Isiris™ a ALFA01Recalls & reported problems
Ambu A/S · Stent Removal Device · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05707480136063), received between Feb 1, 2023 and Aug 14, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Mechanical Problem1
Patient problems
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 14, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
DURING PROCEDURE, THE OPERATOR WANTED TO GRASP THE JJ CATHETER AFTER THE ISIRIS ALFA WAS INTRODUCTION INTO THE BLADDER. THE CLAW OF THE GRASPER DETACHED. A SECOND ENDOSCOPE WAS THEN USED TO RETRIVE THE DETACHED CLAW FROM THE BLADDER CAVITY.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 17539347 ↗MalfunctionFeb 17, 2023Mechanical Problem; Mechanical Problem
DURING PROCEDURE THE LOWER PART OF THE CYSTOSCOPE'S GRASPER FBROKE AND DISLODGED INTO THE PATIENT'S BLADDER. GRASPER WAS REMOVED USING ANOTHER CYSTOSCOPE. PATIENT OUTCOME NOT AFFECTED..
Patient problems: Foreign Body In Patient
MAUDE report 16395745 ↗MalfunctionFeb 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
DURING AN ATTEMPT TO REMOVE A STENT WITH AN ISIRIS CYSTOSCOPE, THE OPERATOR HAD DIFFICULTIES TO REMOVE THE STENT, THE GRASPER OF THE SCOPE, AND DISLODGE TO PATIENT'S BLADDER. A FLEXIBLE CYSTOSCOPE WAS USED TO RETRIEVE THE GRASPER IN THE BLADDER CAVITY AND TO REMOVE THE STENT. THE PROCEDURE WAS PROLONGED WITH 30 MINUTES. NO HARM TO PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 16276635 ↗
Other Cystoscope and accessories, flexible/rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall54 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.