Skip to content

Titan ES8916Recalls & reported problems

Coloplast A/S · Scrotal Zero Degree Angle Cylinder Set with Pump · Class 3 · Implant

All models of Titan (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Titan ES8916 is a Coloplast scrotal cylinder set with a pump, used as a penile implant for impotence. Most of the 68 reports describe the tubing cracking or breaking, which led to the implant being taken out and replaced, or the cylinder being the wrong size or too short. Others mention erosion and infection with tubing exposed through the scrotum, leaking fluid, and pain. There were 60 injuries and no deaths.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
67
Deaths reported
0
Injuries reported
59
Recalls naming it
0

Summary

The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 05708932487320), received between Jul 24, 2017 and May 21, 2026 — classified as 59 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Fracture” (14 reports).

18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture14
  • Adverse Event Without Identified Device or Use Problem10
  • Insufficient Device Problem Information8
  • Inadequacy of Device Shape and/or Size7
  • Break6
  • Fluid/Blood Leak6
  • Deflation Problem3
  • Inflation Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • No Information16
  • Unspecified Infection11
  • Erosion7
  • No Known Impact Or Consequence To Patient4
  • Pain4
  • Bacterial Infection1
  • Deformity/ Disfigurement1

Recent adverse event reports

The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 21, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO A BREAK IN TUBING BY THE RESERVOIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241876 ↗
  • InjuryMay 21, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANTS WERE REMOVED AND REPLACED DUE TO INCORRECT SIZING. THE PHYSICIAN ONLY WANTED TO REPLACE THE CYLINDERS SO CUT THEM IN. THE RESERVOIR WAS NOT REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241105 ↗
  • InjuryApr 17, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A FRACTURE IN THE LEFT TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912742 ↗
  • InjuryApr 11, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS REMOVED DUE TO ERODED AND INFECTED. IT WAS NOTED THAT THE TUBING WAS EXPOSED FROM THE SCROTUM.

    Patient problems: Erosion; Unspecified Infection

    MAUDE report 24852057 ↗
  • InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO EROSION.

    Patient problems: Erosion

    MAUDE report 24661278 ↗
  • InjuryFeb 5, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING FRACTURE MALFUNCTION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24273436 ↗
  • InjuryDec 31, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE IMPLANT BEING TOO SHORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23937182 ↗
  • InjuryNov 3, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A CRACK IN THE TUBING NEAR THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23450812 ↗
  • InjuryJun 2, 2025Malposition of Device; Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A CYLINDER CROSSOVER/EROSION. THE CYLINDER WAS OUTSIDE THE CORPORA.

    Patient problems: Erosion

    MAUDE report 22124533 ↗
  • InjuryMay 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22110043 ↗
  • InjuryMay 1, 2025Deflation Problem; Inflation Problem; Inadequacy of Device Shape and/or Size; Physical Resistance/Sticking; Deflation Problem; Inflation Problem; Inadequacy of Device Shape and/or Size; Physical Resistance/Sticking

    ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT'S PRIVATE PARTS ARE VERY SENSITIVE TO TOUCH. INFLATING/ DEFLATING THE PUMP OF THE DEVICE IS VERY HARD AND THE PUMP IS NOT EASY TO SQUEEZE. THE HEAD OF THE PATIENT¿S PENIS IS DANGLING DOWN AND HE HAS LOST 2 INCHES FROM HIS PENIS. THE PATIENT IS CONCERNED THAT THE IMPLANT IS NOT SIZED RIGHT.

    Patient problems: Deformity/ Disfigurement; Discomfort

    MAUDE report 21943776 ↗
  • InjuryApr 30, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS MALFUNCTIONING. THE DEVICE WAS LATER EXPLANTED AND REPLACED DUE TO A TUBING BREAK FROM THE LEFT CYLINDER TO PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21934893 ↗
  • InjuryApr 24, 2025Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED AS THE PUMP WAS LEAKING, WHICH WAS CAUSED BY A TUBING TEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21904322 ↗
  • InjuryFeb 20, 2025Fracture; Air/Gas in Device; Fracture; Air/Gas in Device

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO AIR IN SYSTEM CAUSED BY TEAR ON CYLINDER TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21426827 ↗
  • InjuryNov 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED DUE TO INFECTION. INTRAVENOUS ANTIBIOTICS WERE PROVIDED.

    Patient problems: Unspecified Infection

    MAUDE report 20612652 ↗
  • InjuryOct 30, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20568803 ↗
  • InjuryMar 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO SCROTAL INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 18851952 ↗
  • InjuryDec 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE BEGAN LEAKING AND WAS NO LONGER OPERATIONAL. NO ADVERSE EFFECTS TO THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18324503 ↗
  • InjuryNov 28, 2023Break; Break

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO BREAK IN TUBING RIGHT AT THE INPUT JUNCTION ON PROXIMAL PORTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18220871 ↗
  • InjuryMay 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED EXPLANTATION DUE TO FLUID LOSS. A MALLEABLE DEVICE WAS IMPLANTED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16992155 ↗
  • InjuryApr 23, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO SIZE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16792767 ↗
  • InjuryApr 19, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO AN UNSPECIFIED REASON. THE RESERVOIR WAS RETAINED AND REUSED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16773994 ↗
  • InjuryMar 30, 2023Material Puncture/Hole; Inadequacy of Device Shape and/or Size; Material Puncture/Hole; Inadequacy of Device Shape and/or Size

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A HOLE. AN ANEURYSM WAS ALSO NOTED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16653076 ↗
  • InjuryMar 28, 2023Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING BREAK. THE RIGHT CYLINDER TUBING WAS RUPTURED. THE RESERVOIR WAS RETAINED AND REUSED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16625354 ↗
  • InjuryFeb 27, 2023Fracture; Pumping Problem; Fracture; Pumping Problem

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO NOT PUMPING. A FRACTURE WAS ALSO NOTED NEAR THE PUMP. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16448440 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis

…and 3 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page