Titan 90-9918SNCRecalls & reported problems
Coloplast A/S · Scrotal Bioflex® Cylinder Set with Pump - Narrow · Class 3 · Implant
All models of Titan (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 05708932323901), received between Jul 24, 2017 and Mar 3, 2026 — classified as 16 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (7 reports).
18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture7
- Fluid/Blood Leak2
- Insufficient Device Problem Information2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Device Inoperable1
- Disconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Erosion2
- No Information2
- Insufficient Information1
- No Code Available1
- No Consequences Or Impact To Patient1
- Unspecified Infection1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2026Fracture
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO CRACKED TUBING. IT WAS NOTED THAT THE DEVICE WAS 8 YEARS OLD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24497457 ↗InjuryJan 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE AVAILABLE INFORMATION, THE LEFT CYLINDER WAS REMOVED AS THE IMPLANT HAD IMPENDING EROSION ON THE LEFT CYLINDER.
Patient problems: Erosion
MAUDE report 21137427 ↗InjuryNov 25, 2024Fluid/Blood Leak; Inflation Problem; Insufficient Device Problem Information; Fluid/Blood Leak; Inflation Problem; Insufficient Device Problem Information
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNKNOWN REASON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20772039 ↗InjuryOct 18, 2024Fracture; Fracture
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A PUMP TUBING BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20487498 ↗InjuryJul 8, 2024Fracture; Fracture
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19690225 ↗InjuryMay 22, 2023Fracture; Fracture
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING BREAK. THERE WAS A BREAK IN THE TUBING OFF THE PUMP. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16981481 ↗InjuryJan 27, 2023Fracture; Fracture
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING TEAR. A 1 CM TEAR ALONG THE TUBING WAS NOTED AT THE RIGHT CYLINDER AT THE POINT WHERE THE EXIT TUBING WAS COMING FROM THE TUNICA. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16255379 ↗InjuryOct 7, 2022Fracture; Fracture
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING BREAK. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15565269 ↗InjuryDec 28, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A MALFUNCTION. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13101376 ↗InjuryNov 30, 2021Fracture; Fracture
AS REPORTED TO COLOPLAST, THOUGH NOT VERIFIED, TUBING BREAK NEAR THE PUMP. THE PUMP AND TUBING WERE REPLACED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12900622 ↗InjuryAug 18, 2021Fluid/Blood Leak; Fluid/Blood Leak
ACCORDING TO THE AVAILABLE INFORMATION, A TUBING LEAK WAS NOTED ON THE CYLINDER TO PUMP TUBING. THE DEVICE WAS EXPLANTED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12337439 ↗InjuryJul 2, 2021Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS EXPLANTED AND REPLACED FOR A REAR TIP EXCHANGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12109070 ↗MalfunctionOct 18, 2020Disconnection; Disconnection
THIS FOLLOW-UP MDR IS CREATED TO DOCUMENT THE ADDITIONAL EVENT INFORMATION RECEIVED FOR RECORD #615730. ACCORDING TO THE AVAILABLE INFORMATION THIS INFLATABLE DEVICE WAS IMPLANTED IN 2013 AND REMOVED/REPLACED ON (B)(6) 2020. INFORMATION RECEIVED FROM THE CLINICAL SALES REPRESENTATIVE INDICATED: ¿COLLARS ON RESERVOIR TUBING WAS DISCONNECTED. MINIMAL SALINE IN THE SYSTEM.¿ A TITAN PUMP, TWO CYLINDERS, AND RESERVOIR WERE RECEIVED FOR EVALUATION. EXAMINATION AND TESTING OF THE RETURNED COMPONENTS REVEALED A SEPARATION IN THE BLADDER OF CYLINDER 1. TESTING REVEALED THIS TO BE A SITE OF LEAKAGE. MICROSCOPIC EXAMINATION OF THE SURFACES REVEALED THEM TO HAVE BURN LIKE MARKS, INDICATING CONTACT WAS MADE WITH CAUTERIZING INSTRUMENTATION. PARTIAL SEPARATIONS WITHIN ABRASION WERE NOTED ON BOTH PUMP EXHAUST TUBES, THE EXHAUST TUBING OF CYLINDER 1, AND THE RESERVOIR INLET TUBING. TESTING REVEALED THESE TO NOT BE SITES OF LEAKAGE. SURFACE ABRASION WAS NOTED ON THE PUMP INLET TUBING. BURN LIKE MARKS WERE NOTED ON THE BLADDER OF CYLINDER 2. TESTING REVEALED THIS TO NOT BE A SITE OF LEAKAGE. NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE PUMP, CYLINDER 2, OR RESERVOIR. BECAUSE THESE COMPONENTS WERE
Patient problems: Insufficient Information
MAUDE report 10695075 ↗MalfunctionJun 10, 2020Break; Leak/Splash; Break; Leak/Splash
ACCORDING TO THE AVAILABLE INFORMATION, THERE WAS LEAKAGE DUE TO A TUBING BREAK NEAR THE RESERVOIR. THE RESERVOIR WAS REMOVED/REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10139253 ↗InjuryApr 18, 2019Material Puncture/Hole; Material Puncture/Hole
ACCORDING TO THE AVAILABLE INFORMATION, MALFUNCTION - CYLINDERS PERFORATED.
Patient problems: No Information
MAUDE report 8529724 ↗InjuryDec 4, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
ACCORDING TO THE AVAILABLE INFORMATION, INFECTION.
Patient problems: Unspecified Infection
MAUDE report 8130968 ↗InjuryMar 20, 2018Device Inoperable
ACCORDING TO THE AVAILABLE INFORMATION, PATIENT WITH A HISTORY OF ERECTILE DYSFUNCTION AND PEYRONIE'S DISEASE UNDERWENT PENILE IMPLANT IN 2017. THE PENILE IMPLANT IS NOT FUNCTIONING AND THE DEVICE WAS EXPLANTED.
Patient problems: No Code Available; No Information
MAUDE report 7352919 ↗InjuryJul 24, 2017Extrusion
ACCORDING TO THE AVAILABLE INFORMATION: EROSION/EXTRUSION, NON-MALFUNCTION ISSUE. PATIENT PREFERENCE.
Patient problems: Erosion
MAUDE report 6737400 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.
- Z-1367-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
- Z-1374-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
- Z-1368-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
- Z-1372-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
- Z-1365-2023 · Mar 1, 2023 · Open, ClassifiedTitan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
- Z-1363-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
- Z-1360-2023 · Mar 1, 2023 · Open, ClassifiedTitan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
- Z-1370-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
- Z-1371-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
- Z-1369-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
- Z-1375-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
- Z-1362-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
- Z-1366-2023 · Mar 1, 2023 · Open, ClassifiedTitan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
- Z-1373-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
- Z-1359-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
…and 3 more.
Other Device, Impotence, Mechanical/Hydraulic devices
- Titan ES2920Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall862 reports
- Titan ES2922Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall1 death758 reports
- Titan ES2918Coloplast A/S · Device, Impotence, Mechanical/Hydraulic577 reports
- Titan ES8920Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall476 reports
- Titan ES8922Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall334 reports
- Titan 517750Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall258 reports
- Titan ES8918Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall225 reports
- Titan ES2818Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page