Titan ES2916Recalls & reported problems

Coloplast A/S · Titan® Touch Scrotal Zero Degree Angle Cylinder Set with Pump · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Coloplast Titan Touch penile implant is an inflatable pump implant used for erectile dysfunction. Many of the 179 reports describe fluid leaks and broken tubing, and several implants were removed. The reports include 154 injuries, 25 malfunctions and no deaths. Reports from the last 12 months number 5, compared with 20 in the 12 months before. One recall, started in March 2023 and still open, concerns a thinner pump wall that may fail early.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
178
Deaths reported
0
Injuries reported
153
Recalls naming it
1

Summary

The FDA's MAUDE database contains 178 adverse event reports that list this exact device (DI 05708932539173), received between Jul 24, 2017 and Jun 11, 2026 — classified as 153 injuries, 25 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (28 reports).

1 FDA recall record name this device's identifier, 1 still open.

17 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 3 reportsJan 2024: 6 reportsFeb 2024: 8 reportsMar 2024: 5 reportsApr 2024: 6 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 4 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

    Z-1369-2023 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak28
  • Fracture26
  • Adverse Event Without Identified Device or Use Problem25
  • Insufficient Device Problem Information18
  • Inflation Problem16
  • Pumping Problem15
  • Inadequacy of Device Shape and/or Size13
  • Break10

Patient problems

  • No Clinical Signs, Symptoms or Conditions91
  • No Information28
  • Unspecified Infection24
  • No Consequences Or Impact To Patient8
  • No Known Impact Or Consequence To Patient6
  • Erosion5
  • Bacterial Infection4
  • Insufficient Information3

Recent adverse event reports

The 25 most recent of 178 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2026Malposition of Device

    ACCORDING TO THE AVAILABLE INFORMATION, THE PUMP WAS REMOVED AND REPLACED DUE TO PATIENT'S PUMP WAS RIDING UP, NEEDED MORE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477369 ↗
  • InjuryApr 21, 2026Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24938167 ↗
  • MalfunctionFeb 12, 2026Fluid/Blood Leak; Air/Gas in Device

    ACCORDING TO THE AVAILABLE INFORMATION, DURING INTRAOPERATIVE PREPARATION OF A 16 CM TOUCH SCROTAL DEVICE, A PERSISTENT AIR INFILTRATION ISSUE WAS OBSERVED. DESPITE MULTIPLE PREPARATION ATTEMPTS, AIR CONTINUED TO RETURN TO THE PUMP MECHANISM. THE DEVICE WAS DETERMINED TO HAVE A SUSPECTED MANUFACTURING DEFECT [LEAK] AND WAS REPLACED WITH A NEW UNIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322743 ↗
  • InjuryNov 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED WITH A MALLEABLE DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 23555139 ↗
  • InjuryOct 31, 2025Material Frayed; Material Frayed

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO FRAYED TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23438983 ↗
  • InjuryOct 24, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO IT BEING TOO SHORT AND PATIENT PREFERENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23384491 ↗
  • InjurySep 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED WITH A MALLEABLE DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 22979087 ↗
  • InjuryAug 19, 2025Pumping Problem; Pumping Problem

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO AN UNKNOWN PUMP DIFFICULTY. REPLACED BOTH PUMP AND CYLINDERS AND CONNECTED TO EXISTING RESERVOIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22839473 ↗
  • InjuryAug 11, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A BREAK [FRACTURE] IN THE CYLINDER TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22770759 ↗
  • InjuryJul 15, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO DEFECTIVE TUBING [FRACTURED TUBING].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22510332 ↗
  • InjuryJul 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED WITH A MALLEABLE DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 22463611 ↗
  • InjuryMay 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED FOR AN UNKNOWN REASON. DETAILS SURROUNDING THE EXPLANT WERE NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22080824 ↗
  • InjuryMay 13, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS REMOVED AND REPLACED DUE TO A BREAK (FRACTURE) IN THE TUBING WHERE THE TUBING MEETS THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22012805 ↗
  • InjuryMay 7, 2025Inflation Problem; Material Perforation; Inflation Problem; Material Perforation

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS REMOVED AND REPLACED DUE TO NOT PUMPING UP. THERE WAS A POSSIBLE SUTURE PERF IN THE LEFT CYLINDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21976599 ↗
  • MalfunctionMay 1, 2025Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE AVAILABLE INFORMATION, DURING VIRGIN IPP IMPLANTATION, THERE WAS A SUTURE NEEDLE PERFORATION ON THE RIGHT SIDE OF THE IMPLANT, WHICH THEN CAUSED THE SYSTEM TO REMOVE AND REPLACE THAT CYLINDER WITH A DIFFERENT LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21944286 ↗
  • InjuryMar 17, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A PUMP TUBING FRACTURE AT THE JUNCTION OF THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21621354 ↗
  • InjuryMar 17, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO A TUBING MALFUNCTION [FRACTURE].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21620833 ↗
  • InjuryMar 11, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO EROSION / EXTRUSION.

    Patient problems: Erosion

    MAUDE report 21572594 ↗
  • InjuryFeb 7, 2025Material Frayed; Material Frayed

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO TUBING FRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21333845 ↗
  • InjuryFeb 4, 2025Fracture; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO BROKEN TUBING THAT LEFT THE RESERVOIR EMPTY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21304197 ↗
  • InjuryNov 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO EROSION.

    Patient problems: Erosion

    MAUDE report 20784319 ↗
  • InjuryNov 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO FLUID LOSS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20781067 ↗
  • MalfunctionNov 4, 2024Pumping Problem; Pumping Problem

    ACCORDING TO THE AVAILABLE INFORMATION, AFTER PREPARATION OF AN IMPLANT WITHOUT ISSUE, AND SUCCESSFUL IMPLANTATION OF THE PUMP AND CYLINDERS, THE PUMP WOULD NOT REFILL. IMPLANT WAS REMOVED AND REPLACED WITH ANOTHER DEVICE SUCCESSFULLY DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20599885 ↗
  • InjuryNov 4, 2024Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING BREAK THAT LED TO FLUID LOSS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20598226 ↗
  • InjuryOct 30, 2024Deflation Problem; Inadequacy of Device Shape and/or Size; Deflation Problem; Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE IMPLANT BEING UNDERSIZED & THE PATIENT COULD NOT USE IT CORRECTLY TO DEFLATE THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20566734 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis

…and 2 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →