EnSite™ ENSITE-SW-4.0.1Recalls & reported problems
St. Jude Medical, Inc. · X EP System Software Upgrade · Class 2
All models of EnSite (37)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067053114), received between Nov 21, 2025 and Dec 1, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch1
- Display or Visual Feedback Problem1
- Erratic or Intermittent Display1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
DURING THE PFA PROCEDURE, OUTPUT ISSUES RESULTED IN A PROCEDURAL DELAY. AFTER COLLECTING THE VOLTAGE MAP WITH HD GRID, THE HD GRID WAS WITHDRAWN AND THE NON-ABBOTT PFA CATHETER (FARAWAVE BY BOSTON SCIENTIFIC) WAS INSERTED. CONTACT INDEX BASELINE WAS STARTED AND THE NON-ABBOTT CATHETER (FARAWAVE) PFA CATHETER FLIPPED BACKWARDS IN APPEARANCE EVEN THOUGH IT HAD NOT BEEN PHYSICALLY MOVED. PIN LOCATION WAS CHECKED; THE CATHETER WAS REMOVED FROM ENSITEX AND A NEW ONE WAS LOADED. THE CATHETER PRESETS WERE CHANGED AND RESPIRATORY COMPENSATION WAS PERFORMED WITH NO RESOLUTION. THE AMPLIFIER WAS RESET AND THIS DID NOT FIX THE ISSUE. THE CASE WAS CLOSED; THE AMPLIFIER WAS TURNED OFF AND THE DWS9 COMPUTER WAS REBOOTED. THE CASE WAS RESTARTED; RENDERING WAS APPROPRIATE BUT COULD NOT TAKE SHADOWS OF THE NON-ABBOTT (FARAWAVE) PFA CATHETER. NEXT THE STUDY WAS CLOSED AND THE AMPLIFIER AND DWS9 WERE REBOOTED AGAIN AND STARTED A COMPLETELY NEW CASE WHICH RESOLVED THE ISSUE. RENDERING WAS APPROPRIATE AND ALL TAGS AND MAPPING WORKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23681617 ↗MalfunctionNov 21, 2025Display or Visual Feedback Problem; Application Program Freezes or Fails to Launch; Display or Visual Feedback Problem; Application Program Freezes or Fails to Launch
DURING A PFA PROCEDURE, THERE WAS A SELF-TEST ISSUE WITH THE AMPLIFIER AND SOFTWARE ISSUES RESULTING IN A DELAY. AFTER SETTING UP FOR THE CASE, THE MODE WAS CHANGED FROM VOXEL TO NAVX TO VISUALIZE THE CS AND HIS CATHETERS THAT WERE POSITIONED IN THE HEART UNDER FLUOROSCOPY. THE CATHETERS FROZE AT THE BASE OF THE SCREEN BUT THERE WERE CS AND HIS EGM SIGNALS ON BOTH ENSITE X AND THE NON-ABBOTT (PRUCKA) RECORDING SYSTEM. ADDITIONALLY, WHEN RESPIRATION COMPENSATION WAS PERFORMED, THE FOLLOWING ERROR DISPLAYED, ¿RESPIRATION DATA COLLECTION FAILED. MAY BE CAUSED BY A DAMAGED FIBER CABLE, COMPLEX MODELS, OR RAPID SUCCESSIVE RECORDINGS. RECOLLECT RESPIRATION DATA BEFORE PROCEEDING." THE AMPLIFIER WAS RESTARTED, WHICH RESULTED IN A FLASHING ORANGE LIGHT. THE AMPLIFIER WAS TURNED OFF AGAIN AND THE CASE WAS CLOSED. THE CASE WAS REOPENED AND THE AMPLIFIER TURNED BACK ON, BUT THE FLASHING ORANGE LIGHT REMAINED. EVERYTHING WAS SHUT OFF AND ALL CONNECTIONS TO THE FRONT PANEL OF THE AMPLIFIER WERE RECONNECTED. THE AMPLIFIER WAS THEN TURNED ON AND THE LIGHT WAS GREEN. REVALIDATION WAS PERFORMED, METAL CHECKS WERE DONE AND THE PRS'S WERE GREEN. THE CATHETERS WERE POSITIONED IN THE HEART. THE ADVISOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23618360 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.