Skip to content

Titan® Titan Touch (REF EN2816)Recalls & reported problems

Coloplast A/S · Titan Touch Narrow Base Infrapubic Zero Angle 16cm · Class 3 · Implant

All models of Titan (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
13
Deaths reported
0
Injuries reported
13
Recalls naming it
1

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 05708932072519), received between Aug 6, 2021 and Apr 8, 2026 — classified as 13 injuries.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

17 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

    Z-1361-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size4
  • Device Appears to Trigger Rejection2
  • Fluid/Blood Leak2
  • Fracture2
  • Malposition of Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Erosion2
  • Deformity/ Disfigurement1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 8, 2026Device Appears to Trigger Rejection

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS REMOVED AND REPLACED DUE TO EXTRUSION.

    Patient problems: Erosion

    MAUDE report 24815653 ↗
  • InjuryAug 29, 2025Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING FRACTURE AND SUBSEQUENT FLUID LOSS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22924104 ↗
  • InjuryJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 22412403 ↗
  • InjuryMay 28, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO INCORRECT SIZING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22100104 ↗
  • InjuryApr 3, 2025Device Appears to Trigger Rejection; Inadequacy of Device Shape and/or Size; Device Appears to Trigger Rejection; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO INCORRECT SIZE WHICH LED TO EROSION / EXTRUSION OF THE DEVICE.

    Patient problems: Erosion

    MAUDE report 21762501 ↗
  • InjuryMar 31, 2025Malposition of Device; Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED AS THE PATIENT HAD SST (GLANS PENIS DEFORMITY) AND THE TUBING WAS SUPERFICIAL.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 21734154 ↗
  • InjuryOct 10, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO SIZE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20417041 ↗
  • InjuryMay 16, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE PUMP WAS REMOVED AND REPLACED DUE TO A TUBING LENGTH ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19332494 ↗
  • InjuryApr 23, 2024Malposition of Device; Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED WITH A MALLEABLE DUE TO PAIN AND POSSIBLE CROSSOVER.

    Patient problems: Pain

    MAUDE report 19165627 ↗
  • InjuryJan 20, 2023Fluid/Blood Leak; Pumping Problem; Fluid/Blood Leak; Pumping Problem

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A LEAK. IT WAS REPORTED THAT THE PUMP WOULD PUMP FOR A FEW PUMPS AND THEN IT WOULD DIMPLE AND COULD NOT PUMP ANYMORE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16216419 ↗
  • InjurySep 26, 2022Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING TEAR. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15487327 ↗
  • InjuryDec 23, 2021Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REVISION DUE TO A HOLE IN A CYLINDER. THE RESERVOIR WAS RETAINED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13082795 ↗
  • InjuryAug 6, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE AVAILABLE INFORMATION, A REVISION PROCEDURE WAS PERFORMED TO REPOSITION THE RESERVOIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12290144 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis

…and 2 more.

Other Device, impotence, mechanical/hydraulic devices