Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05707480152032), received between Aug 21, 2024 and Aug 21, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2024Mechanical Jam; Mechanical Jam
ACCORDING TO THE COMPLAINT INFORMATION, DURING THE EVENT OF ADMINISTER RESCUE BREATHS TO A PATIENT IN DESTRESS, THE SPUR II DID NOT DEPRESS PROPERLY. IT WAS CHECHED THAT THE AMBU MASK WAS PROPERLY PLACED MULTIPLE TIMES, AS WELL AS CHECKING THE CONNECTION TO THE OXYGEN FLOWMETER. THE SPUR II WAS THROWN AWAY AND ANOTHER SPUR II WAS OPENED WITH THE SAME ISSUE. PER CUSTOMER: THERE DID NOT APPEAR TO BE ANY DEFORMITIES TO EITHER BAG. EVERYTHING APPEARED TO BE AS IT SHOULD BE. WHEN PLACED ON THE PATIENT'S FACE WE CHECKED SEVERAL TIMES FOR A GOOD SEAL BETWEEN AMBU MASK AND THE PATIENT, A GOOD SEAL WAS MADE, HOWEVER WHEN WE SQUEEZED THE BAG, WE COULD NOT PUSH AIR AT ALL. WE COULDN'T HEAR ANY AIR, NOR COULD WE SEE CHEST RISE AND FALL, THIS HAPPENED WITH BOTH BAGS. TOO MUCH TIME WAS SPENT TRYING TO GET IT TO WORK AND THEREFORE THE ANESTHESIA TEAM ASKED THAT WE DISCONTINUE USE AND ORDER A DIFFERENT BRAND. THE PATIENT DID STABILIZE DESPITE THE MALFUNCTION OF THE AMBU BAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20023061 ↗MalfunctionAug 21, 2024Mechanical Jam; Mechanical Jam
ACCORDING TO THE COMPLAINT INFORMATION, DURING THE EVENT OF ADMINISTER RESCUE BREATHS TO A PATIENT IN DISTRESS, THE SPUR II DID NOT DEPRESS PROPERLY. IT WAS CHECKED THAT THE AMBU MASK WAS PROPERLY PLACED MULTIPLE TIMES, AS WELL AS CHECKING THE CONNECTION TO THE OXYGEN FLOWMETER. THE SPUR II WAS THROWN AWAY AND ANOTHER SPUR II WAS OPENED WITH THE SAME ISSUE. PER CUSTOMER: THERE DID NOT APPEAR TO BE ANY DEFORMITIES TO EITHER BAG. EVERYTHING APPEARED TO BE AS IT SHOULD BE. WHEN PLACED ON THE PATIENT'S FACE WE CHECKED SEVERAL TIMES FOR A GOOD SEAL BETWEEN AMBU MASK AND THE PATIENT, A GOOD SEAL WAS MADE, HOWEVER WHEN WE SQUEEZED THE BAG, WE COULD NOT PUSH AIR AT ALL. WE COULDN'T HEAR ANY AIR, NOR COULD WE SEE CHEST RISE AND FALL, THIS HAPPENED WITH BOTH BAGS. TOO MUCH TIME WAS SPENT TRYING TO GET IT TO WORK AND THEREFORE THE ANESTHESIA TEAM ASKED THAT WE DISCONTINUE USE AND ORDER A DIFFERENT BRAND. THE PATIENT DID STABILIZE DESPITE THE MALFUNCTION OF THE AMBU BAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20020981 ↗
Other Cardiopulmonary resuscitation aid kit devices
- Ambu® SPUR® II Pediatric Resuscitator 531600051Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall10 reports
- AirLife™ 2K8004Airlife · Cardiopulmonary resuscitation aid kit1 recall5 reports
- AirLife™ 2K8005Airlife · Cardiopulmonary resuscitation aid kit1 recall1 death4 reports
- Ambu® SPUR® II Adult Resuscitator 523611051Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall3 reports
- AirLife™ 2K8017Airlife · Cardiopulmonary resuscitation aid kit1 recall1 death2 reports
- AirLife™ 2K8004FAirlife · Cardiopulmonary resuscitation aid kit1 recall1 report
- Ambu® SPUR® II Pediatric Resuscitator 530213031Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall0 reports
- Ambu® SPUR® II Pediatric Resuscitator 530613071Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page