Areus(TM) Endobronchial Ultrasound Aspiration Needle EBUS00001Recalls & reported problems
Micro-Tech (Nanjing) Co., Ltd. · Maximum insertion portion diameter:1.8mm,Needle:19G,Maximum needle length:40mm,Material: Nitinol, Working channel: is more than or equal to 2.0mm,Working Length:720~760mm,FNA needle · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06932503549219), received between Jan 20, 2026 and Sep 21, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Display or Visual Feedback Problem1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Display or Visual Feedback Problem
It is reported during endobronchial ultrasound case, endobronchial ultrasound scope (b)(6) broke related to thoracent/areus endobronchial ultrasound aspiration needle.Scope working at beginning of case. Passed needle through channel, scope still working. Attempted to deploy needle and image on screen became pixelated. Removed needle and attempted to reinsert/redeploy, and pixelation on screen became worse. Pixelation on screen became worse with every attempt to redeploy needle.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50835423 ↗MalfunctionJan 20, 2026Material Puncture/Hole
IT WAS REPORTED THAT "NEEDLE CAME THROUGH SHEATH- REMOVED FROM FIELD AND REPLACED WITH NEW ONE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24112405 ↗
Other Bronchoscope Accessory devices
- ViziShot 2 FLEX NA-U403SX-4019Gyrus ACMI, LLC · Bronchoscope accessory1 recall349 reports
- PeriView FLEX NA-403D-2021Gyrus ACMI, LLC · Bronchoscope accessory54 reports
- INSPIRA AIR BC1040AZAcclarent, Inc. · BRONCHOSCOPE ACCESSORY1 recall0 reports
- CRE Pulmonary M00550340Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY26 reports
- CRE Pulmonary M00550350Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY21 reports
- CRE Pulmonary M00550310Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY21 reports
- CRE Pulmonary M00550330Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY20 reports
- CRE Pulmonary M00550320Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY15 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.