HUDSON RCI IPN048996Recalls & reported problems
Teleflex Incorporated · ET TUBE, SHER-I-BRONCH, LS, 37 FR · Class 2
All models of HUDSON RCI (152)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Teleflex Sher-I-Bronch endotracheal tube is a single-use breathing tube for intubation. The 64 reports mostly describe the tube or its cuff giving way or failing to hold air, plus connectors that would not tighten and leaked air during use. There are 64 malfunctions, no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 14026704617408), received between Feb 25, 2021 and Jul 31, 2023 — classified as 64 malfunctions.
The most frequently reported specific device problem is “Break” (42 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break42
- Gas/Air Leak7
- Inflation Problem5
- Fracture3
- Leak/Splash2
- Contamination1
- Fitting Problem1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions57
- Insufficient Information8
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 31, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT "THE DOCTOR FOUND CUFF CANNOT BE INFLATED WHEN DOING CLINICAL SETTING BEFORE USING ON PATIENT. THEN CHANGED NEW ONE, NO IMPACT ON PATIENT. INVOLVED 1 PC." NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17429576 ↗MalfunctionJul 21, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED THAT DURING USE ON A PATIENT, "THE Y CONNECTOR COULD NOT BE ATTACHED TO THE MACHINE TIGHTLY, AIR LEAKING HAPPENED." NO MEDICAL INTERVENTION REQUIRED, "THE SITUATION WAS FOUND IN TIME". NO PATIENT HARM, DESATURATION, OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17372314 ↗MalfunctionJun 15, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT "THE DOCTOR FOUND CUFF CANNOT BE INFLATED WHEN DOING CLINICAL SETTING BEFORE USING ON PATIENT. THEN CHANGED NEW ONE, NO IMPACT ON PATIENT". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17139071 ↗MalfunctionMar 24, 2023Inflation Problem; Inflation Problem
REPORTED ISSUE: THE DOCTOR FOUND CUFF CANNOT BE INFLATED BEFORE USING IT ON THE PATIENT; IT WAS CHANGED TO A NEW ONE. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16616374 ↗MalfunctionMar 24, 2023Inflation Problem; Inflation Problem
REPORTED ISSUE: THE DOCTOR FOUND CUFF CANNOT BE INFLATED BEFORE USING IT ON THE PATIENT; IT WAS CHANGED TO A NEW ONE. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16616141 ↗MalfunctionMar 24, 2023Inflation Problem; Inflation Problem
REPORTED ISSUE: THE DOCTOR FOUND CUFF CANNOT BE INFLATED BEFORE USING IT ON THE PATIENT; IT WAS CHANGED TO A NEW ONE. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16616076 ↗MalfunctionMar 24, 2023Gas/Air Leak; Gas/Air Leak
REPORTED ISSUE: THE DOCTOR FOUND CUFF LEAKAGE WHEN DOING CLINICAL SETTING BEFORE USING ON PATIENT. THEN CHANGED NEW ONE, NO IMPACT ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16615481 ↗MalfunctionMar 17, 2023Break; Break
REPORTED ISSUE: THE DOCTOR FOUND THE CONNECTOR BROKEN WHEN USING IT ON A PATIENT. IT WAS CHANGED TO A NEW ONE; NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16567460 ↗MalfunctionMar 17, 2023Break; Break
REPORTED ISSUE: THE DOCTOR FOUND THE CONNECTOR BROKEN WHEN USING IT ON A PATIENT. IT WAS CHANGED TO A NEW ONE; NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16567364 ↗MalfunctionFeb 27, 2023Break; Break
REPORTED ISSUE: THE DOCTOR FOUND THE CONNECTOR BROKEN WHEN IN A CLINICAL SETTING BEFORE USING IT ON A PATIENT. A NEW ONE WAS USED. NO IMPACT ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16448456 ↗MalfunctionOct 11, 2022Break; Break
IT WAS REPORTED THAT: "THE NURSE FOUND CONNECTOR BROKEN DURING CLINICAL SETTING BEFORE USING ON PATIENT. NO IMPACT ON PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15576970 ↗MalfunctionAug 29, 2022Leak/Splash; Leak/Splash
COMPLAINT REPORTS THAT: THE BRONCHIAL BALLOON LEAKED, DURING USE ON A PATIENT.
Patient problems: Insufficient Information
MAUDE report 15310229 ↗MalfunctionAug 24, 2022Break; Break
ALLEGED QUALITY ISSUE/INCIDENT: "PATIENT WAS INTUBATED, WHEN CLINICIAN ATTEMPTED TO ATTACH TO Y CONNECTOR, IT JUST BROKE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15282262 ↗MalfunctionAug 23, 2022Break; Break
IT WAS REPORTED THAT: "THE DOCTOR FOUND CUFF DETACHED FROM THE TUBE DURING CLINICAL SETTING BEFORE USING ON PATIENT. NO IMPACT ON PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15277623 ↗MalfunctionAug 2, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT: "THE DOCTOR FOUND CUFF LEAKAGE WHEN USING ON PATIENT. THEN CHANGED NEW ONE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145498 ↗MalfunctionAug 2, 2022Break; Break
IT WAS REPORTED THAT: "THE DOCTOR FOUND Y CONNECTOR BOKEN DURING CLINICAL SETTING BEFORE USING ON PATIENT. NO IMPACT ON PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145335 ↗MalfunctionAug 2, 2022Break; Break
IT WAS REPORTED THAT: "THE DOCTOR FOUND CONNECTOR BOKEN WHEN USING ON PATIENT.THEN CHANGED NEW ONE. NO IMPACT ON PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145240 ↗MalfunctionAug 2, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT: "THE DOCTOR FOUND CUFF LEAKAGE DURING CLINICAL SETTING BEFORE USING ON PATIENT. NO IMPACT ON PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145215 ↗MalfunctionJun 8, 2022Break; Break
COMPLAINT FROM MAUDE DATABASE: "NARRATIVE FROM STAFF: AFTER INTUBATING WITH LEFT DOUBLE LUMEN TUBE THE TWO-WAY SPLITTER FROM THE Y-TUBING BROKE". NO PATIENT HARM OR INJURY REPORTED. PATIENT CONDITION NOT REPORTED.
Patient problems: Insufficient Information
MAUDE report 14631484 ↗MalfunctionApr 22, 2022Break; Break
IT WAS REPORTED THAT "THE DOCTOR FOUND CONNECTOR BOKEN DURING CLINICAL SETTING BEFORE USING ON PATIENT". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14189502 ↗MalfunctionApr 13, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "THE DOCTOR FOUND CUFF LEAKAGE WHEN USING ON PATIENT AND CHANGED NEW ONE". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14102065 ↗MalfunctionFeb 7, 2022Break; Break
IT WAS REPORTED "THE Y CONNECTOR WAS FOUND BROKEN WHEN CLINICAL STAFF OPEN IT". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13475884 ↗MalfunctionDec 16, 2021Fracture; Fracture
IT WAS REPORTED "PURCHASED DOUBLE LUMEN BRONCH TUBE AND THE FO DOUBLE SWIVEL CONNECTOR ASSEMBLY/ 1 WYE HAS A CRACK IN THE PLASTIC AT THE ELBOW". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13022495 ↗MalfunctionDec 16, 2021Fracture; Fracture
IT WAS REPORTED "PURCHASED DOUBLE LUMEN BRONCH TUBE AND THE FO DOUBLE SWIVEL CONNECTOR ASSEMBLY/ 1 WYE HAS A CRACK IN THE PLASTIC AT THE ELBOW". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13022438 ↗MalfunctionDec 16, 2021Fracture; Fracture
IT WAS REPORTED "PURCHASED DOUBLE LUMEN BRONCH TUBE AND THE FO DOUBLE SWIVEL CONNECTOR ASSEMBLY/ 1 WYE HAS A CRACK IN THE PLASTIC AT THE ELBOW". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13022420 ↗
Other TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR) devices
- VivaSight™ 2 DLT Kit 37 Fr - LEFT 412371000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall10 reports
- HUDSON RCI IPN048995Teleflex Incorporated · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)58 reports
- VivaSight™ 2 DLT Kit 39 Fr - LEFT 412391000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall6 reports
- VivaSight™ 2 DLT Kit 35 Fr - LEFT 412351000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall2 reports
- VivaSight™ 2 DLT Kit 41 Fr - LEFT 412411000Ambu A/S · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)1 recall2 reports
- Portex 198-35LIcu Medical, Inc. · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)48 reports
- Portex 198-37LIcu Medical, Inc. · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)44 reports
- HUDSON RCI IPN049000Teleflex Incorporated · TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR)40 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page