EnSite™ ENSITE-DWS-04Recalls & reported problems

St. Jude Medical, Inc. · X EP System Display Workstation · Class 2

All models of EnSite (37)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05415067051257), received between Feb 6, 2026 and Aug 17, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Freezes or Fails to Launch5
  • Communication or Transmission Problem2
  • Application Program Problem1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Output Problem

    During an atrial fibrillation procedure, after patches were placed on the patient, an attempt was made to create a new study using an existing patient record. At that time, the system displayed an "unexpected error" message, and the study could not be created. The EnSite X DWS was exchanged with a replacement system, resulting in a procedural delay. Following replacement of the DWS, the procedure was completed successfully with no adverse patient consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50287646 ↗
  • MalfunctionMar 31, 2026Application Program Problem; Application Program Freezes or Fails to Launch

    DURING THE RADIOFREQUENCY ABLATION SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THERE WAS A PROCEDURAL DELAY. IT WAS NOTED THAT THE ENSITE X SYSTEM WAS "FREEZING". IT WAS NOTED THAT THE MOUSE WAS ABLE TO MOVE, BUT WAS UNABLE TO CLICK ANY MENUS, WINDOWS, OR FUNCTIONS. TROUBLESHOOTING INCLUDED REBOOTING THE ENSITE X SYSTEM AND REPLACING THE ENSITE X DWS, WHICH RESOLVED THE ISSUE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24744659 ↗
  • MalfunctionFeb 25, 2026Communication or Transmission Problem

    DURING A SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE PROCEDURE WAS ABORTED. VALIDATION COULD NOT BE PERFORMED. A LEFT ARM AND LEFT LEG ERROR, ALONG WITH SYSTEM REFERENCE PATCH ERRORS WERE DISPLAYED. THE PATCHES WERE EXCHANGED, AND THE AMPLIFIER AND DWS WERE REBOOTED TWICE. THREE NEW STUDIES WERE STARTED, THE SURFACELINK WAS RESEATED, AND PATCHES WERE RESEATED INTO THE SURFACELINK. THE SURFACELINK WAS EXCHANGED WITH NO RESOLUTION. THE PROCEDURE WAS ABORTED, AND NO PROCEDURE WAS PERFORMED (NOT EVEN GROIN ACCESS). THOUGH THE PATIENT HAD BEEN UNDER ANESTHESIA. FURTHER TROUBLESHOOTING AT A LATER DATE DETERMINED THAT THE ISSUE WAS THE DWS WHICH HAS BEEN SINCE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24445116 ↗
  • MalfunctionFeb 11, 2026Communication or Transmission Problem

    DURING AN ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA PROCEDURE THERE WAS A DELAY. THE PROCEDURE WAS STARTED WITH ENSITEX DWS 9 (V5.0.1) AND HIS, CR, AND HRA CATHETERS WERE INSERTED IN THE PATIENT. ALL THE CATHETERS APPEARED ELONGATED AND STACKED ON EACH OTHER. THOUGH THE EGMS AND SURFACE SIGNALS WERE FINE. THE SYSTEM WAS REBOOTED WITH NO RESOLUTION. A NEW CASE WAS STARTED BUT AN AMPLIFIER ISSUE OCCURRED AND TECH SUPPORT WAS CALLED. EVERYTHING HAD TO BE DISCONNECTED FROM THE AMPLIFIER TO GET IT TO BOOT UP CORRECTLY. AFTER THAT, THERE WERE SEVERAL IMPEDANCE ERRORS AND GEOGRAPHY COULD NOT BE COLLECTED. THE CATHETERS WERE UNABLE TO SHOW OR WERE DISTORTED. THE NAVX PATCHES WERE REPLACED WITH NO RESOLUTION. A TACTIFLEX CATHETER WAS USED BUT THE SAME ISSUE OCCURRED WITH THE CATHETER APPEARING DISTORTED IN NAVX. IN VOXEL MODE, THE SE GUMBALL KEPT HAVING "INVALID SHAFT ELECTRODE DATA" AND "POOR IMPEDANCE DATA" ERRORS. SHADOWS WERE DROPPED, AND THE FORCE ARROW WAS WORKING AS EXPECTED BUT COULD NOT COLLECT DATA. THE PROCEDURE WAS COMPLETED WITH USING FLUOROSCOPY AND SIGNALS. FURTHER TROUBLESHOOTING AT A LATER DATE DETERMINED THAT THE ISSUE WAS THE DWS WHICH HAS BEEN SINCE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24315139 ↗
  • MalfunctionFeb 6, 2026Application Program Freezes or Fails to Launch

    DURING A PFA PROCEDURE, THERE WAS A DELAY DUE TO A COMMUNICATION AND SELF-TEST ISSUE WITH THE AMPLIFIER AND A FREEZING ISSUE WITH THE DISPLAY WORKSTATION. THE AMPLIFIER DISCONNECTED AND DISPLAYED AN AMBER LIGHT. THERE WERE ABOUT 10 REBOOTS OF THE AMPLIFIER AND 2-3 REBOOTS OF THE DISPLAY WORKSTATION WHILE THE PATIENT WAS ON THE TABLE WITH CORRECT BOOT UP SEQUENCE BEFORE THE CASE COULD BE STARTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24275397 ↗
  • MalfunctionFeb 6, 2026Application Program Freezes or Fails to Launch

    DURING A PFA PROCEDURE, THERE WAS A DELAY DUE TO A COMMUNICATION AND SELF-TEST ISSUE WITH THE AMPLIFIER AND A FREEZING ISSUE WITH THE DISPLAY WORKSTATION. THE AMPLIFIER DISCONNECTED AND DISPLAYED AN AMBER LIGHT. THERE WERE ABOUT 10 REBOOTS OF THE AMPLIFIER AND 2-3 REBOOTS OF THE DISPLAY WORKSTATION WHILE THE PATIENT WAS ON THE TABLE WITH CORRECT BOOT UP SEQUENCE BEFORE THE CASE COULD BE STARTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24275390 ↗
  • MalfunctionFeb 6, 2026Application Program Freezes or Fails to Launch

    DURING A PFA PROCEDURE, THERE WAS A DELAY DUE TO A COMMUNICATION AND SELF-TEST ISSUE WITH THE AMPLIFIER AND A FREEZING ISSUE WITH THE DISPLAY WORKSTATION. THE AMPLIFIER DISCONNECTED AND DISPLAYED AN AMBER LIGHT. THERE WERE ABOUT 10 REBOOTS OF THE AMPLIFIER AND 2-3 REBOOTS OF THE DISPLAY WORKSTATION WHILE THE PATIENT WAS ON THE TABLE WITH CORRECT BOOT UP SEQUENCE BEFORE THE CASE COULD BE STARTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24275384 ↗
  • MalfunctionFeb 6, 2026Application Program Freezes or Fails to Launch

    DURING A PFA PROCEDURE, THERE WAS A DELAY DUE TO A COMMUNICATION AND SELF-TEST ISSUE WITH THE AMPLIFIER AND A FREEZING ISSUE WITH THE DISPLAY WORKSTATION. THE AMPLIFIER DISCONNECTED MID APPLICATION OF PFA. THE AMPLIFIER DISPLAYED AN AMBER LIGHT. THE SYSTEM WAS REBOOTED WITH EVERYTHING DISCONNECTED. THE CATHETER WOULD THEN NOT VISUALIZE ON ENSITE AND FROZE REQUIRING LOGGING OUT OF THE SYSTEM. THERE WAS AN INTERNAL "RECONNECT TO PFA GENERATOR" WHEN THE VOLT ICON WAS STILL GREEN AND "RECONNECT TO SE SYSTEM" ERROR. THE CURRENT PFA GENERATOR OPTIONS WERE GREYED OUT AND DESPITE RESTARTING THE CURRENT PFA GENERATOR THERE WAS NO RESOLUTION. THE CASE WAS EXITED AND RELOADED, BUT IT WAS NOT POSSIBLE TO RESUME THE CASE. THE PROCEDURE WAS COMPLETED WITHOUT ENSITE USING X-RAY AND WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24275381 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →