Titan ESR916Recalls & reported problems

Coloplast A/S · Scrotal OTR Zero Degree Angle Cylinder Set with Pump · Class 3 · Implant

All models of Titan (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 05708932487641), received between Jul 24, 2017 and Nov 11, 2025 — classified as 20 injuries.

The most frequently reported specific device problem is “Fracture” (5 reports).

18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Inflation Problem4
  • Fluid/Blood Leak3
  • Insufficient Device Problem Information2
  • Break1
  • Crack1
  • Failure to Pump1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Information6
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO WITH A MALLEABLE DUE TO A TUBING BREAK BY THE CYLINDER AND PATIENT PREFERENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23522637 ↗
  • InjurySep 9, 2025Inflation Problem; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS NOT INFLATING. THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23009752 ↗
  • InjuryApr 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND NOT REPLACED DUE TO NON-USE UNSPECIFIED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21762173 ↗
  • InjuryJul 15, 2024Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A MALFUNCTION OF THE TUBING TO THE LEFT CYLINDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19741540 ↗
  • InjuryMar 21, 2024Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING BREAK BY THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18950808 ↗
  • InjuryFeb 12, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18691739 ↗
  • InjuryDec 26, 2023Premature Activation; Premature Activation

    ACCORDING TO THE AVAILABLE INFORMATION THERE WAS A FAILURE IN THE VALVE IN THAT IT ACTIVATED ITSELF WITHOUT MANIPULATION OF THE PATIENT AND THE CYLINDERS WERE FILLED. THE PRODUCT WAS REMOVED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18393532 ↗
  • InjuryNov 16, 2023Fluid/Blood Leak; Inflation Problem; Fluid/Blood Leak; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO UNABLE TO INFLATE THE PROSTHESIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18152806 ↗
  • InjuryJun 6, 2023Failure to Pump; Failure to Pump

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A PUMP FAILURE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17075874 ↗
  • InjuryJun 5, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO FLUID LOSS. THERE WAS A HOLE IN THE TUBING. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17059995 ↗
  • InjuryAug 29, 2022Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED EXPLANTATION DUE TO A TUBING BREAK. THE BREAK WAS AT THE PUMP. A MALLEABLE DEVICE WAS IMPLANTED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15316696 ↗
  • InjuryMar 16, 2022Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS EXPLANTED AND REPLACED DUE TO TUBING LEADING FROM THE RESERVOIR BROKEN IN MULTIPLE LOCATIONS. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13783178 ↗
  • InjuryFeb 24, 2022Material Protrusion/Extrusion; Material Protrusion/Extrusion

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS EXPLANTED AND REPLACED BECAUSE THE RIGHT SIDE CYLINDER EXTRUDED THROUGH THE DISTAL URETHRA. A MALLEABLE DEVICE WAS IMPLANTED ON THE LEFT SIDE ONLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593865 ↗
  • InjuryJun 8, 2021Fluid/Blood Leak; Pumping Problem; Fluid/Blood Leak; Pumping Problem

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS EXPLANTED AND REPLACED DUE TO A POTENTIAL LEAD IN THE PUMP. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11958748 ↗
  • InjuryApr 17, 2019Inflation Problem; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, PUMP WOULD NOT INFLATE CYLINDERS.

    Patient problems: No Information

    MAUDE report 8524334 ↗
  • InjuryApr 2, 2019Inflation Problem; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT STOPPED INFLATING.

    Patient problems: No Information

    MAUDE report 8473256 ↗
  • InjuryAug 22, 2018Leak/Splash; Leak/Splash

    ACCORDING TO THE AVAILABLE INFORMATION, THE PHYSICIAN REMOVED AN OTR DEVICE AND REPLACED ALL COMPONENTS. THERE WAS A FLUID LOSS OF UNKNOWN CAUSE IN ORIGINAL DEVICE.

    Patient problems: No Information

    MAUDE report 7804412 ↗
  • InjuryApr 13, 2018Break

    ACCORDING TO THE AVAILABLE INFORMATION, RESERVOIR TUBING BREAK 2" ABOVE THE REINFORCED TUBING.

    Patient problems: No Information

    MAUDE report 7428376 ↗
  • InjuryNov 14, 2017Material Rupture; Torn Material; Material Rupture

    ACCORDING TO THE AVAILABLE INFORMATION TITAN OTR WAS EXPLANTED DUE TO TORN TUBING NEAR PUMP.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 7029309 ↗
  • InjuryJul 24, 2017Crack

    ACCORDING TO AVAILABLE INFORMATION, TUBING CRACKED FROM PUMP.

    Patient problems: No Information

    MAUDE report 6736188 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis

…and 3 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page