Skip to content

Titan 90-9916SNCRecalls & reported problems

Coloplast A/S · Scrotal Bioflex® Cylinder Set with Pump - Narrow · Class 3 · Implant

All models of Titan (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 05708932323604), received between Jul 24, 2017 and Aug 6, 2021 — classified as 10 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Insufficient Device Problem Information2
  • Break1
  • Extrusion1
  • Fracture1
  • Inadequacy of Device Shape and/or Size1
  • Leak/Splash1
  • Material Erosion1

Patient problems

  • No Information3
  • No Known Impact Or Consequence To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • Erosion1
  • Failure of Implant1
  • Insufficient Information1
  • No Code Available1
  • Scar Tissue1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE STOPPED WORKING. IT WAS EXPLANTED AND REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 12288831 ↗
  • InjuryJun 30, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED DUE TO AN UNSPECIFIED FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12096542 ↗
  • InjuryMay 18, 2021Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS EXPLANTED DUE TO A LEAK IN THE TUBING NEAR THE PUMP. A NEW DEVICE WAS SUCCESSFULLY IMPLANTED AND NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11841121 ↗
  • MalfunctionFeb 8, 2021Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THOUGH NOT VERIFIED, RESERVOIR TUBING LEAK (TEAR)THE LOT # WAS REVIEWED FOR COMPLAINT TREND, NONCONFORMING REPORT AND CAPA. DEVICES MET SPECIFICATION PRIOR TO RELEASE AND NO TRENDS WERE NOTED.

    Patient problems: Failure of Implant

    MAUDE report 11291954 ↗
  • MalfunctionDec 30, 2020Migration; Migration

    ACCORDING TO THE AVAILABLE INFORMATION, THE RESERVOIR MIGRATED INTO THE SCROTUM AND THE RIGHT CYLINDER CAME OUT OF THE CORPORA. THE DEVICE WAS REMOVED AND REPLACED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 11096164 ↗
  • InjuryApr 19, 2019Mechanical Problem; Mechanical Problem

    ACCORDING TO THE AVAILABLE INFORMATION, MALFUNCTION.

    Patient problems: No Information

    MAUDE report 8533882 ↗
  • InjuryJan 12, 2018Material Erosion; Extrusion; Extrusion

    ACCORDING TO THE AVAILABLE INFORMATION, PATIENT EXPERIENCED DEVICE EROSION/EXTRUSION.

    Patient problems: Erosion

    MAUDE report 7186127 ↗
  • InjuryOct 26, 2017Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, A TUBING LEAK AT CYLINDER JUNCTION OCCURRED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6980844 ↗
  • InjurySep 21, 2017Material Frayed; Leak/Splash

    ACCORDING TO THE AVAILABLE INFORMATION, THE PUMP TUBING FRAYED AND LEAKS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6882873 ↗
  • InjuryJul 24, 2017Positioning Problem

    ACCORDING TO THE AVAILABLE INFORMATION, REPOSITION.

    Patient problems: No Information

    MAUDE report 6739115 ↗
  • InjuryJul 24, 2017Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, INCORRECT SIZE.

    Patient problems: No Code Available; No Information

    MAUDE report 6739028 ↗
  • InjuryJul 24, 2017Break

    ACCORDING TO THE AVAILABLE INFORMATION, BROKEN TUBING AND SCAR TISSUE.

    Patient problems: Scar Tissue

    MAUDE report 6737726 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis

…and 3 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page