Coloplast A/S · Scrotal Bioflex® Cylinder Set with Pump - Narrow · Class 3 · Implant

All models of Titan (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 05708932323901), received between Jul 24, 2017 and Mar 3, 2026 — classified as 16 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (7 reports).

18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture7
  • Fluid/Blood Leak2
  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Device Inoperable1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Erosion2
  • No Information2
  • Insufficient Information1
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO CRACKED TUBING. IT WAS NOTED THAT THE DEVICE WAS 8 YEARS OLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24497457 ↗
  • InjuryJan 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE LEFT CYLINDER WAS REMOVED AS THE IMPLANT HAD IMPENDING EROSION ON THE LEFT CYLINDER.

    Patient problems: Erosion

    MAUDE report 21137427 ↗
  • InjuryNov 25, 2024Fluid/Blood Leak; Inflation Problem; Insufficient Device Problem Information; Fluid/Blood Leak; Inflation Problem; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20772039 ↗
  • InjuryOct 18, 2024Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A PUMP TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20487498 ↗
  • InjuryJul 8, 2024Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19690225 ↗
  • InjuryMay 22, 2023Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING BREAK. THERE WAS A BREAK IN THE TUBING OFF THE PUMP. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16981481 ↗
  • InjuryJan 27, 2023Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING TEAR. A 1 CM TEAR ALONG THE TUBING WAS NOTED AT THE RIGHT CYLINDER AT THE POINT WHERE THE EXIT TUBING WAS COMING FROM THE TUNICA. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16255379 ↗
  • InjuryOct 7, 2022Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING BREAK. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15565269 ↗
  • InjuryDec 28, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A MALFUNCTION. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13101376 ↗
  • InjuryNov 30, 2021Fracture; Fracture

    AS REPORTED TO COLOPLAST, THOUGH NOT VERIFIED, TUBING BREAK NEAR THE PUMP. THE PUMP AND TUBING WERE REPLACED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12900622 ↗
  • InjuryAug 18, 2021Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, A TUBING LEAK WAS NOTED ON THE CYLINDER TO PUMP TUBING. THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12337439 ↗
  • InjuryJul 2, 2021Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS EXPLANTED AND REPLACED FOR A REAR TIP EXCHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12109070 ↗
  • MalfunctionOct 18, 2020Disconnection; Disconnection

    THIS FOLLOW-UP MDR IS CREATED TO DOCUMENT THE ADDITIONAL EVENT INFORMATION RECEIVED FOR RECORD #615730. ACCORDING TO THE AVAILABLE INFORMATION THIS INFLATABLE DEVICE WAS IMPLANTED IN 2013 AND REMOVED/REPLACED ON (B)(6) 2020. INFORMATION RECEIVED FROM THE CLINICAL SALES REPRESENTATIVE INDICATED: ¿COLLARS ON RESERVOIR TUBING WAS DISCONNECTED. MINIMAL SALINE IN THE SYSTEM.¿ A TITAN PUMP, TWO CYLINDERS, AND RESERVOIR WERE RECEIVED FOR EVALUATION. EXAMINATION AND TESTING OF THE RETURNED COMPONENTS REVEALED A SEPARATION IN THE BLADDER OF CYLINDER 1. TESTING REVEALED THIS TO BE A SITE OF LEAKAGE. MICROSCOPIC EXAMINATION OF THE SURFACES REVEALED THEM TO HAVE BURN LIKE MARKS, INDICATING CONTACT WAS MADE WITH CAUTERIZING INSTRUMENTATION. PARTIAL SEPARATIONS WITHIN ABRASION WERE NOTED ON BOTH PUMP EXHAUST TUBES, THE EXHAUST TUBING OF CYLINDER 1, AND THE RESERVOIR INLET TUBING. TESTING REVEALED THESE TO NOT BE SITES OF LEAKAGE. SURFACE ABRASION WAS NOTED ON THE PUMP INLET TUBING. BURN LIKE MARKS WERE NOTED ON THE BLADDER OF CYLINDER 2. TESTING REVEALED THIS TO NOT BE A SITE OF LEAKAGE. NO FUNCTIONAL ABNORMALITIES WERE NOTED WITH THE PUMP, CYLINDER 2, OR RESERVOIR. BECAUSE THESE COMPONENTS WERE

    Patient problems: Insufficient Information

    MAUDE report 10695075 ↗
  • MalfunctionJun 10, 2020Break; Leak/Splash; Break; Leak/Splash

    ACCORDING TO THE AVAILABLE INFORMATION, THERE WAS LEAKAGE DUE TO A TUBING BREAK NEAR THE RESERVOIR. THE RESERVOIR WAS REMOVED/REPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10139253 ↗
  • InjuryApr 18, 2019Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE AVAILABLE INFORMATION, MALFUNCTION - CYLINDERS PERFORATED.

    Patient problems: No Information

    MAUDE report 8529724 ↗
  • InjuryDec 4, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 8130968 ↗
  • InjuryMar 20, 2018Device Inoperable

    ACCORDING TO THE AVAILABLE INFORMATION, PATIENT WITH A HISTORY OF ERECTILE DYSFUNCTION AND PEYRONIE'S DISEASE UNDERWENT PENILE IMPLANT IN 2017. THE PENILE IMPLANT IS NOT FUNCTIONING AND THE DEVICE WAS EXPLANTED.

    Patient problems: No Code Available; No Information

    MAUDE report 7352919 ↗
  • InjuryJul 24, 2017Extrusion

    ACCORDING TO THE AVAILABLE INFORMATION: EROSION/EXTRUSION, NON-MALFUNCTION ISSUE. PATIENT PREFERENCE.

    Patient problems: Erosion

    MAUDE report 6737400 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis

…and 3 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →