Skip to content

EntrantRecalls & reported problems across all models

St. Jude Medical, Inc. · 8 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one Entrant model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
658
Deaths reported
7
Injuries reported
112
Recalls
2

We found 8 models of Entrant by St. Jude Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 658 adverse event reports in the FDA's MAUDE database, received between Dec 3, 2020 and Sep 30, 2026 — classified as 7 deaths, 112 injuries, 539 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (287 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 12 reportsDec 2023: 12 reportsJan 2024: 8 reportsFeb 2024: 10 reportsMar 2024: 6 reportsApr 2024: 14 reportsMay 2024: 12 reportsJun 2024: 17 reportsJul 2024: 12 reportsAug 2024: 6 reportsSep 2024: 11 reportsOct 2024: 9 reportsNov 2024: 18 reportsDec 2024: 10 reportsJan 2025: 15 reportsFeb 2025: 11 reportsMar 2025: 12 reportsApr 2025: 4 reportsMay 2025: 16 reportsJun 2025: 19 reportsJul 2025: 14 reportsAug 2025: 15 reportsSep 2025: 8 reportsOct 2025: 14 reportsNov 2025: 15 reportsDec 2025: 18 reportsJan 2026: 12 reportsFeb 2026: 29 reportsMar 2026: 19 reportsApr 2026: 25 reportsMay 2026: 20 reportsJun 2026: 13 reportsJul 2026: 4 reportsAug 2026: 14 reportsSep 2026: 22 reportsOct 2026: 0 reports20242025202629
Last 36 months. The most recent month may be incomplete.

Entrant recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 16, 2023Open, Classified

    Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

    Z-2615-2023 · Root cause: Component design/selection · Recall details · FDA record ↗

  2. Aug 16, 2023Open, Classified

    Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

    Z-2614-2023 · Root cause: Component design/selection · names 2 of these models · Recall details · FDA record ↗

Most reported Entrant problems

Device problems

  • Over-Sensing287
  • Inappropriate or Unexpected Reset99
  • Environmental Compatibility Problem64
  • Loose or Intermittent Connection52
  • Failure to Interrogate49
  • Inappropriate/Inadequate Shock/Stimulation46
  • High impedance31
  • Incorrect Interpretation of Signal29

Patient problems

  • No Clinical Signs, Symptoms or Conditions593
  • Arrhythmia25
  • Unspecified Infection19
  • Insufficient Information17
  • Discomfort11
  • Pocket Erosion5
  • Hematoma4
  • No Consequences Or Impact To Patient3

Entrant models

8 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Entrant adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other St. Jude Medical, Inc. products

All St. Jude Medical, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page