St. Jude Medical, Inc. · X EP System Software Upgrade · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05415067053121), received between Sep 17, 2025 and Feb 3, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem1
  • Display or Visual Feedback Problem1
  • Erratic or Intermittent Display1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Application Program Problem

    DURING A PULSED FIELD ABLATION PROCEDURE, THERE WAS A DELAY DUE TO A GRAPHICAL USER INTERFACE ISSUE. WHEN USING THE ENSITE X SYSTEM (V3.1.1) IN VOXEL MODE, THE VOXEL POINT ACQUISITION AND GEOMETRY ACQUISITION WERE POSSIBLE. DESPITE ACQUIRING MULTIPLE VOXEL POINTS, THE CATHETER ELECTRODE NAVIGATION DISPLAY REMAINED GRAYED - PREVENTING MAP ACQUISITION. WHEN THE MODE WAS SWITCHED FROM VOXEL TO NAVX, THE CATHETER NAVIGATION DISPLAY APPEARED TO HAVE COLLAPSED AND COULD NOT BE DISPLAYED PROPERLY. ADDITIONAL TROUBLESHOOTING INCLUDED REPLACING THE CONNECTION CABLE AND CATHETERS, RESTARTING THE SYSTEM, AND PERFORMING REVALIDATION, BUT THE ISSUE PERSISTED. THE ISSUE WAS RESOLVED BY REPLACING THE ENSITE X SYSTEM WITH A NON-ABBOTT 3D SYSTEM. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24254984 ↗
  • MalfunctionJan 8, 2026Erratic or Intermittent Display

    DURING TYPICAL ATRIAL FLUTTER PROCEDURE USING ADVISOR HD GRID IN VOXEL MODE, THE CATHETERS BEGAN TO INTERMITTENTLY DISAPPEAR FROM THE SCREEN UNTIL THEY NO LONGER APPEARED AT ALL RESULTING IN A DELAY. THE EXTENSIONS WERE CHANGED, THE AMPLIFIER WAS RESTARTED, AND THE CATHETERS WERE SWITCHED TO DIFFERENT CHANNELS, BUT THEY STILL DID NOT APPEAR. THE MODE WAS SWITCHED TO NAVX, BUT THE CATHETERS CONTINUED TO BE ABSENT FROM THE SCREEN. THE CASE WAS RESTARTED; HOWEVER, WHEN ATTEMPTING TO RESUME IT, THE SYSTEM DID NOT ALLOW IT BECAUSE A MESSAGE APPEARED ON THE SCREEN STATING THAT THE PATCHES HAD ALREADY EXCEEDED THE TIME LIMIT TO RESUME THE PROCEDURE, EVEN THOUGH THE PATCH VALIDATION HAD BEEN PERFORMED LESS THAN TWO HOURS EARLIER. IT WAS ATTEMPTED TO RESUME UNDER A NEW CASE, BUT IT WAS NOT POSSIBLE BECAUSE THE SAME MESSAGE APPEARED ON THE SCREEN. THE PATCHES WERE REPLACED, AND SINCE PATCHES HAD ALREADY BEEN VALIDATED WITH THE HD GRID MAPPING CATHETER AND THE TACTIFLEX ABLATION CATHETER, BOTH CATHETERS HAD TO BE REPLACED BECAUSE THE SYSTEM DID NOT ALLOW A NEW VALIDATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24012154 ↗
  • MalfunctionSep 17, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    DURING A PREMATURE VENTRICULAR CONTRACTION PROCEDURE, THERE WAS A MAPPING SHIFT RESULTING IN A CLINICALLY SIGNIFICANT DELAY. AN RV MODEL AND MAP WAS CREATED THE MAPPING CATHETER. WHEN THE MAPPING CATHETER WAS EXCHANGED WITH AN ABLATION CATHETER TO BEGIN ABLATION, IT WAS NOTED THAT ALL OF THE PRS-P SENSORS TURNED RED. IT APPEARED AS THOUGH THEY WERE STILL IN THE SAME INITIAL LOCATION, HOWEVER, THEY WERE ALL RED. TROUBLESHOOTING INCLUDED ATTEMPTING TO MANUALLY ADJUST THE PRS-PS TO TURN GREEN, BUT THE ISSUE PERSISTED. IN ORDER TO CONTINUE THE PROCEDURE, A NEW MODEL WAS CREATED, FOLLOWED BY RESETTING THE METAL BASELINE, AND CONTINUING TO MAP AND ABLATE. AFTER THE ABLATION WAS COMPLETED, THE PHYSICIAN WAITED TWENTY MINUTES TO MAKE SURE THAT THE PVC¿S WERE GONE. AFTER THE WAITING PERIOD WAS COMPLETED, THE SHEATHS WERE PULLED AND PLANNED TO GET THE PATIENT OFF THE TABLE. AT THIS POINT, THE PVC¿S RETURNED, SO THE PHYSICIAN DECIDED TO GET ACCESS AND PUT THE CATHETERS BACK IN TO ABLATE MORE. HOWEVER, WHEN THE CATHETERS WERE PLACED BACK INTO THE PATIENT, IT WAS CLEAR THAT THE MODEL HAS SHIFTED. THERE WAS NO ERROR MESSAGES DISPLAYED AND NO INDICATORS THAT WOULD SUGGEST THAT THERE WAS ANYTHING

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23077886 ↗