VivaSight™ 2 DLT Kit 39 Fr - LEFT 412391000Recalls & reported problems

Ambu A/S · VivaSight 2 DLT Kit 39 Fr - LEFT · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05707480145768), received between Dec 5, 2024 and Feb 19, 2026 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Structural Problem” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 3, 2022Class IIOpen, Classified

    Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.

    Z-1420-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Structural Problem4
  • Inflation Problem2

Patient problems

  • Upper Respiratory Tract Injury4
  • Aspiration Pneumonitis2
  • Insufficient Information2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2026Inflation Problem

    DOCTOR ATTEMPTED STANDARD THORACIC PROCEDURE - INTUBATION OF A PATIENT USING VIVASIGHT 2 DLT. FIRST, THEY USED A 41FR AND THEN DOWNSIZED TO A 39FR. BOTH DLTS HAD THE CUFF BURST. CONSULTANT INFLATED THE CUFF WITH 2-3ML OF AIR AND THE CUFF WOULD NATURALLY DEFLATE. ADDITIONAL INFORMATION FROM THE HOSPITAL WAS PROVIDED BY THE DOCTOR: 'INTERMITTENTLY WE RUN INTO AN ISSUE IN WHICH DESPITE THE OPERATIVE LUNG BEING ISOLATED IT IS NOT DEFLATING WHEN OPENED TO AIR. CONDITION PERSISTS EVEN WITH APPLICATION OF DIRECT SUCTION. AIRWAY MANAGED DESPITE COMPROMISES TO THE SURGERY. MAY HAVE RESULTED IN THE NEED FOR LARGER INCISIONS AND EVEN UNNECESSARY THORACOTOMIES. TUBE CHANGED TO SHILEY BRONCOCATH OR BRONCHOSCOPE USED TO DOUBLE CHECK PLACEMENT DESPITE TUBE SEEMINGLY BEING WELL PLACED ACCORDING TO THE INBUILT CAMERA.'' IT IS NOT INDICATED THAT THE PATIENT EXPERIENCED ASSOCIATED HARM OF HYPOXIA, SINCE PATIENT'S AIRWAY WAS MANAGED. HOWEVER, DUE TO LACK OF INFORMATION ON THE COMPROMISES TO THE SURGERY, THE INCIDENT IS BEING REPORTED. NOTE THAT TWO REPORTS ARE BEING SUBMITTED DUE TO THE INVOLVEMENT OF TWO DEVICES. THE MANUFACTURER REFERENCE NUMBER OF THE SECOND REPORT IS 1220828-2026-00004.

    Patient problems: Insufficient Information

    MAUDE report 24393046 ↗
  • MalfunctionFeb 19, 2026Inflation Problem

    DOCTOR ATTEMPTED STANDARD THORACIC PROCEDURE - INTUBATION OF A PATIENT USING VIVASIGHT 2 DLT. FIRST, THEY USED A 41FR AND THEN DOWNSIZED TO A 39FR. BOTH DLTS HAD THE CUFF BURST. CONSULTANT INFLATED THE CUFF WITH 2-3ML OF AIR AND THE CUFF WOULD NATURALLY DEFLATE. ADDITIONAL INFORMATION FROM THE HOSPITAL WAS PROVIDED BY THE DOCTOR: 'INTERMITTENTLY WE RUN INTO AN ISSUE IN WHICH DESPITE THE OPERATIVE LUNG BEING ISOLATED IT IS NOT DEFLATING WHEN OPENED TO AIR. CONDITION PERSISTS EVEN WITH APPLICATION OF DIRECT SUCTION. AIRWAY MANAGED DESPITE COMPROMISES TO THE SURGERY. MAY HAVE RESULTED IN THE NEED FOR LARGER INCISIONS AND EVEN UNNECESSARY THORACOTOMIES. TUBE CHANGED TO SHILEY BRONCOCATH OR BRONCHOSCOPE USED TO DOUBLE CHECK PLACEMENT DESPITE TUBE SEEMINGLY BEING WELL PLACED ACCORDING TO THE INBUILT CAMERA.'' IT IS NOT INDICATED THAT THE PATIENT EXPERIENCED ASSOCIATED HARM OF HYPOXIA, SINCE PATIENT'S AIRWAY WAS MANAGED. HOWEVER, DUE TO LACK OF INFORMATION ON THE COMPROMISES TO THE SURGERY, THE INCIDENT IS BEING REPORTED. NOTE THAT TWO REPORTS ARE BEING SUBMITTED DUE TO THE INVOLVEMENT OF TWO DEVICES. THE MANUFACTURER REFERENCE NUMBER OF THE SECOND REPORT IS 9610691-2026-00004.

    Patient problems: Insufficient Information

    MAUDE report 24387197 ↗
  • InjuryDec 5, 2024Structural Problem; Structural Problem

    THE INCIDENT LEADING TO TONSIL INJURIES OCCURRED IN TWO SEPARATE PROCEDURES (THE SECOND INCIDENT IS COVERED BY 1220828-2024- 00024). A CRNA WAS TRYING TO INTUBATE WITH THE VIVASIGHT 2 DLT PRODUCT AND THEY HAD TO ADJUST THE DISTANCE OF TIP OF THE TUBE PAST THE VOCAL CORDS. THE CURVE AT THE END OF THE BRONCHIAL LUMEN WAS MORE ANGULATED THAN A TYPICAL ROBERT-SHAW DOUBLE LUMEN ENDOTRACHEAL TUBE, AND GOT MORE EASILY CAUGHT ON THE TRACHEAL RINGS. ONE COULD EITHER ADVANCE THE BRONCHIAL CUFF FURTHER PAST THE VOCAL CORDS WITH THE STYLET, OR IF THE STYLET WAS REMOVED AFTER THE TIP WAS IN THE TRACHEA, ONE COULD TURN THE TUBE 120-180 DEGREES LEFTWARD WHILE ADVANCING AND THEN RETURN IT TO 90 DEGREES ONCE PAST THE RING THAT THE TIP GOT HUNG UP ON.

    Patient problems: Upper Respiratory Tract Injury

    MAUDE report 20854058 ↗
  • InjuryDec 5, 2024Structural Problem; Structural Problem

    THE INCIDENT LEADING TO TONSIL INJURIES OCCURRED IN TWO SEPARATE PROCEDURES (THE SECOND INCIDENT IS COVERED BY 1220828-2024-00025). A CRNA WAS TRYING TO INTUBATE WITH THE VIVASIGHT 2 DLT PRODUCT AND THEY HAD TO ADJUST THE DISTANCE OF TIP OF THE TUBE PAST THE VOCAL CORDS. THE CURVE AT THE END OF THE BRONCHIAL LUMEN WAS MORE ANGULATED THAN A TYPICAL ROBERT-SHAW DOUBLE LUMEN ENDOTRACHEAL TUBE, AND GOT MORE EASILY CAUGHT ON THE TRACHEAL RINGS. ONE COULD EITHER ADVANCE THE BRONCHIAL CUFF FURTHER PAST THE VOCAL CORDS WITH THE STYLET, OR IF THE STYLET WAS REMOVED AFTER THE TIP WAS IN THE TRACHEA, ONE COULD TURN THE TUBE 120-180 DEGREES LEFTWARD WHILE ADVANCING AND THEN RETURN IT TO 90 DEGREES ONCE PAST THE RING THAT THE TIP GOT HUNG UP ON.

    Patient problems: Aspiration Pneumonitis; Upper Respiratory Tract Injury

    MAUDE report 20853676 ↗
  • InjuryDec 5, 2024Structural Problem; Structural Problem

    THE INCIDENT LEADING TO TONSIL INJURIES OCCURRED IN TWO SEPARATE PROCEDURES (THE SECOND INCIDENT IS COVERED BY 9610691-2024-00023). A CRNA WAS TRYING TO INTUBATE WITH THE VIVASIGHT 2 DLT PRODUCT AND THEY HAD TO ADJUST THE DISTANCE OF TIP OF THE TUBE PAST THE VOCAL CORDS. THE CURVE AT THE END OF THE BRONCHIAL LUMEN WAS MORE ANGULATED THAN A TYPICAL ROBERT-SHAW DOUBLE LUMEN ENDOTRACHEAL TUBE AND GOT MORE EASILY CAUGHT ON THE TRACHEAL RINGS. ONE COULD EITHER ADVANCE THE BRONCHIAL CUFF FURTHER PAST THE VOCAL CORDS WITH THE STYLET, OR IF THE STYLET WAS REMOVED AFTER THE TIP WAS IN THE TRACHEA, ONE COULD TURN THE TUBE 120-180 DEGREES LEFTWARD WHILE ADVANCING AND THEN RETURN IT TO 90 DEGREES ONCE PAST THE RING THAT THE TIP GOT HUNG UP ON.

    Patient problems: Upper Respiratory Tract Injury

    MAUDE report 20850556 ↗
  • InjuryDec 5, 2024Structural Problem; Structural Problem

    THE INCIDENT LEADING TO TONSIL INJURIES OCCURRED IN TWO SEPARATE PROCEDURES (THE SECOND INCIDENT IS COVERED BY 9610691-2024-00024). A CRNA WAS TRYING TO INTUBATE WITH THE VIVASIGHT 2 DLT PRODUCT AND THEY HAD TO ADJUST THE DISTANCE OF TIP OF THE TUBE PAST THE VOCAL CORDS. THE CURVE AT THE END OF THE BRONCHIAL LUMEN WAS MORE ANGULATED THAN A TYPICAL ROBERT-SHAW DOUBLE LUMEN ENDOTRACHEAL TUBE, AND GOT MORE EASILY CAUGHT ON THE TRACHEAL RINGS. ONE COULD EITHER ADVANCE THE BRONCHIAL CUFF FURTHER PAST THE VOCAL CORDS WITH THE STYLET, OR IF THE STYLET WAS REMOVED AFTER THE TIP WAS IN THE TRACHEA, ONE COULD TURN THE TUBE 120-180 DEGREES LEFTWARD WHILE ADVANCING AND THEN RETURN IT TO 90 DEGREES ONCE PAST THE RING THAT THE TIP GOT HUNG UP ON.

    Patient problems: Aspiration Pneumonitis; Upper Respiratory Tract Injury

    MAUDE report 20850550 ↗

Other TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR) devices

All TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION (W/WO CONNECTOR) devices →