This Coloplast Titan ESR920 scrotal cylinder implant set is a penile prosthesis used for impotence. The 76 reports mostly describe the implant's tubing cracking, fracturing or breaking, and fluid leaking out, sometimes with erosion of surrounding tissue; 70 reports involved injuries and no deaths. In many cases the implant was removed and replaced because of the broken tubing. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
76
Deaths reported
0
Injuries reported
70
Recalls naming it
0

Summary

The FDA's MAUDE database contains 76 adverse event reports that list this exact device (DI 05708932487689), received between Jul 19, 2017 and Jun 26, 2026 — classified as 70 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Fracture” (18 reports).

18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture18
  • Break12
  • Fluid/Blood Leak12
  • Leak/Splash9
  • Insufficient Device Problem Information6
  • Inflation Problem4
  • Adverse Event Without Identified Device or Use Problem3
  • Material Erosion3

Patient problems

  • No Clinical Signs, Symptoms or Conditions35
  • No Information25
  • No Known Impact Or Consequence To Patient5
  • Erosion3
  • No Code Available3
  • No Consequences Or Impact To Patient3
  • Unspecified Infection2
  • Bacterial Infection1

Recent adverse event reports

The 25 most recent of 76 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO A TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25635594 ↗
  • InjuryMar 25, 2026Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE TITAN OTR IMPLANT WAS REMOVED AND REPLACED DUE TO LOSS OF FLUID IN IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24689525 ↗
  • InjuryFeb 2, 2026Fluid/Blood Leak; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS REMOVED AND REPLACED DUE TO BROKEN TUBING MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24233592 ↗
  • InjuryDec 23, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE TUBING CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23879791 ↗
  • InjurySep 3, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING FRACTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951606 ↗
  • InjuryAug 6, 2025Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO LOSS OF FLUID.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22733015 ↗
  • InjuryApr 3, 2025Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO LOSS OF FLUID DUE TO TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21762296 ↗
  • InjuryFeb 20, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A MALFUNCTION IN THE CLEAR TUBING OF THE PROSTHETIC, WHICH IS ASSUMED TO BE A TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21422009 ↗
  • InjuryFeb 19, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21413950 ↗
  • InjuryOct 7, 2024Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A HOLE IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20384543 ↗
  • InjuryAug 26, 2024Inflation Problem; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE IMPLANT WAS NOT FULLY INFLATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20066203 ↗
  • InjuryAug 8, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE TITAN OTR IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNSPECIFIED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19942736 ↗
  • InjuryApr 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO EROSION/EXTRUSION.

    Patient problems: Erosion

    MAUDE report 19013039 ↗
  • InjuryNov 28, 2023Break; Break

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18218338 ↗
  • InjuryNov 20, 2023Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A LEAK IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18174159 ↗
  • InjuryNov 16, 2023Fluid/Blood Leak; Inflation Problem; Fluid/Blood Leak; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO NO LONGER BEING ABLE TO INFLATE THE PROSTHESIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18153933 ↗
  • InjuryNov 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO TUBING LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18124896 ↗
  • InjuryNov 9, 2023Material Erosion; Material Erosion

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS REMOVED AND REPLACED DUE TO TUBE EROSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18101139 ↗
  • InjuryJun 2, 2023Inflation Problem; Inflation Problem

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO NOT INFLATING. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17056584 ↗
  • InjuryMay 1, 2023Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING FRACTURE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16844964 ↗
  • InjuryApr 19, 2023Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING FRACTURE. THE FRACTURE WAS FROM THE PUMP TO THE CYLINDERS. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16772560 ↗
  • InjuryApr 18, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED EXPLANTATION DUE TO AN UNKNOWN REASON. A MALLEABLE DEVICE WAS IMPLANTED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16760840 ↗
  • InjuryFeb 24, 2023Pumping Problem; Appropriate Device Problem Term/Code Not Available; Pumping Problem; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO PATIENT PREFERENCE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16440510 ↗
  • InjuryFeb 10, 2023Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO FLUID LOSS. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16353431 ↗
  • InjuryJan 27, 2023Fracture; Fracture

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A TUBING FRACTURE. THE FRACTURE WAS AT THE PUMP. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16253549 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis

…and 3 more.

Other Device, Impotence, Mechanical/Hydraulic devices