Skip to content

Titan ER8075Recalls & reported problems

Coloplast A/S · CL Reservoir · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Coloplast CL Reservoir is the implanted fluid reservoir used in a penile implant for impotence. Most of the 67 reports describe the reservoir shifting out of place or being pushed out of position, sometimes from weight loss, along with leaking fluid or blood, and pain or erosion where it sits; many also mention the reservoir filling or emptying on its own. 57 reports describe injuries and no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
66
Deaths reported
0
Injuries reported
56
Recalls naming it
0

Summary

The FDA's MAUDE database contains 66 adverse event reports that list this exact device (DI 05708932473910), received between Jul 19, 2017 and Jun 9, 2026 — classified as 56 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Migration” (10 reports).

18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 4 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 4 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration10
  • Insufficient Device Problem Information9
  • Fluid/Blood Leak8
  • Adverse Event Without Identified Device or Use Problem5
  • Migration or Expulsion of Device4
  • Self-Activation or Keying4
  • Device Appears to Trigger Rejection3
  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions37
  • No Information10
  • Pain7
  • Erosion4
  • No Code Available2
  • No Consequences Or Impact To Patient2
  • Bacterial Infection1
  • Capsular Contracture1

Recent adverse event reports

The 25 most recent of 66 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2026Device Appears to Trigger Rejection

    ACCORDING TO THE AVAILABLE INFORMATION, RESERVOIR REMOVED AND REPLACED DUE TO EXTRUSION.

    Patient problems: Erosion

    MAUDE report 25444959 ↗
  • InjuryMay 5, 2026Migration

    ACCORDING TO THE AVAILABLE INFORMATION, RESERVOIR REMOVED AND REPLACED DUE TO LARGE WEIGHT LOSS DISPLACING THE RESERVOIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25082226 ↗
  • InjuryApr 22, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT RESERVOIR WAS REMOVED AND REPLACED DUE TO UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957597 ↗
  • InjuryFeb 26, 2026Self-Activation or Keying

    ACCORDING TO THE AVAILABLE INFORMATION, THE RESERVOIR WAS REMOVED AND REPLACED DUE TO AUTO INFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24457534 ↗
  • MalfunctionJan 14, 2026Material Separation

    ACCORDING TO THE AVAILABLE INFORMATION, DURING REPLACEMENT OF THE DEVICE FOR AN ISSUE RELATED TO THE CYLINDER SET/ TUBING, A MALFUNCTION OF THE RESERVOIR WAS NOTED. THE GLUE SEEMINGLY FAILED. A PICTURE OF THE DEVICE SHOWS THE LOCKOUT VALVE SEPARATED FROM THE RESERVOIR BLADDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24072242 ↗
  • InjuryJan 6, 2026Unintended Deflation

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AUTO DEFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23976084 ↗
  • InjuryDec 30, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO EXTRUSION.

    Patient problems: Erosion

    MAUDE report 23928857 ↗
  • InjuryDec 17, 2025Migration; Migration

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A HERNIATED RESERVOIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23831472 ↗
  • InjuryOct 28, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    ACCORDING TO THE AVAILABLE INFORMATION THE RESERVOIR WAS EXPLANTED AND REPLACED DUE TO EXTRUSION. THE RESERVOIR WAS IN WORKING CONDITION, BUT POOR ANATOMICAL POSITION.

    Patient problems: Erosion

    MAUDE report 23406491 ↗
  • InjurySep 15, 2025Self-Activation or Keying; Self-Activation or Keying

    ACCORDING TO THE AVAILABLE INFORMATION, THE RESERVOIR WAS EXPLANTED AND REPLACED DUE TO AUTO INFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23058900 ↗
  • InjurySep 12, 2025Disconnection; Disconnection

    ACCORDING TO THE AVAILABLE INFORMATION, THE RESERVOIR WAS EXPLANTED AND REPLACED DUE TO THE RESERVOIR NOT BEING CONNECTED AT THE WHITE CONNECTION JUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23039631 ↗
  • InjurySep 10, 2025Migration; Migration

    ACCORDING TO THE AVAILABLE INFORMATION THE PUMP AND RESERVOIR WERE EXPLANTED AND REPLACED DUE TO THE RESERVOIR MIGRATING INTO THE SCROTUM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23021032 ↗
  • InjurySep 2, 2025Self-Activation or Keying; Self-Activation or Keying

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO AUTO INFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22937662 ↗
  • InjuryJun 23, 2025Migration; Migration

    ACCORDING TO THE AVAILABLE INFORMATION, THE RESERVOIR WAS EXPLANTED AND REPLACED DUE TO IT MIGRATING FROM ITS ORIGINAL PLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22311396 ↗
  • InjuryMay 13, 2025Migration; Migration

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS REMOVED AND REPLACED DUE TO RESERVOIR HERNIATION [MIGRATION].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22013254 ↗
  • MalfunctionMay 1, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE LABEL ON THE INNER PACKAGE OF THE KIT CLAIMS THAT THE DEVICE IS A '125 CC TITAN CL RESERVOIR', BUT THE DEVICE IS A 75 CC RESERVOIR. ALL OF THE LOT NUMBERS ON THE BOX, INNER PACKAGING, AND DEVICE MATCH AND REPRESENT A 75CC TITAN CL RESERVOIR DEVICE, BUT THE LABEL ON THE INNER PACKAGE INCORRECTLY STATES THAT IT IS THE LARGER TITAN RESERVOIR (125CC). THE DEVICE WAS NOT IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941755 ↗
  • InjuryMar 26, 2025Migration; Migration

    ACCORDING TO THE AVAILABLE INFORMATION, DURING THIS REVISION, THE RESERVOIR WAS REPLACED AND REPOSITIONED AS IT HAD MIGRATED TOO LOW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21700424 ↗
  • MalfunctionFeb 20, 2025Fluid/Blood Leak; Air/Gas in Device; Fluid/Blood Leak; Air/Gas in Device

    ACCORDING TO THE AVAILABLE INFORMATION, DURING PREP, ALL OF THE AIR DID NOT COME OUT OF THE RESERVOIR AND THE PHYSICIAN SUSPECTED THERE WAS A LEAK. THE RESERVOIR WAS NOT IMPLANTED AND A DIFFERENT 75 CC RESERVOIR WAS PLACED INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21427341 ↗
  • InjuryOct 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE RESERVOIR WAS REMOVED AND REPLACED DUE TO A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20398658 ↗
  • InjuryJul 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO LOWER ABDOMEN PAIN.

    Patient problems: Pain

    MAUDE report 19793721 ↗
  • InjuryJul 5, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS REMOVED AND REPLACED DUE TO A RESERVOIR TUBING TEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19680755 ↗
  • InjuryJun 18, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO TUBING LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19562809 ↗
  • MalfunctionMay 14, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, A LEAK IN THE RESERVOIR WAS FOUND DURING PRE-SURGICAL ASSEMBLY AND TESTING OF THE PROSTHESIS COMPONENTS. THE NURSE REPLACED THE RESERVOIR, AND THE SURGERY WAS CONTINUED WITHOUT COMPLICATIONS. NO HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19311749 ↗
  • InjuryMay 6, 2024Malposition of Device; Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION, REVISION SURGERY WAS PERFORMED TO RE-POSITION THE SUBMUSCULAR RESERVOIR DEEPER AND REMOVE SOME OF THE FLUID (10-20ML) FROM THE RESERVOIR. THE TUBING WAS CUT AND RECONNECTED IN ORDER TO DO THIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19248172 ↗
  • InjuryApr 30, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS REMOVED AND REPLACED DUE TO A TUBING LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19213285 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis

…and 3 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page