ON THE DEVICE. THE DEVICE WAS EXPLANTED AND REPLACED TO RESOLVE EVENT. THE PATIENT WAS STABLE WITH NO CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24438710 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM WITH WOUND DEHISCENCE AND A SUSPECTED INFECTION. THE DEVICE AND LEADS WERE ERODED. TO RESOLVE THE ISSUE, THE PATIENT UNDERWENT A SURGICAL PROCEDURE ON (B)(6) 2026 TO REMOVE THE ENTIRE SYSTEM, INCLUDING THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), THE RIGHT ATRIAL (RA) LEAD, AND THE RIGHT VENTRICULAR (RV) LEAD. THE PATIENT REMAINED IN STABLE CONDITION.

    Patient problems: Pocket Erosion; Unspecified Infection; Wound Dehiscence

    MAUDE report 24381934 ↗
  • MalfunctionFeb 9, 2026Failure to Capture

    IT WAS REPORTED THAT A LOSS OF CAPTURE WAS OBSERVED ON THE DEVICE. IT IS UNKNOWN IF THE PATIENT IS PACEMAKER DEPENDENT. TECHNICAL SUPPORT WAS CONTACTED AND THE DEVICE WAS REPROGRAMMED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24297606 ↗
  • MalfunctionJan 7, 2026High impedance

    DURING THE INITIAL IMPLANT PROCEDURE, THE PACING IMPEDANCE INCREASED WHEN THE LEAD WAS CONNECTED TO THE HEADER OF THE DEVICE. A NEW DEVICE WAS SELECTED AND CONNECTED TO THE LEAD TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23987114 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →