VivaSight™ 2 DLT Kit 35 Fr - LEFT 412351000Recalls & reported problems
Ambu A/S · VivaSight 2 DLT Kit 35 Fr - LEFT · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05707480145706), received between Aug 28, 2024 and Aug 28, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 3, 2022Class IIOpen, Classified
Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.
Z-1420-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult to Insert2
- Structural Problem2
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2024Difficult to Insert; Structural Problem; Difficult to Insert; Structural Problem
ACCORDING TO INFORMATION FROM THE CUSTOMER. IN CONNECTION WITH INTUBATING WITH AMBU VIVASIGHT DOUBLE-LUMEN TUBE WITH CAMERA, DIFFICULT INTUBATION OCCURED. THE TUBE WAS HYPERANGULATED, WHICH MADE IT DIFFICULT TO PASS THE EPIGLOTTIS AND TRACHEA. THERE WAS BLEEDING AROUND THE EPIGLOTTIS, AND UPON EXTUBATION, BLOOD WAS SEEN ON THE CUFF AND THE TIP OF THE TUBE. ACCORDING TO FOLLOW-UP INFORMATION FROM THE CUSTOMER, THE PATIENT WAS FINE THE NEXT DAY AND DID NOT COMPLAIN ABOUT SORE THROAT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20088095 ↗MalfunctionAug 28, 2024Difficult to Insert; Structural Problem; Difficult to Insert; Structural Problem
ACCORDING TO INFORMATION FROM THE CUSTOMER. IN CONNECTION WITH INTUBATING WITH AMBU VIVASIGHT DOUBLE-LUMEN TUBE WITH CAMERA, DIFFICULT INTUBATION OCCURED. THE TUBE WAS HYPERANGULATED, WHICH MADE IT DIFFICULT TO PASS THE EPIGLOTTIS AND TRACHEA. THERE WAS BLEEDING AROUND THE EPIGLOTTIS, AND UPON EXTUBATION, BLOOD WAS SEEN ON THE CUFF AND THE TIP OF THE TUBE. ACCORDING TO FOLLOW-UP INFORMATION FROM THE CUSTOMER, THE PATIENT WAS FINE THE NEXT DAY AND DID NOT COMPLAINT ABOUT SORE THROAT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20085639 ↗
Other TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR) devices
- HUDSON RCI IPN048996Teleflex Incorporated · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)64 reports
- VivaSight™ 2 DLT Kit 37 Fr - LEFT 412371000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall10 reports
- HUDSON RCI IPN048995Teleflex Incorporated · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)58 reports
- VivaSight™ 2 DLT Kit 39 Fr - LEFT 412391000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall6 reports
- VivaSight™ 2 DLT Kit 41 Fr - LEFT 412411000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall2 reports
- Portex 198-35LIcu Medical, Inc. · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)48 reports
- Portex 198-37LIcu Medical, Inc. · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)44 reports
- HUDSON RCI IPN049000Teleflex Incorporated · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)40 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page