Jade PTA Balloon Dilatation Catheter 585015022Recalls & reported problems

OrbusNeich Medical (ShenZhen) Co., Ltd. · Percutaneous Transluminal Angioplasty (PTA) Catheter · Class 2

All models of Jade PTA Balloon Dilatation Catheter (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06934955901270), received between Mar 29, 2025 and Apr 4, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture1
  • Material Separation1
  • Material Split, Cut or Torn1

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 4, 2025Material Rupture; Material Separation; Material Rupture; Material Separation

    IT WAS REPORTED THAT A NON-DILATABLE, HEAVILY CALCIFIED, HEAVILY TORTUOUS SUPERFICIAL FEMORAL ARTERY WITH CHRONIC TOTAL OCCLUSION WAS ATTEMPTED TO BE TREATED WITH SEVERAL UNSPECIFIED BALLOONS, ALL UNSUCCESSFULLY. WHEN A JADE PTA OTW 5 X 150 MM BALLOON WAS ATTEMPTED TO BE USED, IT RUPTURED WHEN IT REACHED ROUGHLY 22-24 ATMOSPHERES. THE RUPTURED COMPONENT CAME ENTIRELY OFF THE SHAFT. THERE WAS A DELAY OF A FEW HOURS WHILE THE PHYSICIAN PLACED STENTS. ADDITIONALLY, A CUT DOWN OF A CLOT IN THE COMMON FEMORAL WAS PERFORMED. THE HYPOTUBE WAS ABLE TO BE REMOVED FROM THE COMMON FEMORAL DURING THIS TIME, HOWEVER, THE RUPTURED COMPONENTS REMAINED IN-VIVO. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE COMPONENTS; HOWEVER, THEY WERE DIFFICULT TO BE IDENTIFIED WITH ANGIOGRAPHIC IMAGING AND THEREFORE, STENTS WERE PLACED IN THE AREA WHERE THE BALLOON RUPTURED. THE PHYSICIAN BELIEVED THE STENTS WERE ABLE TO COVER THE RUPTURED COMPONENT. AT THE END OF THE PROCEDURE, IT WAS INDICATED THAT THE FLOW WAS SUFFICIENT AND THE PROCEDURE WAS COMPLETED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: Foreign Body In Patient

    MAUDE report 21775326 ↗
  • MalfunctionMar 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT A NON-DILATABLE, HEAVILY CALCIFIED, HEAVILY TORTUOUS SUPERFICIAL FEMORAL ARTERY WITH CHRONIC TOTAL OCCLUSION WAS ATTEMPTED TO BE TREATED WITH SEVERAL UNSPECIFIED BALLOONS, ALL UNSUCCESSFULLY. WHEN A JADE PTA OTW 5 X 150 MM BALLOON WAS ATTEMPTED TO BE USED, IT RUPTURED WHEN IT REACHED ROUGHLY 22-24 ATMOSPHERES. THE RUPTURED COMPONENT CAME ENTIRELY OFF THE SHAFT. THERE WAS A DELAY OF A FEW HOURS WHILE THE PHYSICIAN PLACED STENTS. ADDITIONALLY, A CUT DOWN OF A CLOT IN THE COMMON FEMORAL WAS PERFORMED. THE HYPOTUBE WAS ABLE TO BE REMOVED FROM THE COMMON FEMORAL DURING THIS TIME, HOWEVER, THE RUPTURED COMPONENTS REMAINED IN-VIVO. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE COMPONENTS; HOWEVER, THEY WERE DIFFICULT TO BE IDENTIFIED WITH ANGIOGRAPHIC IMAGING AND THEREFORE, STENTS WERE PLACED IN THE AREA WHERE THE BALLOON RUPTURED. THE PHYSICIAN BELIEVED THE STENTS WERE ABLE TO COVER THE RUPTURED COMPONENT. AT THE END OF THE PROCEDURE, IT WAS INDICATED THAT THE FLOW WAS SUFFICIENT AND THE PROCEDURE WAS COMPLETED. THE PATIENT WAS IN STABLE CONDITION. NO CASE IMAGE/CD AND OTHER INFORMATION WAS PROVIDED FOR ANALYSIS.

    Patient problems: Foreign Body In Patient

    MAUDE report 21728429 ↗