Titan ES2820Recalls & reported problems
Coloplast A/S · Titan® Touch Infrapubic Zero Degree Angle Cylinder Set with Pump · Class 3 · Implant
Quick summary
This Coloplast Titan Touch penile implant has 232 adverse event reports, including 204 injuries and no deaths. The reports mostly describe leaks, pumping and inflation problems, and implants that were the wrong size, with some noting erosion. Unspecified infection is listed in 32 reports. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 232
- Deaths reported
- 0
- Injuries reported
- 204
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 232 adverse event reports that list this exact device (DI 05708932539135), received between Jul 24, 2017 and Jun 9, 2026 — classified as 204 injuries, 28 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (29 reports).
18 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem37
- Fluid/Blood Leak29
- Inadequacy of Device Shape and/or Size26
- Pumping Problem26
- Insufficient Device Problem Information19
- Inflation Problem18
- Fracture16
- Appropriate Device Problem Term/Code Not Available10
Patient problems
- No Clinical Signs, Symptoms or Conditions112
- No Information37
- Unspecified Infection32
- Erosion11
- No Consequences Or Impact To Patient9
- Pain9
- No Known Impact Or Consequence To Patient7
- Bacterial Infection5
Recent adverse event reports
The 25 most recent of 232 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2026Inadequacy of Device Shape and/or Size
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO INCORRECT SIZING (TOO LONG).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25448513 ↗InjuryMay 20, 2026Inadequacy of Device Shape and/or Size
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO INCORRECT SIZING (TOO SHORT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229610 ↗InjuryMay 4, 2026Fluid/Blood Leak
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25072483 ↗InjuryApr 9, 2026Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE AVAILABLE INFORMATION THE TITAN TOUCH WAS REMOVED AND REPLACED WITH A GENESIS DUE TO EROSION.
Patient problems: Erosion
MAUDE report 24830095 ↗InjuryMar 30, 2026Fluid/Blood Leak; Insufficient Device Problem Information
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH DUE TO AN UNSPECIFIED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24715960 ↗InjuryMar 23, 2026Fluid/Blood Leak
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED AS THE DOCTOR SUSPECTED A LEAK; HOWEVER, THE LEAK WAS NOT IDENTIFIED DURING THE EXPLANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24662340 ↗InjuryMar 16, 2026Fracture
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED DUE TO BROKEN TUBING FROM THE RESERVOIR TO THE PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24611981 ↗InjuryFeb 24, 2026Insufficient Device Problem Information
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED DUE TO PAIN AND AN UNSPECIFIED MALFUNCTION.
Patient problems: Pain
MAUDE report 24430574 ↗InjuryFeb 20, 2026Malposition of Device; Pumping Problem
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PUMP DIFFICULT TO PUMP & RESERVOIR PLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24404531 ↗InjuryFeb 18, 2026Inadequacy of Device Shape and/or Size
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO INCORRECT SIZING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24377655 ↗InjuryFeb 18, 2026Insufficient Device Problem Information
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS REMOVED AND REPLACED DUE TO AN UNKNOWN REASON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24377451 ↗InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A GENESIS DUE TO A SCROTAL INFECTION.
Patient problems: Unspecified Infection
MAUDE report 24200493 ↗InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE AVAILABLE INFORMATION, THE PUMP BULB HAS ATTACHED TO PATIENT'S INNER SCROTUM. THE PATIENT HAD A STAPH INFECTION POST IMPLANTATION PROCEDURE THAT LASTED MULTIPLE MONTHS. THE FIRST DOCTOR REPORTEDLY SAW NO ISSUES; HOWEVER, A SECOND DOCTOR NOTED POSSIBILITY FOR EROSION ISSUES AND RECOMMENDED REPLACEMENT OF PUMP.
Patient problems: Adhesion(s); Bacterial Infection
MAUDE report 24163949 ↗MalfunctionJan 14, 2026Failure to Deflate
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT IS UNABLE TO DEFLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24073029 ↗MalfunctionJan 13, 2026Unintended Deflation
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WILL NOT STAY RIGID [AUTO-DEFLATION].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24059610 ↗InjuryJan 5, 2026Fluid/Blood Leak; Pumping Problem
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE A LEAK IN THE RESERVOIR TUBING. THE DEVICE WOULDN'T PUMP AFTER 5 CYCLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23956147 ↗InjuryJan 2, 2026Fluid/Blood Leak
ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A FLUID LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23950655 ↗InjuryDec 18, 2025Failure to Pump; Failure to Pump
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PUMP NON-FUNCTIONING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23838302 ↗InjuryDec 11, 2025Material Deformation; Material Deformation
ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO AN ANEURYSM OF THE CYLINDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23778921 ↗InjuryNov 20, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO INCORRECT SIZING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23604909 ↗InjuryNov 3, 2025Fluid/Blood Leak; Fluid/Blood Leak
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO WHAT APPEARED TO BE A LEAK WITH THE CONNECTORS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23450409 ↗InjuryOct 30, 2025Malposition of Device; Malposition of Device
ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO POSSIBLE EXTRUSION [EROSION]. THE IMPLANT ON THE LEFT SIDE FOLDED ON TO ITSELF AND WOULD NOT STRAIGHTEN OUT.
Patient problems: Erosion
MAUDE report 23430454 ↗InjuryOct 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 23254551 ↗InjuryAug 4, 2025Fluid/Blood Leak; Fluid/Blood Leak
ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22706162 ↗InjuryJul 8, 2025Material Puncture/Hole; Material Puncture/Hole
ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A HOLE IN THE CYLINDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22440090 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.
- Z-1361-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
- Z-1363-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
- Z-1369-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
- Z-1372-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
- Z-1374-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
- Z-1367-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
- Z-1365-2023 · Mar 1, 2023 · Open, ClassifiedTitan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
- Z-1371-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
- Z-1370-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
- Z-1366-2023 · Mar 1, 2023 · Open, ClassifiedTitan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
- Z-1360-2023 · Mar 1, 2023 · Open, ClassifiedTitan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
- Z-1364-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
- Z-1373-2023 · Mar 1, 2023 · Open, ClassifiedTitan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
- Z-1362-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
- Z-1359-2023 · Mar 1, 2023 · Open, ClassifiedTitan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
…and 3 more.
Other Device, Impotence, Mechanical/Hydraulic devices
- Titan ES2920Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall862 reports
- Titan ES2922Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall1 death758 reports
- Titan ES2918Coloplast A/S · Device, Impotence, Mechanical/Hydraulic577 reports
- Titan ES8920Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall476 reports
- Titan ES8922Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall334 reports
- Titan 517750Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall258 reports
- Titan ES8918Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall225 reports
- Titan ES2818Coloplast A/S · Device, Impotence, Mechanical/Hydraulic1 recall222 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.