1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Titan Touch hydraulic penile implant is used for erectile dysfunction. Its 105 reports often describe pump tubing fractures, fluid leaks, and implants that were removed and replaced after surgery, along with wound openings and infections. There were no deaths and 99 injuries. 1 open recall, started in March 2023, concerns a thinner pump wall that could cause earlier pump failure.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
105
Deaths reported
0
Injuries reported
99
Recalls naming it
1

Summary

The FDA's MAUDE database contains 105 adverse event reports that list this exact device (DI 05708932072564), received between Sep 6, 2018 and May 21, 2026 — classified as 99 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (27 reports).

1 FDA recall record name this device's identifier, 1 still open.

17 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 10 reportsJan 2024: 7 reportsFeb 2024: 3 reportsMar 2024: 6 reportsApr 2024: 8 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

    Z-1364-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem35
  • Fluid/Blood Leak27
  • Fracture7
  • Insufficient Device Problem Information7
  • Inadequacy of Device Shape and/or Size5
  • Malposition of Device5
  • Inflation Problem4
  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions65
  • Unspecified Infection26
  • Bacterial Infection4
  • Perforation3
  • Erosion2
  • Pain2
  • Wound Dehiscence2
  • Discomfort1

Recent adverse event reports

The 25 most recent of 105 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 21, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING FRACTURE FROM PUMP TO CYLINDERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241849 ↗
  • InjuryMay 19, 2026Material Puncture/Hole; Device Damaged by Another Device

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED WITH A GENESIS DUE TO THE CYLINDER BEING PUNCTURED IN PREVIOUS SURGERY DURING CLOSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216472 ↗
  • InjuryMar 25, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT (PUMP) WAS REMOVED AND REPLACED DUE TO INCORRECT SIZING. THE PUMP WAS PLACED TOO SHORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24689199 ↗
  • InjuryFeb 23, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE PUMP WAS REMOVED AND REPLACED DUE TO A PUMP TUBING FRACTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24418075 ↗
  • InjuryFeb 9, 2026Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO DEVICE NOT FULLY INFLATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24295446 ↗
  • InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED AS THE PATIENT HAD A HOLE IN THEIR SCROTUM [WOUND DEHISCENCE] AFTER THE SURGERY.

    Patient problems: Wound Dehiscence

    MAUDE report 24259093 ↗
  • MalfunctionDec 29, 2025Material Deformation; Material Deformation

    ACCORDING TO THE AVAILABLE INFORMATION THE DOCTOR IMPLANTED 16CM NARROW CYLINDERS WITH 2CM REAR TIP EXTENDERS IN A PATIENT THAT HAD PEYRONIE'S DISEASE WITH A CURVATURE OF 45 DEGREES. MANUAL MODELING WAS PERFORMED FOR 90 SECONDS. DOCTOR NOTICED A BUBBLE HERNIATION (ANEURYSM) ON THE CYLINDER FORMED ON THE PROXIMAL END NEAR THE CLEAR TIP. THE CYLINDER WAS REMOVED AND REPLACED WITH 18CM NARROW CYLINDERS. MANUAL MODELING WAS PERFORMED AGAIN AND AFTER 90 SECONDS, DOCTOR NOTICED A SIMILAR BUBBLE HERNIATION (ANEURYSM) ON THE 18CM NARROW CYLINDERS AT THE PROXIMAL END NEAR THE CLEAR TIP. THE CYLINDER WAS REMOVED AND REPLACED WITH 18CM STANDARD CYLINDERS. THESE WERE SUCCESSFULLY IMPLANTED. MANUAL MOLDING WAS ATTEMPTED WITH NO SUCCESS. THE DOCTOR PROCEEDED TO PERFORM A PLICATION PROCEDURE. THE CASE WAS COMPLETED FOLLOWING PLICATION PROCEDURE WITH NO FURTHER PROBLEMS. THIS REPORT CAPTURES THE 18CM NARROW CYLINDERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905871 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO IMPENDING EROSION ON THE RIGHT SIDE. THE LEFT SIDE WAS RETAINED AND ONLY THE RIGHT SIDE WAS REPLACED, WITH A NEW PUMP.

    Patient problems: Erosion

    MAUDE report 23696882 ↗
  • MalfunctionSep 23, 2025Material Deformation; Material Deformation

    ACCORDING TO THE AVAILABLE INFORMATION THERE WAS AN ANEURYSM ON BOTH NARROW CYLINDERS. DEVICE WAS TESTED TO INFLATE, AND IT STARTED AFTER MILD MODELING. ISSUE HAPPENED DURING IMPLANTATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23121836 ↗
  • InjuryAug 13, 2025Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22792890 ↗
  • InjuryAug 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A GENESIS AS A PLACEHOLDER DUE TO INFECTION. THE PLACEHOLDER WAS REMOVED, AND AN INFLATABLE PENILE PROSTHESIS WAS IMPLANTED AGAIN IN (B)(6) 2025.

    Patient problems: Unspecified Infection

    MAUDE report 22786693 ↗
  • InjuryJul 21, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE PUMP FREEZING UP. UPON EXPLANT, THE PUMP WAS NOTED TO BE WORKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22554249 ↗
  • MalfunctionJul 11, 2025Material Deformation; Material Deformation

    ACCORDING TO THE AVAILABLE INFORMATION, DURING IMPLANTATION THERE WAS AN ANEURYSM ON BOTH NARROW CYLINDERS ON TEST INFLATION. IT STARTED AFTER MILD MODELING. A NEW DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22485185 ↗
  • InjuryJun 23, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A PIN HOLE IN THE DISTAL CYLINDER. THE RESERVOIR REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22306885 ↗
  • InjuryJun 19, 2025Malposition of Device; Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE PUMP SITTING TOO POSTERIOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294134 ↗
  • MalfunctionApr 24, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, A NARROW 18CM REQUIRED STACKING REAR TIP EXTENDERS [RTES] 3CM + 1CM BECAUSE THE PATIENT MEASURED 22CM. DUE TO THE USAGE OF 4CM OF RTES, THE DOCTOR NEEDED TO CUT THE PUMP OFF AND RECONNECT BECAUSE THERE WAS TOO MUCH TUBING. THERE WERE NO ISSUES WITH THE PATIENT OR THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21901190 ↗
  • MalfunctionMar 20, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE DOCTOR USED A NARROW 18CM AND HAD TO STACK 3CM PLUS 1CM BECAUSE THE PATIENT MEASURED 22CM. BECAUSE THE DOCTOR USED 4CM OF RTE, HE NEEDED TO CUT THE PUMP OFF AND RECONNECT BECAUSE THERE WAS TOO MUCH TUBING. THERE WAS NO ISSUE WITH THE PATIENT OR THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652210 ↗
  • MalfunctionMar 5, 2025Failure to Pump; Failure to Pump

    ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT STATED THAT HIS PUMP STOPPED WORKING SOMETIME IN (B)(6) 2024. PATIENT IS SCHEDULED TO SEE A NEW UROLOGIST TO HAVE IT REPLACED ON (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21526185 ↗
  • InjuryJan 15, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED AS THERE WAS A TUBING FRACTURE FROM THE CYLINDER TO THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21162324 ↗
  • InjuryJan 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 21094170 ↗
  • InjuryDec 11, 2024Malposition of Device; Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO CROSSOVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20904837 ↗
  • InjuryNov 25, 2024Inadequacy of Device Shape and/or Size; Deformation Due to Compressive Stress; Inadequacy of Device Shape and/or Size; Deformation Due to Compressive Stress

    ACCORDING TO THE AVAILABLE INFORMATION, A REVISION WAS PERFORMED TO ADJUST THE PUMP AND SHORTEN THE TUBING ON ONE SIDE AS THE TUBING HAD GOTTEN KINKED. NOTHING WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20771460 ↗
  • InjuryOct 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED DUE TO EROSION ON AN UNKNOWN DATE AND LATER REPLACED.

    Patient problems: Erosion

    MAUDE report 20567422 ↗
  • InjuryOct 23, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20514322 ↗
  • InjuryOct 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 20391200 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis

…and 2 more.

Other Device, impotence, mechanical/hydraulic devices